Performance of a System for Rapid Phenotypic Antimicrobial Susceptibility Testing of Gram-Negative Bacteria Directly from Positive Blood Culture Bottles
The rapid administration of optimal antimicrobial treatment is paramount for the treatment of bloodstream infections (BSIs), and rapid antimicrobial susceptibility testing (AST) results are essential. Q-linea has developed the ASTar system, a rapid phenotypic AST device. Here, we report the performa...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
American Society for Microbiology
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10035315/ https://www.ncbi.nlm.nih.gov/pubmed/36852983 http://dx.doi.org/10.1128/jcm.01525-22 |
_version_ | 1784911392954908672 |
---|---|
author | Göransson, J. Sundqvist, M. Ghaderi, E. Lisby, J. G. Molin, Y. Eriksson, E. Carlsson, S. Cederlöf, A. Ellis, L. Melin, J. |
author_facet | Göransson, J. Sundqvist, M. Ghaderi, E. Lisby, J. G. Molin, Y. Eriksson, E. Carlsson, S. Cederlöf, A. Ellis, L. Melin, J. |
author_sort | Göransson, J. |
collection | PubMed |
description | The rapid administration of optimal antimicrobial treatment is paramount for the treatment of bloodstream infections (BSIs), and rapid antimicrobial susceptibility testing (AST) results are essential. Q-linea has developed the ASTar system, a rapid phenotypic AST device. Here, we report the performance of the ASTar BC G− (Gram-negative) kit when assessed according to the ISO 20776-2:2007 standard for performance evaluation of in vitro diagnostic AST devices. The evaluated ASTar BC G− kit uses a broad panel of 23 antimicrobials for the treatment of BSIs caused by Gram-negative fastidious and nonfastidious bacteria across a range of 6 to 14 2-fold dilutions, including cefoxitin as a screening agent for AmpC-producing Enterobacterales. The ASTar system processes blood culture samples to generate data on MICs and susceptible, intermediate, or resistant (SIR) category. The automated protocol includes concentration determination and concentration adjustment to enable a controlled inoculum, followed by broth microdilution (BMD) and microscopy performed continuously to generate MIC values within approximately 6 h once the test is run on the ASTar system. The performance of the ASTar system was assessed against the ISO 20776-2:2007 standard BMD reference method. Testing was performed across three sites, with results from 412 contrived blood cultures and 74 fresh clinical blood cultures. The ASTar system was also tested for reproducibility, with triplicate testing of 11 strains. The accuracy study comprised 8,650 data points of bacterium-antimicrobial tests. The ASTar system demonstrated an overall essential agreement (EA) of 95.8% (8,283/8,650) and a categorical agreement (CA) of 97.6% (8,433/8,639) compared to the reference BMD method. The overall rate of major discrepancies (MDs) was 0.9% (62/6,845), and that of very major discrepancies (VMDs) was 2.4% (30/1,239). This study shows that the ASTar system delivers reproducible results with overall EA and CA of >95%. |
format | Online Article Text |
id | pubmed-10035315 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | American Society for Microbiology |
record_format | MEDLINE/PubMed |
spelling | pubmed-100353152023-03-24 Performance of a System for Rapid Phenotypic Antimicrobial Susceptibility Testing of Gram-Negative Bacteria Directly from Positive Blood Culture Bottles Göransson, J. Sundqvist, M. Ghaderi, E. Lisby, J. G. Molin, Y. Eriksson, E. Carlsson, S. Cederlöf, A. Ellis, L. Melin, J. J Clin Microbiol Bacteriology The rapid administration of optimal antimicrobial treatment is paramount for the treatment of bloodstream infections (BSIs), and rapid antimicrobial susceptibility testing (AST) results are essential. Q-linea has developed the ASTar system, a rapid phenotypic AST device. Here, we report the performance of the ASTar BC G− (Gram-negative) kit when assessed according to the ISO 20776-2:2007 standard for performance evaluation of in vitro diagnostic AST devices. The evaluated ASTar BC G− kit uses a broad panel of 23 antimicrobials for the treatment of BSIs caused by Gram-negative fastidious and nonfastidious bacteria across a range of 6 to 14 2-fold dilutions, including cefoxitin as a screening agent for AmpC-producing Enterobacterales. The ASTar system processes blood culture samples to generate data on MICs and susceptible, intermediate, or resistant (SIR) category. The automated protocol includes concentration determination and concentration adjustment to enable a controlled inoculum, followed by broth microdilution (BMD) and microscopy performed continuously to generate MIC values within approximately 6 h once the test is run on the ASTar system. The performance of the ASTar system was assessed against the ISO 20776-2:2007 standard BMD reference method. Testing was performed across three sites, with results from 412 contrived blood cultures and 74 fresh clinical blood cultures. The ASTar system was also tested for reproducibility, with triplicate testing of 11 strains. The accuracy study comprised 8,650 data points of bacterium-antimicrobial tests. The ASTar system demonstrated an overall essential agreement (EA) of 95.8% (8,283/8,650) and a categorical agreement (CA) of 97.6% (8,433/8,639) compared to the reference BMD method. The overall rate of major discrepancies (MDs) was 0.9% (62/6,845), and that of very major discrepancies (VMDs) was 2.4% (30/1,239). This study shows that the ASTar system delivers reproducible results with overall EA and CA of >95%. American Society for Microbiology 2023-02-28 /pmc/articles/PMC10035315/ /pubmed/36852983 http://dx.doi.org/10.1128/jcm.01525-22 Text en Copyright © 2023 Göransson et al. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 International license (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Bacteriology Göransson, J. Sundqvist, M. Ghaderi, E. Lisby, J. G. Molin, Y. Eriksson, E. Carlsson, S. Cederlöf, A. Ellis, L. Melin, J. Performance of a System for Rapid Phenotypic Antimicrobial Susceptibility Testing of Gram-Negative Bacteria Directly from Positive Blood Culture Bottles |
title | Performance of a System for Rapid Phenotypic Antimicrobial Susceptibility Testing of Gram-Negative Bacteria Directly from Positive Blood Culture Bottles |
title_full | Performance of a System for Rapid Phenotypic Antimicrobial Susceptibility Testing of Gram-Negative Bacteria Directly from Positive Blood Culture Bottles |
title_fullStr | Performance of a System for Rapid Phenotypic Antimicrobial Susceptibility Testing of Gram-Negative Bacteria Directly from Positive Blood Culture Bottles |
title_full_unstemmed | Performance of a System for Rapid Phenotypic Antimicrobial Susceptibility Testing of Gram-Negative Bacteria Directly from Positive Blood Culture Bottles |
title_short | Performance of a System for Rapid Phenotypic Antimicrobial Susceptibility Testing of Gram-Negative Bacteria Directly from Positive Blood Culture Bottles |
title_sort | performance of a system for rapid phenotypic antimicrobial susceptibility testing of gram-negative bacteria directly from positive blood culture bottles |
topic | Bacteriology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10035315/ https://www.ncbi.nlm.nih.gov/pubmed/36852983 http://dx.doi.org/10.1128/jcm.01525-22 |
work_keys_str_mv | AT goranssonj performanceofasystemforrapidphenotypicantimicrobialsusceptibilitytestingofgramnegativebacteriadirectlyfrompositivebloodculturebottles AT sundqvistm performanceofasystemforrapidphenotypicantimicrobialsusceptibilitytestingofgramnegativebacteriadirectlyfrompositivebloodculturebottles AT ghaderie performanceofasystemforrapidphenotypicantimicrobialsusceptibilitytestingofgramnegativebacteriadirectlyfrompositivebloodculturebottles AT lisbyjg performanceofasystemforrapidphenotypicantimicrobialsusceptibilitytestingofgramnegativebacteriadirectlyfrompositivebloodculturebottles AT moliny performanceofasystemforrapidphenotypicantimicrobialsusceptibilitytestingofgramnegativebacteriadirectlyfrompositivebloodculturebottles AT erikssone performanceofasystemforrapidphenotypicantimicrobialsusceptibilitytestingofgramnegativebacteriadirectlyfrompositivebloodculturebottles AT carlssons performanceofasystemforrapidphenotypicantimicrobialsusceptibilitytestingofgramnegativebacteriadirectlyfrompositivebloodculturebottles AT cederlofa performanceofasystemforrapidphenotypicantimicrobialsusceptibilitytestingofgramnegativebacteriadirectlyfrompositivebloodculturebottles AT ellisl performanceofasystemforrapidphenotypicantimicrobialsusceptibilitytestingofgramnegativebacteriadirectlyfrompositivebloodculturebottles AT melinj performanceofasystemforrapidphenotypicantimicrobialsusceptibilitytestingofgramnegativebacteriadirectlyfrompositivebloodculturebottles |