Comparison of central corneal thickness measurements obtained by OrbscanIIz(®) and ultrasound pachymetry: A concordance study in the usual clinical practice
PURPOSE: The objective of this study was to assess the concordance between the values obtained in measuring central corneal thickness using the OrbscanIIz(®) and the contact ultrasonic pachymeter available in our public ophthalmology service. METHODS: Measurements were taken from 88 eyes of 44 patie...
Autores principales: | Romero-Trevejo, José L., Sánchez-Pérez, Andrés, Muñoz-García, Erika, Fernández-Romero, Lourdes, Jiménez-Navarro, Manuel F. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10032292/ https://www.ncbi.nlm.nih.gov/pubmed/36968780 http://dx.doi.org/10.4103/sjopt.Sjopt_82_21 |
Ejemplares similares
-
Measurement of Central Corneal Thickness Using Ultrasound Pachymetry and Orbscan II in Normal Eyes
por: Sadoughi, Mohammad Mehdi, et al.
Publicado: (2015) -
Agreement between Clinical History Method, Orbscan IIz, and Pentacam in Estimating Corneal Power after Myopic Excimer Laser Surgery
por: Lekhanont, Kaevalin, et al.
Publicado: (2015) -
Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and Orbscan
por: Khaja, Wassia A, et al.
Publicado: (2015) -
Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy
por: Nassiri, Nader, et al.
Publicado: (2014) -
Central Corneal Thickness in Highly Myopic Eyes: Interdevice Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy: Erratum
Publicado: (2015)