In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating
Background and Objectives: Fractures of the proximal humerus are common, particularly in elderly populations. Anatomical locking plates target stabilization with a multitude of screws spanning into the humeral head. Sound implant placement and screw length determination are key for a successful clin...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10055017/ https://www.ncbi.nlm.nih.gov/pubmed/36984451 http://dx.doi.org/10.3390/medicina59030450 |
_version_ | 1785015805210001408 |
---|---|
author | Windolf, Markus Knierzinger, Dominik Nijs, Stefaan Sermon, An Blauth, Michael Richards, Robert Geoff Buschbaum, Jan |
author_facet | Windolf, Markus Knierzinger, Dominik Nijs, Stefaan Sermon, An Blauth, Michael Richards, Robert Geoff Buschbaum, Jan |
author_sort | Windolf, Markus |
collection | PubMed |
description | Background and Objectives: Fractures of the proximal humerus are common, particularly in elderly populations. Anatomical locking plates target stabilization with a multitude of screws spanning into the humeral head. Sound implant placement and screw length determination are key for a successful clinical outcome but are difficult to obtain from planar X-rays. A novel implant positioning technology for proximal humerus plating (Xin1) outputs screw lengths suggestions and plate position based on hole projections in conventional X-ray images. This study investigated the performance of a prototype Xin1 system in a postmortem (in vitro) experiment as well as in a clinical handling test. Materials and Methods: For in vitro testing, twelve shoulders from six anatomical specimens were randomized into two groups to compare the Xin1 technique to the conventional operation in terms of surgical precision, procedure time and X-ray exposure. For the clinical trial, 11 patients undergoing plating of the proximal humerus were included. The aim was to investigate clinical handling of the Xin1 marker clip and to retrospectively evaluate the system performance in a real-life fracture situation. Image pairs before and after insertion of the proximal screws were retrospectively processed to investigate the influence of potential bone fragment shifts on the system output. Results: In the postmortem experiment, the use of the system significantly improved the surgical precision (52% error reduction), procedure time (38% shorter) and radiation exposure (64% less X-rays). Clinical handling demonstrated seamless embedding of the marker clip into existing clinical workflows without adverse events reported. Retrospective X-ray analysis on six eligible patients revealed differences in the calculated screw lengths of ≤2 mm before and after screw insertion for five patients. In one patient, the screw lengths differed up to 8 mm, which might indicate displacement of the head fragment. Conclusions: Results suggest a strong potential of the Xin1 assistance technology to enhance the surgical procedure and patient outcomes in the rising incidence of osteoporotic humeral fractures. Robust performance in a real-life fracture situation was observed. In-depth validation of the system is, however, needed before placing it into clinical practice. |
format | Online Article Text |
id | pubmed-10055017 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-100550172023-03-30 In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating Windolf, Markus Knierzinger, Dominik Nijs, Stefaan Sermon, An Blauth, Michael Richards, Robert Geoff Buschbaum, Jan Medicina (Kaunas) Article Background and Objectives: Fractures of the proximal humerus are common, particularly in elderly populations. Anatomical locking plates target stabilization with a multitude of screws spanning into the humeral head. Sound implant placement and screw length determination are key for a successful clinical outcome but are difficult to obtain from planar X-rays. A novel implant positioning technology for proximal humerus plating (Xin1) outputs screw lengths suggestions and plate position based on hole projections in conventional X-ray images. This study investigated the performance of a prototype Xin1 system in a postmortem (in vitro) experiment as well as in a clinical handling test. Materials and Methods: For in vitro testing, twelve shoulders from six anatomical specimens were randomized into two groups to compare the Xin1 technique to the conventional operation in terms of surgical precision, procedure time and X-ray exposure. For the clinical trial, 11 patients undergoing plating of the proximal humerus were included. The aim was to investigate clinical handling of the Xin1 marker clip and to retrospectively evaluate the system performance in a real-life fracture situation. Image pairs before and after insertion of the proximal screws were retrospectively processed to investigate the influence of potential bone fragment shifts on the system output. Results: In the postmortem experiment, the use of the system significantly improved the surgical precision (52% error reduction), procedure time (38% shorter) and radiation exposure (64% less X-rays). Clinical handling demonstrated seamless embedding of the marker clip into existing clinical workflows without adverse events reported. Retrospective X-ray analysis on six eligible patients revealed differences in the calculated screw lengths of ≤2 mm before and after screw insertion for five patients. In one patient, the screw lengths differed up to 8 mm, which might indicate displacement of the head fragment. Conclusions: Results suggest a strong potential of the Xin1 assistance technology to enhance the surgical procedure and patient outcomes in the rising incidence of osteoporotic humeral fractures. Robust performance in a real-life fracture situation was observed. In-depth validation of the system is, however, needed before placing it into clinical practice. MDPI 2023-02-23 /pmc/articles/PMC10055017/ /pubmed/36984451 http://dx.doi.org/10.3390/medicina59030450 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Windolf, Markus Knierzinger, Dominik Nijs, Stefaan Sermon, An Blauth, Michael Richards, Robert Geoff Buschbaum, Jan In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating |
title | In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating |
title_full | In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating |
title_fullStr | In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating |
title_full_unstemmed | In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating |
title_short | In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating |
title_sort | in vitro testing and clinical handling of a novel implant positioning technology for proximal humeral plating |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10055017/ https://www.ncbi.nlm.nih.gov/pubmed/36984451 http://dx.doi.org/10.3390/medicina59030450 |
work_keys_str_mv | AT windolfmarkus invitrotestingandclinicalhandlingofanovelimplantpositioningtechnologyforproximalhumeralplating AT knierzingerdominik invitrotestingandclinicalhandlingofanovelimplantpositioningtechnologyforproximalhumeralplating AT nijsstefaan invitrotestingandclinicalhandlingofanovelimplantpositioningtechnologyforproximalhumeralplating AT sermonan invitrotestingandclinicalhandlingofanovelimplantpositioningtechnologyforproximalhumeralplating AT blauthmichael invitrotestingandclinicalhandlingofanovelimplantpositioningtechnologyforproximalhumeralplating AT richardsrobertgeoff invitrotestingandclinicalhandlingofanovelimplantpositioningtechnologyforproximalhumeralplating AT buschbaumjan invitrotestingandclinicalhandlingofanovelimplantpositioningtechnologyforproximalhumeralplating |