Pyloric Functional Lumen Imaging Probe Measurements Are Dependent on Balloon Position

BACKGROUND/AIMS: The functional lumen imaging probe (FLIP) device has been used to assess pyloric dysfunction in patients with gastroparesis. We aim to investigate whether varying FLIP catheter positions affect pyloric FLIP measurements. METHODS: Patients undergoing endoscopy for chronic unexplained...

Descripción completa

Detalles Bibliográficos
Autores principales: Yim, Brandon, Gregor, Lennon, Siwiec, Robert M, Al-Haddad, Mohammad, Nowak, Thomas V, Wo, John M
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Society of Neurogastroenterology and Motility 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10083104/
https://www.ncbi.nlm.nih.gov/pubmed/37019864
http://dx.doi.org/10.5056/jnm22053
_version_ 1785021436195241984
author Yim, Brandon
Gregor, Lennon
Siwiec, Robert M
Al-Haddad, Mohammad
Nowak, Thomas V
Wo, John M
author_facet Yim, Brandon
Gregor, Lennon
Siwiec, Robert M
Al-Haddad, Mohammad
Nowak, Thomas V
Wo, John M
author_sort Yim, Brandon
collection PubMed
description BACKGROUND/AIMS: The functional lumen imaging probe (FLIP) device has been used to assess pyloric dysfunction in patients with gastroparesis. We aim to investigate whether varying FLIP catheter positions affect pyloric FLIP measurements. METHODS: Patients undergoing endoscopy for chronic unexplained nausea and vomiting (CUNV) or gastroparesis were prospectively enrolled. FLIP balloon was adjusted for 3 positions within the pylorus: (1) proximal position, 75% of FLIP balloon in the duodenum and 25% in the antrum; (2) middle position, 50% in the duodenum and 50% in the antrum; and (3) distal position, 25% in the duodenum and 75% in the antrum. Pylorus cross-sectional area (CSA), intra-bag pressure (P), and distensibility indices (DI) were measured for 30, 40, and 50-mL balloon volumes. Fluoroscopic images were obtained to confirm FLIP balloon geometry. Data was analyzed separately using FLIP Analytic and customized MATLAB software. RESULTS: Twenty-two patients with CUNV (n = 4) and gastroparesis (n = 18) were enrolled. Pressures were significantly higher in the proximal position compared to the middle and distal positions. CSA measurements were significantly higher at the proximal and middle positions for 30-mL and 40-mL volume compared to the distal position values. DI values were significantly lower at the proximal positions for 40-mL and 50-mL distensions when compared to the middle and distal positions. Fluoroscopic images confirmed increased balloon bending when placed mostly in the duodenum. CONCLUSIONS: FLIP balloon position within the pylorus directly affects balloon geometry which significantly affects P, CSA, and DI measurements. Standardized pyloric FLIP protocols and balloon design adjustments are needed for the continued application of this technology to the pylorus.
format Online
Article
Text
id pubmed-10083104
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher The Korean Society of Neurogastroenterology and Motility
record_format MEDLINE/PubMed
spelling pubmed-100831042023-04-30 Pyloric Functional Lumen Imaging Probe Measurements Are Dependent on Balloon Position Yim, Brandon Gregor, Lennon Siwiec, Robert M Al-Haddad, Mohammad Nowak, Thomas V Wo, John M J Neurogastroenterol Motil Original Article BACKGROUND/AIMS: The functional lumen imaging probe (FLIP) device has been used to assess pyloric dysfunction in patients with gastroparesis. We aim to investigate whether varying FLIP catheter positions affect pyloric FLIP measurements. METHODS: Patients undergoing endoscopy for chronic unexplained nausea and vomiting (CUNV) or gastroparesis were prospectively enrolled. FLIP balloon was adjusted for 3 positions within the pylorus: (1) proximal position, 75% of FLIP balloon in the duodenum and 25% in the antrum; (2) middle position, 50% in the duodenum and 50% in the antrum; and (3) distal position, 25% in the duodenum and 75% in the antrum. Pylorus cross-sectional area (CSA), intra-bag pressure (P), and distensibility indices (DI) were measured for 30, 40, and 50-mL balloon volumes. Fluoroscopic images were obtained to confirm FLIP balloon geometry. Data was analyzed separately using FLIP Analytic and customized MATLAB software. RESULTS: Twenty-two patients with CUNV (n = 4) and gastroparesis (n = 18) were enrolled. Pressures were significantly higher in the proximal position compared to the middle and distal positions. CSA measurements were significantly higher at the proximal and middle positions for 30-mL and 40-mL volume compared to the distal position values. DI values were significantly lower at the proximal positions for 40-mL and 50-mL distensions when compared to the middle and distal positions. Fluoroscopic images confirmed increased balloon bending when placed mostly in the duodenum. CONCLUSIONS: FLIP balloon position within the pylorus directly affects balloon geometry which significantly affects P, CSA, and DI measurements. Standardized pyloric FLIP protocols and balloon design adjustments are needed for the continued application of this technology to the pylorus. The Korean Society of Neurogastroenterology and Motility 2023-04-30 2023-04-30 /pmc/articles/PMC10083104/ /pubmed/37019864 http://dx.doi.org/10.5056/jnm22053 Text en © 2023 The Korean Society of Neurogastroenterology and Motility https://creativecommons.org/licenses/by-nc/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0 (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Yim, Brandon
Gregor, Lennon
Siwiec, Robert M
Al-Haddad, Mohammad
Nowak, Thomas V
Wo, John M
Pyloric Functional Lumen Imaging Probe Measurements Are Dependent on Balloon Position
title Pyloric Functional Lumen Imaging Probe Measurements Are Dependent on Balloon Position
title_full Pyloric Functional Lumen Imaging Probe Measurements Are Dependent on Balloon Position
title_fullStr Pyloric Functional Lumen Imaging Probe Measurements Are Dependent on Balloon Position
title_full_unstemmed Pyloric Functional Lumen Imaging Probe Measurements Are Dependent on Balloon Position
title_short Pyloric Functional Lumen Imaging Probe Measurements Are Dependent on Balloon Position
title_sort pyloric functional lumen imaging probe measurements are dependent on balloon position
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10083104/
https://www.ncbi.nlm.nih.gov/pubmed/37019864
http://dx.doi.org/10.5056/jnm22053
work_keys_str_mv AT yimbrandon pyloricfunctionallumenimagingprobemeasurementsaredependentonballoonposition
AT gregorlennon pyloricfunctionallumenimagingprobemeasurementsaredependentonballoonposition
AT siwiecrobertm pyloricfunctionallumenimagingprobemeasurementsaredependentonballoonposition
AT alhaddadmohammad pyloricfunctionallumenimagingprobemeasurementsaredependentonballoonposition
AT nowakthomasv pyloricfunctionallumenimagingprobemeasurementsaredependentonballoonposition
AT wojohnm pyloricfunctionallumenimagingprobemeasurementsaredependentonballoonposition