Development of a multi-purpose quality control phantom for MRI-based treatment planning in high-dose-rate brachytherapy of cervical cancer

PURPOSE: Suitable commissioning and quality control (QC) tests for high-dose-rate brachytherapy (HDR-BT) is necessary to ensure dosimetric and geometric accuracy of the treatment. This study aimed to present the methodology of developing a novel multi-purpose QC phantom (AQuA-BT) and examples of its...

Descripción completa

Detalles Bibliográficos
Autores principales: Kanani, Abolfazl, Owrangi, Amir, Yazdi, Mehran, Fatemi-Ardekani, Ali, Mosleh-Shirazi, Mohammad Amin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10034728/
https://www.ncbi.nlm.nih.gov/pubmed/36970435
http://dx.doi.org/10.5114/jcb.2023.125014
_version_ 1784911270096404480
author Kanani, Abolfazl
Owrangi, Amir
Yazdi, Mehran
Fatemi-Ardekani, Ali
Mosleh-Shirazi, Mohammad Amin
author_facet Kanani, Abolfazl
Owrangi, Amir
Yazdi, Mehran
Fatemi-Ardekani, Ali
Mosleh-Shirazi, Mohammad Amin
author_sort Kanani, Abolfazl
collection PubMed
description PURPOSE: Suitable commissioning and quality control (QC) tests for high-dose-rate brachytherapy (HDR-BT) is necessary to ensure dosimetric and geometric accuracy of the treatment. This study aimed to present the methodology of developing a novel multi-purpose QC phantom (AQuA-BT) and examples of its’ application in 3D image-based (particularly magnetic resonance imaging [MRI]-based) planning for cervix BT. MATERIAL AND METHODS: Design criteria led to a phantom with sufficient size waterproof box for dosimetry and capability for inserting other components inside the phantom for: (A) Validating dose calculation algorithms in treatment planning systems (TPSs) using a small-volume ionization chamber; (B) Testing volume calculation accuracy in TPSs for bladder, rectum, and sigmoid organs at risk (OARs) constructed by 3D printing; (C) Quantification of MRI distortions using 17 semi-elliptical plates with 4,317 control points to mimic a realistic female’s pelvis size; and (D) Quantification of image distortions and artifacts induced by MRI-compatible applicators using a specific radial fiducial marker. The utility of the phantom was tested in various QC procedures. RESULTS: The phantom was successfully implemented for examples of intended QC procedures. The maximum deviation between the absorbed doses to water assessed with our phantom and those calculated by SagiPlan TPS was 1.7%. The mean discrepancy in volumes of TPS-calculated OARs was 1.1%. The differences between known distances within the phantom on MR imaging were within 0.7 mm compared with computed tomography. CONCLUSIONS: This phantom is a promising useful tool for dosimetric and geometric quality assurance (QA) in MRI-based cervix BT.
format Online
Article
Text
id pubmed-10034728
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Termedia Publishing House
record_format MEDLINE/PubMed
spelling pubmed-100347282023-03-24 Development of a multi-purpose quality control phantom for MRI-based treatment planning in high-dose-rate brachytherapy of cervical cancer Kanani, Abolfazl Owrangi, Amir Yazdi, Mehran Fatemi-Ardekani, Ali Mosleh-Shirazi, Mohammad Amin J Contemp Brachytherapy Original Paper PURPOSE: Suitable commissioning and quality control (QC) tests for high-dose-rate brachytherapy (HDR-BT) is necessary to ensure dosimetric and geometric accuracy of the treatment. This study aimed to present the methodology of developing a novel multi-purpose QC phantom (AQuA-BT) and examples of its’ application in 3D image-based (particularly magnetic resonance imaging [MRI]-based) planning for cervix BT. MATERIAL AND METHODS: Design criteria led to a phantom with sufficient size waterproof box for dosimetry and capability for inserting other components inside the phantom for: (A) Validating dose calculation algorithms in treatment planning systems (TPSs) using a small-volume ionization chamber; (B) Testing volume calculation accuracy in TPSs for bladder, rectum, and sigmoid organs at risk (OARs) constructed by 3D printing; (C) Quantification of MRI distortions using 17 semi-elliptical plates with 4,317 control points to mimic a realistic female’s pelvis size; and (D) Quantification of image distortions and artifacts induced by MRI-compatible applicators using a specific radial fiducial marker. The utility of the phantom was tested in various QC procedures. RESULTS: The phantom was successfully implemented for examples of intended QC procedures. The maximum deviation between the absorbed doses to water assessed with our phantom and those calculated by SagiPlan TPS was 1.7%. The mean discrepancy in volumes of TPS-calculated OARs was 1.1%. The differences between known distances within the phantom on MR imaging were within 0.7 mm compared with computed tomography. CONCLUSIONS: This phantom is a promising useful tool for dosimetric and geometric quality assurance (QA) in MRI-based cervix BT. Termedia Publishing House 2023-02-10 2023-02 /pmc/articles/PMC10034728/ /pubmed/36970435 http://dx.doi.org/10.5114/jcb.2023.125014 Text en Copyright © 2023 Termedia https://creativecommons.org/licenses/by-nc-sa/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/ (https://creativecommons.org/licenses/by-nc-sa/4.0/) )
spellingShingle Original Paper
Kanani, Abolfazl
Owrangi, Amir
Yazdi, Mehran
Fatemi-Ardekani, Ali
Mosleh-Shirazi, Mohammad Amin
Development of a multi-purpose quality control phantom for MRI-based treatment planning in high-dose-rate brachytherapy of cervical cancer
title Development of a multi-purpose quality control phantom for MRI-based treatment planning in high-dose-rate brachytherapy of cervical cancer
title_full Development of a multi-purpose quality control phantom for MRI-based treatment planning in high-dose-rate brachytherapy of cervical cancer
title_fullStr Development of a multi-purpose quality control phantom for MRI-based treatment planning in high-dose-rate brachytherapy of cervical cancer
title_full_unstemmed Development of a multi-purpose quality control phantom for MRI-based treatment planning in high-dose-rate brachytherapy of cervical cancer
title_short Development of a multi-purpose quality control phantom for MRI-based treatment planning in high-dose-rate brachytherapy of cervical cancer
title_sort development of a multi-purpose quality control phantom for mri-based treatment planning in high-dose-rate brachytherapy of cervical cancer
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10034728/
https://www.ncbi.nlm.nih.gov/pubmed/36970435
http://dx.doi.org/10.5114/jcb.2023.125014
work_keys_str_mv AT kananiabolfazl developmentofamultipurposequalitycontrolphantomformribasedtreatmentplanninginhighdoseratebrachytherapyofcervicalcancer
AT owrangiamir developmentofamultipurposequalitycontrolphantomformribasedtreatmentplanninginhighdoseratebrachytherapyofcervicalcancer
AT yazdimehran developmentofamultipurposequalitycontrolphantomformribasedtreatmentplanninginhighdoseratebrachytherapyofcervicalcancer
AT fatemiardekaniali developmentofamultipurposequalitycontrolphantomformribasedtreatmentplanninginhighdoseratebrachytherapyofcervicalcancer
AT moslehshirazimohammadamin developmentofamultipurposequalitycontrolphantomformribasedtreatmentplanninginhighdoseratebrachytherapyofcervicalcancer