Agreement between cardiac output estimation by multi-beat analysis of arterial blood pressure waveforms and continuous thermodilution in post cardiac surgery intensive care unit patients
We sought to assess agreement of cardiac output estimation between continuous pulmonary artery catheter (PAC) guided thermodilution (CO-CTD) and a novel pulse wave analysis (PWA) method that performs an analysis of multiple beats of the arterial blood pressure waveform (CO-MBA) in post-operative car...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Netherlands
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10068656/ https://www.ncbi.nlm.nih.gov/pubmed/36269451 http://dx.doi.org/10.1007/s10877-022-00924-z |
Sumario: | We sought to assess agreement of cardiac output estimation between continuous pulmonary artery catheter (PAC) guided thermodilution (CO-CTD) and a novel pulse wave analysis (PWA) method that performs an analysis of multiple beats of the arterial blood pressure waveform (CO-MBA) in post-operative cardiac surgery patients. PAC obtained CO-CTD measurements were compared with CO-MBA measurements from the Argos monitor (Retia Medical; Valhalla, NY, USA), in prospectively enrolled adult cardiac surgical intensive care unit patients. Agreement was assessed via Bland-Altman analysis. Subgroup analysis was performed on data segments identified as arrhythmia, or with low CO (less than 5 L/min). 927 hours of monitoring data from 79 patients was analyzed, of which 26 had arrhythmia. Mean CO-CTD was 5.29 ± 1.14 L/min (bias ± precision), whereas mean CO-MBA was 5.36 ± 1.33 L/min, (4.95 ± 0.80 L/min and 5.04 ± 1.07 L/min in the arrhythmia subgroup). Mean of differences was 0.04 ± 1.04 L/min with an error of 38.2%. In the arrhythmia subgroup, mean of differences was 0.14 ± 0.90 L/min with an error of 35.4%. In the low CO subgroup, mean of differences was 0.26 ± 0.89 L/min with an error of 40.4%. In adult patients after cardiac surgery, including those with low cardiac output and arrhythmia CO-MBA is not interchangeable with the continuous thermodilution method via a PAC, when using a 30% error threshold. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s10877-022-00924-z. |
---|