Accuracy Validation of the Elecsys HBsAg II Quant Assay and Its Utility in Resolving Equivocal Qualitative HBsAg Results
Background and Objectives: There are reports of false qualitative HBsAg results, because of various causes, such as samples with low HBsAg concentrations that may produce false positives. The main aims of this study were to validate the analytical accuracy and to assess the utility of the Elecsys as...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10056079/ https://www.ncbi.nlm.nih.gov/pubmed/36984443 http://dx.doi.org/10.3390/medicina59030443 |
_version_ | 1785016038065176576 |
---|---|
author | Lee, Jaehyeon Lee, Seung Yeob Cho, Yong Gon Kim, Dal Sik Park, Joonhong |
author_facet | Lee, Jaehyeon Lee, Seung Yeob Cho, Yong Gon Kim, Dal Sik Park, Joonhong |
author_sort | Lee, Jaehyeon |
collection | PubMed |
description | Background and Objectives: There are reports of false qualitative HBsAg results, because of various causes, such as samples with low HBsAg concentrations that may produce false positives. The main aims of this study were to validate the analytical accuracy and to assess the utility of the Elecsys assay compared to that of the qualitative HbsAg assay as a screening test in resolving equivocal qualitative HbsAg results. Materials and Methods: The limit of blank (LoB), the limit of detection (LoD), the limit of quantification (LoQ), and linearity were estimated to validate the analytical accuracy of the Elecsys HBsAg II Quant assay. A total of 449 serum samples showing initial equivocal results (1–50 index) were evaluated by Elecsys HBsAg II Quant and ADVIA Centaur HBsAg II assays. Results: The LoQ of the assay was determined to be 0.050 IU/mL, as provided by the manufacturer. The Kappa agreement between the two assays was almost perfect, at 0.9669, despite seven discordant results. With a specificity of 100% at new cut-off index value ≥5.42, about 78 samples (17%, 78/449) with index value ≥5.42 were interpreted as positives without further duplicate tests, however the remaining 371 samples with index value <5.42 need to be confirmed with additional HBV marker assays. Conclusions: We confirm that the Elecsys HBsAg II Quant assay is accurate and sensitive for HBV infection and recommend it as an alternative confirmatory HBsAg assay for resolving equivocal qualitative HBsAg results. |
format | Online Article Text |
id | pubmed-10056079 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-100560792023-03-30 Accuracy Validation of the Elecsys HBsAg II Quant Assay and Its Utility in Resolving Equivocal Qualitative HBsAg Results Lee, Jaehyeon Lee, Seung Yeob Cho, Yong Gon Kim, Dal Sik Park, Joonhong Medicina (Kaunas) Article Background and Objectives: There are reports of false qualitative HBsAg results, because of various causes, such as samples with low HBsAg concentrations that may produce false positives. The main aims of this study were to validate the analytical accuracy and to assess the utility of the Elecsys assay compared to that of the qualitative HbsAg assay as a screening test in resolving equivocal qualitative HbsAg results. Materials and Methods: The limit of blank (LoB), the limit of detection (LoD), the limit of quantification (LoQ), and linearity were estimated to validate the analytical accuracy of the Elecsys HBsAg II Quant assay. A total of 449 serum samples showing initial equivocal results (1–50 index) were evaluated by Elecsys HBsAg II Quant and ADVIA Centaur HBsAg II assays. Results: The LoQ of the assay was determined to be 0.050 IU/mL, as provided by the manufacturer. The Kappa agreement between the two assays was almost perfect, at 0.9669, despite seven discordant results. With a specificity of 100% at new cut-off index value ≥5.42, about 78 samples (17%, 78/449) with index value ≥5.42 were interpreted as positives without further duplicate tests, however the remaining 371 samples with index value <5.42 need to be confirmed with additional HBV marker assays. Conclusions: We confirm that the Elecsys HBsAg II Quant assay is accurate and sensitive for HBV infection and recommend it as an alternative confirmatory HBsAg assay for resolving equivocal qualitative HBsAg results. MDPI 2023-02-23 /pmc/articles/PMC10056079/ /pubmed/36984443 http://dx.doi.org/10.3390/medicina59030443 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Lee, Jaehyeon Lee, Seung Yeob Cho, Yong Gon Kim, Dal Sik Park, Joonhong Accuracy Validation of the Elecsys HBsAg II Quant Assay and Its Utility in Resolving Equivocal Qualitative HBsAg Results |
title | Accuracy Validation of the Elecsys HBsAg II Quant Assay and Its Utility in Resolving Equivocal Qualitative HBsAg Results |
title_full | Accuracy Validation of the Elecsys HBsAg II Quant Assay and Its Utility in Resolving Equivocal Qualitative HBsAg Results |
title_fullStr | Accuracy Validation of the Elecsys HBsAg II Quant Assay and Its Utility in Resolving Equivocal Qualitative HBsAg Results |
title_full_unstemmed | Accuracy Validation of the Elecsys HBsAg II Quant Assay and Its Utility in Resolving Equivocal Qualitative HBsAg Results |
title_short | Accuracy Validation of the Elecsys HBsAg II Quant Assay and Its Utility in Resolving Equivocal Qualitative HBsAg Results |
title_sort | accuracy validation of the elecsys hbsag ii quant assay and its utility in resolving equivocal qualitative hbsag results |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10056079/ https://www.ncbi.nlm.nih.gov/pubmed/36984443 http://dx.doi.org/10.3390/medicina59030443 |
work_keys_str_mv | AT leejaehyeon accuracyvalidationoftheelecsyshbsagiiquantassayanditsutilityinresolvingequivocalqualitativehbsagresults AT leeseungyeob accuracyvalidationoftheelecsyshbsagiiquantassayanditsutilityinresolvingequivocalqualitativehbsagresults AT choyonggon accuracyvalidationoftheelecsyshbsagiiquantassayanditsutilityinresolvingequivocalqualitativehbsagresults AT kimdalsik accuracyvalidationoftheelecsyshbsagiiquantassayanditsutilityinresolvingequivocalqualitativehbsagresults AT parkjoonhong accuracyvalidationoftheelecsyshbsagiiquantassayanditsutilityinresolvingequivocalqualitativehbsagresults |