Paper 21: All-Inside Soft Suture Anchor Meniscus Repair Devices vs. PEEK Anchor Devices: A Biomechanical Analysis
OBJECTIVES: To evaluate the load to failure and meniscus gapping of longitudinal meniscus repairs with Soft Suture Based all-inside devices and compare them to a polyether ether ketone (PEEK) anchored device and inside-out repair. METHODS: After IRB review 12 human cadaver knees were dissected leavi...