Accuracy Validation of the Elecsys HBsAg II Quant Assay and Its Utility in Resolving Equivocal Qualitative HBsAg Results

Background and Objectives: There are reports of false qualitative HBsAg results, because of various causes, such as samples with low HBsAg concentrations that may produce false positives. The main aims of this study were to validate the analytical accuracy and to assess the utility of the Elecsys as...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Jaehyeon, Lee, Seung Yeob, Cho, Yong Gon, Kim, Dal Sik, Park, Joonhong
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10056079/
https://www.ncbi.nlm.nih.gov/pubmed/36984443
http://dx.doi.org/10.3390/medicina59030443
Descripción
Sumario:Background and Objectives: There are reports of false qualitative HBsAg results, because of various causes, such as samples with low HBsAg concentrations that may produce false positives. The main aims of this study were to validate the analytical accuracy and to assess the utility of the Elecsys assay compared to that of the qualitative HbsAg assay as a screening test in resolving equivocal qualitative HbsAg results. Materials and Methods: The limit of blank (LoB), the limit of detection (LoD), the limit of quantification (LoQ), and linearity were estimated to validate the analytical accuracy of the Elecsys HBsAg II Quant assay. A total of 449 serum samples showing initial equivocal results (1–50 index) were evaluated by Elecsys HBsAg II Quant and ADVIA Centaur HBsAg II assays. Results: The LoQ of the assay was determined to be 0.050 IU/mL, as provided by the manufacturer. The Kappa agreement between the two assays was almost perfect, at 0.9669, despite seven discordant results. With a specificity of 100% at new cut-off index value ≥5.42, about 78 samples (17%, 78/449) with index value ≥5.42 were interpreted as positives without further duplicate tests, however the remaining 371 samples with index value <5.42 need to be confirmed with additional HBV marker assays. Conclusions: We confirm that the Elecsys HBsAg II Quant assay is accurate and sensitive for HBV infection and recommend it as an alternative confirmatory HBsAg assay for resolving equivocal qualitative HBsAg results.