Cargando…
Application of Multi-Criteria Optimization Methods in the Calibration Process of Digital Measuring Instruments
The article describes the use of multi-criteria optimization methods during the calibration of digital multimeters. Currently, calibration is based on a single measurement of a specific value. The aim of this research was to confirm the possibility of using a series of measurements in order to reduc...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10052159/ https://www.ncbi.nlm.nih.gov/pubmed/36991697 http://dx.doi.org/10.3390/s23062984 |
_version_ | 1785015093483798528 |
---|---|
author | Klebba, Maciej Adamczyk, Arkadiusz Wąż, Mariusz Iwen, Dominik |
author_facet | Klebba, Maciej Adamczyk, Arkadiusz Wąż, Mariusz Iwen, Dominik |
author_sort | Klebba, Maciej |
collection | PubMed |
description | The article describes the use of multi-criteria optimization methods during the calibration of digital multimeters. Currently, calibration is based on a single measurement of a specific value. The aim of this research was to confirm the possibility of using a series of measurements in order to reduce the measurement uncertainty without significantly extending the calibration time. The automatic measurement loading laboratory stand used during the conducted experiments was essential to obtain results that allowed confirming the thesis. This article presents the applied optimization methods and the results of the calibration of sample digital multimeters obtained thanks to them. As a result of the research, it was found that the use of a series of measurements increased the accuracy of the calibration, reduced the measurement uncertainty, and shortened the calibration time compared to traditional methods. |
format | Online Article Text |
id | pubmed-10052159 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-100521592023-03-30 Application of Multi-Criteria Optimization Methods in the Calibration Process of Digital Measuring Instruments Klebba, Maciej Adamczyk, Arkadiusz Wąż, Mariusz Iwen, Dominik Sensors (Basel) Article The article describes the use of multi-criteria optimization methods during the calibration of digital multimeters. Currently, calibration is based on a single measurement of a specific value. The aim of this research was to confirm the possibility of using a series of measurements in order to reduce the measurement uncertainty without significantly extending the calibration time. The automatic measurement loading laboratory stand used during the conducted experiments was essential to obtain results that allowed confirming the thesis. This article presents the applied optimization methods and the results of the calibration of sample digital multimeters obtained thanks to them. As a result of the research, it was found that the use of a series of measurements increased the accuracy of the calibration, reduced the measurement uncertainty, and shortened the calibration time compared to traditional methods. MDPI 2023-03-09 /pmc/articles/PMC10052159/ /pubmed/36991697 http://dx.doi.org/10.3390/s23062984 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Klebba, Maciej Adamczyk, Arkadiusz Wąż, Mariusz Iwen, Dominik Application of Multi-Criteria Optimization Methods in the Calibration Process of Digital Measuring Instruments |
title | Application of Multi-Criteria Optimization Methods in the Calibration Process of Digital Measuring Instruments |
title_full | Application of Multi-Criteria Optimization Methods in the Calibration Process of Digital Measuring Instruments |
title_fullStr | Application of Multi-Criteria Optimization Methods in the Calibration Process of Digital Measuring Instruments |
title_full_unstemmed | Application of Multi-Criteria Optimization Methods in the Calibration Process of Digital Measuring Instruments |
title_short | Application of Multi-Criteria Optimization Methods in the Calibration Process of Digital Measuring Instruments |
title_sort | application of multi-criteria optimization methods in the calibration process of digital measuring instruments |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10052159/ https://www.ncbi.nlm.nih.gov/pubmed/36991697 http://dx.doi.org/10.3390/s23062984 |
work_keys_str_mv | AT klebbamaciej applicationofmulticriteriaoptimizationmethodsinthecalibrationprocessofdigitalmeasuringinstruments AT adamczykarkadiusz applicationofmulticriteriaoptimizationmethodsinthecalibrationprocessofdigitalmeasuringinstruments AT wazmariusz applicationofmulticriteriaoptimizationmethodsinthecalibrationprocessofdigitalmeasuringinstruments AT iwendominik applicationofmulticriteriaoptimizationmethodsinthecalibrationprocessofdigitalmeasuringinstruments |