Same same but different: dopamine transporter SPECT on scanners with CZT vs. NaI detectors
BACKGROUND: The aims of this study were to establish a normal database (NDB) for semiquantification of dopamine transporter (DAT) single-photon emission computed tomography (SPECT) with [(123)I]FP-CIT on a cadmium zinc telluride (CZT) camera, test the preexisting NaI-derived NDB for use in CZT scans...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10033816/ https://www.ncbi.nlm.nih.gov/pubmed/36949290 http://dx.doi.org/10.1186/s13550-023-00973-8 |
Sumario: | BACKGROUND: The aims of this study were to establish a normal database (NDB) for semiquantification of dopamine transporter (DAT) single-photon emission computed tomography (SPECT) with [(123)I]FP-CIT on a cadmium zinc telluride (CZT) camera, test the preexisting NaI-derived NDB for use in CZT scans, and compare the diagnostic findings in subjects imaged with a CZT scanner with either the preexisting NaI-based NDB or our newly defined CZT NDB. METHODS: The sample comprised 73 subjects with clinically uncertain parkinsonian syndrome (PS) who prospectively underwent [(123)I]FP-CIT SPECT on a CZT camera according to standard guidelines with identical acquisition and reconstruction protocols (DaTQUANT). Two experienced readers visually assessed the images and binarized the subjects into “non-neurodegenerative PS” and “neurodegenerative PS”. Twenty-five subjects from the “non-neurodegenerative PS” subgroup were randomly selected to establish a CZT NDB. The remaining 48 subjects were defined as “test group”. DaTQUANT was used to determine the specific binding ratio (SBR). For the test group, SBR values were transformed to z-scores for the putamen utilizing both the CZT NDB and the manufacturer-provided NaI-based NDB (GE NDB). A predefined fixed cut-off of -2 was used for dichotomization of z-scores to classify neurodegenerative and non-neurodegenerative PS. Performance of semiquantification using the two NDB to identify subjects with neurodegenerative PS was assessed in comparison with the visual rating. Furthermore, a randomized head-to-head comparison of both detector systems was performed semiquantitatively in a subset of 32 out of all 73 subjects. RESULTS: Compared to the visual rating as reference, semiquantification based on the dedicated CZT NDB led to fewer discordant ratings than the GE NDB in CZT scans (3 vs. 8 out of 48 subjects). This can be attributed to the putaminal z-scores being consistently higher with the GE NDB on a CZT camera (median absolute difference of 1.68), suggesting an optimal cut-off of -0.5 for the GE NDB instead of -2.0. Average binding ratios and z-scores were significantly lower in CZT compared to NaI data. CONCLUSIONS: Use of a dedicated, CZT-derived NDB is recommended in [(123)I]FP-CIT SPECT with a CZT camera since it improves agreement between semiquantification and visual assessment. |
---|