Validation of the International Consensus Group Criteria for Slide Review Following Automated Complete Blood Count at Jimma Medical Center, Ethiopia

BACKGROUND: The international consensus group suggested criteria for action following automated complete blood count and white blood cell differential analysis. These criteria were set based on data from laboratories of developed countries. It is highly important to validate the criteria in developi...

Descripción completa

Detalles Bibliográficos
Autores principales: Kiya, Girum Tesfaye, Mamo, Aklilu Getachew, Biya, Sintayehu Asaye, Gobena, Dejene Gebre, Demeke, Natal, Shenkutie, Tilahun Yemane
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10046196/
https://www.ncbi.nlm.nih.gov/pubmed/36998244
http://dx.doi.org/10.2147/JBM.S402940
_version_ 1785013608420212736
author Kiya, Girum Tesfaye
Mamo, Aklilu Getachew
Biya, Sintayehu Asaye
Gobena, Dejene Gebre
Demeke, Natal
Shenkutie, Tilahun Yemane
author_facet Kiya, Girum Tesfaye
Mamo, Aklilu Getachew
Biya, Sintayehu Asaye
Gobena, Dejene Gebre
Demeke, Natal
Shenkutie, Tilahun Yemane
author_sort Kiya, Girum Tesfaye
collection PubMed
description BACKGROUND: The international consensus group suggested criteria for action following automated complete blood count and white blood cell differential analysis. These criteria were set based on data from laboratories of developed countries. It is highly important to validate the criteria in developing countries where infectious diseases are still rampant that can affect blood cell count and morphology. Thus, this study aimed to validate the consensus group criteria for slide review at Jimma Medical Center, Ethiopia from November 1, 2020 to February 30, 2021. METHODS: The study comprised a total of 1685 patient samples from the daily laboratory workload of CBC analysis. The samples were collected in K2-EDTA tubes (Becton Dickinson) and analyzed using Coulter DxH 800 and Sysmex XT-1880 hematology analyzers. A slide review was done on two Wright-stained slides for each sample. All statistical analyses were performed using SPSS version 20 software. RESULTS: There were 39.8% positive findings, the majority of which were related to red blood cells. The false negative and false positive rates for Sysmex and Coulter analyzer were 2.4% vs 4.8%; and 4.6% vs 4.7%, respectively. The false negative rate was unacceptably higher when we used physicians’ triggered slide review, which was 17.3% and 17.9% for Sysmex and Coulter analyzers, respectively. CONCLUSION: Generally, the consensus group rules are suitable to use in our setting. However, we might still need to modify the rules, particularly to reduce the review rates. It is also necessary to confirm the rules with case mixes proportionally derived from the source population.
format Online
Article
Text
id pubmed-10046196
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Dove
record_format MEDLINE/PubMed
spelling pubmed-100461962023-03-29 Validation of the International Consensus Group Criteria for Slide Review Following Automated Complete Blood Count at Jimma Medical Center, Ethiopia Kiya, Girum Tesfaye Mamo, Aklilu Getachew Biya, Sintayehu Asaye Gobena, Dejene Gebre Demeke, Natal Shenkutie, Tilahun Yemane J Blood Med Original Research BACKGROUND: The international consensus group suggested criteria for action following automated complete blood count and white blood cell differential analysis. These criteria were set based on data from laboratories of developed countries. It is highly important to validate the criteria in developing countries where infectious diseases are still rampant that can affect blood cell count and morphology. Thus, this study aimed to validate the consensus group criteria for slide review at Jimma Medical Center, Ethiopia from November 1, 2020 to February 30, 2021. METHODS: The study comprised a total of 1685 patient samples from the daily laboratory workload of CBC analysis. The samples were collected in K2-EDTA tubes (Becton Dickinson) and analyzed using Coulter DxH 800 and Sysmex XT-1880 hematology analyzers. A slide review was done on two Wright-stained slides for each sample. All statistical analyses were performed using SPSS version 20 software. RESULTS: There were 39.8% positive findings, the majority of which were related to red blood cells. The false negative and false positive rates for Sysmex and Coulter analyzer were 2.4% vs 4.8%; and 4.6% vs 4.7%, respectively. The false negative rate was unacceptably higher when we used physicians’ triggered slide review, which was 17.3% and 17.9% for Sysmex and Coulter analyzers, respectively. CONCLUSION: Generally, the consensus group rules are suitable to use in our setting. However, we might still need to modify the rules, particularly to reduce the review rates. It is also necessary to confirm the rules with case mixes proportionally derived from the source population. Dove 2023-03-24 /pmc/articles/PMC10046196/ /pubmed/36998244 http://dx.doi.org/10.2147/JBM.S402940 Text en © 2023 Kiya et al. https://creativecommons.org/licenses/by-nc/3.0/This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/ (https://creativecommons.org/licenses/by-nc/3.0/) ). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php).
spellingShingle Original Research
Kiya, Girum Tesfaye
Mamo, Aklilu Getachew
Biya, Sintayehu Asaye
Gobena, Dejene Gebre
Demeke, Natal
Shenkutie, Tilahun Yemane
Validation of the International Consensus Group Criteria for Slide Review Following Automated Complete Blood Count at Jimma Medical Center, Ethiopia
title Validation of the International Consensus Group Criteria for Slide Review Following Automated Complete Blood Count at Jimma Medical Center, Ethiopia
title_full Validation of the International Consensus Group Criteria for Slide Review Following Automated Complete Blood Count at Jimma Medical Center, Ethiopia
title_fullStr Validation of the International Consensus Group Criteria for Slide Review Following Automated Complete Blood Count at Jimma Medical Center, Ethiopia
title_full_unstemmed Validation of the International Consensus Group Criteria for Slide Review Following Automated Complete Blood Count at Jimma Medical Center, Ethiopia
title_short Validation of the International Consensus Group Criteria for Slide Review Following Automated Complete Blood Count at Jimma Medical Center, Ethiopia
title_sort validation of the international consensus group criteria for slide review following automated complete blood count at jimma medical center, ethiopia
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10046196/
https://www.ncbi.nlm.nih.gov/pubmed/36998244
http://dx.doi.org/10.2147/JBM.S402940
work_keys_str_mv AT kiyagirumtesfaye validationoftheinternationalconsensusgroupcriteriaforslidereviewfollowingautomatedcompletebloodcountatjimmamedicalcenterethiopia
AT mamoaklilugetachew validationoftheinternationalconsensusgroupcriteriaforslidereviewfollowingautomatedcompletebloodcountatjimmamedicalcenterethiopia
AT biyasintayehuasaye validationoftheinternationalconsensusgroupcriteriaforslidereviewfollowingautomatedcompletebloodcountatjimmamedicalcenterethiopia
AT gobenadejenegebre validationoftheinternationalconsensusgroupcriteriaforslidereviewfollowingautomatedcompletebloodcountatjimmamedicalcenterethiopia
AT demekenatal validationoftheinternationalconsensusgroupcriteriaforslidereviewfollowingautomatedcompletebloodcountatjimmamedicalcenterethiopia
AT shenkutietilahunyemane validationoftheinternationalconsensusgroupcriteriaforslidereviewfollowingautomatedcompletebloodcountatjimmamedicalcenterethiopia