Cargando…

Liquid chromatography and differential mobility spectrometry—data-independent mass spectrometry for comprehensive multidimensional separations in metabolomics

The benefits of combining drift time ion mobility (DTIMS) with liquid chromatography–high-resolution mass spectrometry (HRMS) have been reported for metabolomics but the use of differential time mobility spectrometry (DMS) is less obvious due to the need for rapid scanning of the DMS cell. Drift DTI...

Descripción completa

Detalles Bibliográficos
Autores principales: Ekmekciu, Lysi, Hopfgartner, Gérard
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10050028/
https://www.ncbi.nlm.nih.gov/pubmed/36820908
http://dx.doi.org/10.1007/s00216-023-04602-0
Descripción
Sumario:The benefits of combining drift time ion mobility (DTIMS) with liquid chromatography–high-resolution mass spectrometry (HRMS) have been reported for metabolomics but the use of differential time mobility spectrometry (DMS) is less obvious due to the need for rapid scanning of the DMS cell. Drift DTIMS provides additional precursor ion selectivity and collisional cross-section information but the separation resolution between analytes remains cell- and component-dependent. With DMS, the addition of 2-propanol modifier can improve the selectivity but on cost of analyte MS response. In the present work, we investigate the liquid chromatography–mass spectrometry (LC–MS) analysis of a mix of 50 analytes, representative for urine and plasma metabolites, using scanning DMS with the single modifiers cyclohexane (Ch), toluene (Tol), acetonitrile (ACN), ethanol (EtOH), and 2-propanol (IPA), and a binary modifier mixture (cyclohexane/2-propanol) with emphasis on selectivity and signal sensitivity. 1.5% IPA in the N(2) stream was found to suppress the signal of 50% of the analytes which could be partially recovered with the use of IPA to 0.05% as a Ch/IPA mixture. The potential to use the separation voltage/compensation voltage/modifier (SV/CoV/Mod) feature as an additional analyte identifier for qualitative analysis is also presented and applied to a data-independent LCxDMS-SWATH-MS workflow for the analysis of endogenous metabolites and drugs of abuse in human urine samples from traffic control. GRAPHICAL ABSTRACT: [Image: see text] SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00216-023-04602-0.