Cargando…

An Extremely Inhomogeneous Gross Tumor Dose is Suitable for Volumetric Modulated Arc-Based Radiosurgery with a 5-mm Leaf-Width Multileaf Collimator for Single Brain Metastasis

Introduction Single or multi-fraction (mf) stereotactic radiosurgery (SRS) is an indispensable treatment option for brain metastases (BMs). The integration of volumetric modulated arc therapy (VMAT) into linac-based SRS is expected to further enhance efficacy and safety and to expand the indications...

Descripción completa

Detalles Bibliográficos
Autores principales: Ohtakara, Kazuhiro, Suzuki, Kojiro
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10043638/
https://www.ncbi.nlm.nih.gov/pubmed/36999102
http://dx.doi.org/10.7759/cureus.35467
_version_ 1784913195516821504
author Ohtakara, Kazuhiro
Suzuki, Kojiro
author_facet Ohtakara, Kazuhiro
Suzuki, Kojiro
author_sort Ohtakara, Kazuhiro
collection PubMed
description Introduction Single or multi-fraction (mf) stereotactic radiosurgery (SRS) is an indispensable treatment option for brain metastases (BMs). The integration of volumetric modulated arc therapy (VMAT) into linac-based SRS is expected to further enhance efficacy and safety and to expand the indications for the challenging type of BMs. However, the optimal treatment design and relevant optimization method for volumetric modulated arc-based radiosurgery (VMARS) remain unestablished with substantial inter-institutional differences. Therefore, this study was conducted to determine the optimal dose distribution suitable for VMARS of BMs, especially regarding dose inhomogeneity of the gross tumor volume (GTV). The GTV boundary, not margin-added planning target volume, was regarded as a basis for planning optimization and dose prescription. Materials and methods This was a planning study for the clinical scenario of a single BM. Eight sphere-shaped objects with diameters of 5-40 mm in 5-mm increments were assumed as GTVs. The treatment system included a 5-mm leaf width multileaf collimator (MLC) Agility® (Elekta AB, Stockholm, Sweden) and a dedicated planning system Monaco® (Elekta AB). The prescribed dose (PD) was uniformly assigned to just cover 98% of the GTV (D(98%)). Three VMARS plans with different dose inhomogeneities of the GTV were generated for each GTV: the % isodose surfaces (IDSs) of GTV D(98%), normalized to 100% at the maximum dose (D(max)), were ≤70% (extremely inhomogeneous dose, EIH); ≈80% (inhomogeneous dose, IH); and ≈90% (rather homogeneous dose, RH). VMARS plans were optimized using simple and similar cost functions. In particular, no dose constraint to the GTV D(max) was assigned to the EIH plans. Results Intended VMARS plans fulfilling the prerequisites were generated without problems for all GTVs of ≥10 mm, whereas 86.4% was the lowest IDS for the D(98%) for 5-mm GTV. Therefore, additional plans for 9- and 8-mm GTVs were generated, which resulted in 68.6% and 75.1% being the lowest IDSs for the D(98%) values of 9- and 8-mm GTVs, respectively. The EIH plans were the best in terms of the following: 1) dose conformity, i.e., minimum spillage of PD outside the GTV; 2) moderate, not too excessive, dose attenuation outside the GTV, i.e., appropriate marginal dose 2-mm outside the GTV boundary as a function of GTV size; and 3) lowest dose of the surrounding normal tissue outside the GTV. In contrast, the RH plans were the worst based on all of the aforementioned measures. Conclusions On the assumption of uniform dose assignment to the GTV margin, a very inhomogeneous GTV dose is basically the most suitable for SRS of BMs in terms of 1) superior dose conformity; 2) minimizing the dose of the surrounding normal tissue outside the GTV; and 3) moderate dose spillage margin outside the GTV with a tumor volume-dependent rational increase, i.e., appropriate dose of the common PTV boundary. The concentrically laminated steep dose increase inside the GTV boundary for the EIH plan may also be advantageous for achieving superior tumor response, although early and excessive GTV shrinkage caused by the EIH plan during mfSRS can lead to surrounding brain injury.
format Online
Article
Text
id pubmed-10043638
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-100436382023-03-29 An Extremely Inhomogeneous Gross Tumor Dose is Suitable for Volumetric Modulated Arc-Based Radiosurgery with a 5-mm Leaf-Width Multileaf Collimator for Single Brain Metastasis Ohtakara, Kazuhiro Suzuki, Kojiro Cureus Medical Physics Introduction Single or multi-fraction (mf) stereotactic radiosurgery (SRS) is an indispensable treatment option for brain metastases (BMs). The integration of volumetric modulated arc therapy (VMAT) into linac-based SRS is expected to further enhance efficacy and safety and to expand the indications for the challenging type of BMs. However, the optimal treatment design and relevant optimization method for volumetric modulated arc-based radiosurgery (VMARS) remain unestablished with substantial inter-institutional differences. Therefore, this study was conducted to determine the optimal dose distribution suitable for VMARS of BMs, especially regarding dose inhomogeneity of the gross tumor volume (GTV). The GTV boundary, not margin-added planning target volume, was regarded as a basis for planning optimization and dose prescription. Materials and methods This was a planning study for the clinical scenario of a single BM. Eight sphere-shaped objects with diameters of 5-40 mm in 5-mm increments were assumed as GTVs. The treatment system included a 5-mm leaf width multileaf collimator (MLC) Agility® (Elekta AB, Stockholm, Sweden) and a dedicated planning system Monaco® (Elekta AB). The prescribed dose (PD) was uniformly assigned to just cover 98% of the GTV (D(98%)). Three VMARS plans with different dose inhomogeneities of the GTV were generated for each GTV: the % isodose surfaces (IDSs) of GTV D(98%), normalized to 100% at the maximum dose (D(max)), were ≤70% (extremely inhomogeneous dose, EIH); ≈80% (inhomogeneous dose, IH); and ≈90% (rather homogeneous dose, RH). VMARS plans were optimized using simple and similar cost functions. In particular, no dose constraint to the GTV D(max) was assigned to the EIH plans. Results Intended VMARS plans fulfilling the prerequisites were generated without problems for all GTVs of ≥10 mm, whereas 86.4% was the lowest IDS for the D(98%) for 5-mm GTV. Therefore, additional plans for 9- and 8-mm GTVs were generated, which resulted in 68.6% and 75.1% being the lowest IDSs for the D(98%) values of 9- and 8-mm GTVs, respectively. The EIH plans were the best in terms of the following: 1) dose conformity, i.e., minimum spillage of PD outside the GTV; 2) moderate, not too excessive, dose attenuation outside the GTV, i.e., appropriate marginal dose 2-mm outside the GTV boundary as a function of GTV size; and 3) lowest dose of the surrounding normal tissue outside the GTV. In contrast, the RH plans were the worst based on all of the aforementioned measures. Conclusions On the assumption of uniform dose assignment to the GTV margin, a very inhomogeneous GTV dose is basically the most suitable for SRS of BMs in terms of 1) superior dose conformity; 2) minimizing the dose of the surrounding normal tissue outside the GTV; and 3) moderate dose spillage margin outside the GTV with a tumor volume-dependent rational increase, i.e., appropriate dose of the common PTV boundary. The concentrically laminated steep dose increase inside the GTV boundary for the EIH plan may also be advantageous for achieving superior tumor response, although early and excessive GTV shrinkage caused by the EIH plan during mfSRS can lead to surrounding brain injury. Cureus 2023-02-25 /pmc/articles/PMC10043638/ /pubmed/36999102 http://dx.doi.org/10.7759/cureus.35467 Text en Copyright © 2023, Ohtakara et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Medical Physics
Ohtakara, Kazuhiro
Suzuki, Kojiro
An Extremely Inhomogeneous Gross Tumor Dose is Suitable for Volumetric Modulated Arc-Based Radiosurgery with a 5-mm Leaf-Width Multileaf Collimator for Single Brain Metastasis
title An Extremely Inhomogeneous Gross Tumor Dose is Suitable for Volumetric Modulated Arc-Based Radiosurgery with a 5-mm Leaf-Width Multileaf Collimator for Single Brain Metastasis
title_full An Extremely Inhomogeneous Gross Tumor Dose is Suitable for Volumetric Modulated Arc-Based Radiosurgery with a 5-mm Leaf-Width Multileaf Collimator for Single Brain Metastasis
title_fullStr An Extremely Inhomogeneous Gross Tumor Dose is Suitable for Volumetric Modulated Arc-Based Radiosurgery with a 5-mm Leaf-Width Multileaf Collimator for Single Brain Metastasis
title_full_unstemmed An Extremely Inhomogeneous Gross Tumor Dose is Suitable for Volumetric Modulated Arc-Based Radiosurgery with a 5-mm Leaf-Width Multileaf Collimator for Single Brain Metastasis
title_short An Extremely Inhomogeneous Gross Tumor Dose is Suitable for Volumetric Modulated Arc-Based Radiosurgery with a 5-mm Leaf-Width Multileaf Collimator for Single Brain Metastasis
title_sort extremely inhomogeneous gross tumor dose is suitable for volumetric modulated arc-based radiosurgery with a 5-mm leaf-width multileaf collimator for single brain metastasis
topic Medical Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10043638/
https://www.ncbi.nlm.nih.gov/pubmed/36999102
http://dx.doi.org/10.7759/cureus.35467
work_keys_str_mv AT ohtakarakazuhiro anextremelyinhomogeneousgrosstumordoseissuitableforvolumetricmodulatedarcbasedradiosurgerywitha5mmleafwidthmultileafcollimatorforsinglebrainmetastasis
AT suzukikojiro anextremelyinhomogeneousgrosstumordoseissuitableforvolumetricmodulatedarcbasedradiosurgerywitha5mmleafwidthmultileafcollimatorforsinglebrainmetastasis
AT ohtakarakazuhiro extremelyinhomogeneousgrosstumordoseissuitableforvolumetricmodulatedarcbasedradiosurgerywitha5mmleafwidthmultileafcollimatorforsinglebrainmetastasis
AT suzukikojiro extremelyinhomogeneousgrosstumordoseissuitableforvolumetricmodulatedarcbasedradiosurgerywitha5mmleafwidthmultileafcollimatorforsinglebrainmetastasis