Assessment of subjective image quality, contrast to noise ratio and modulation transfer function in the middle ear using a novel full body cone beam computed tomography device

BACKGROUND: Multi slice computed tomography (MSCT) is the most common used method in middle ear imaging. However, MSCT lacks the ability to distinguish the ossicular chain microstructures in detail resulting in poorer diagnostic outcomes. Novel cone beam computed tomography (CBCT) devices’ image res...

Descripción completa

Detalles Bibliográficos
Autores principales: Heikkinen, Anssi-Kalle, Rissanen, Valtteri, Aarnisalo, Antti A., Nyman, Kristofer, Sinkkonen, Saku T., Koivisto, Juha
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10084678/
https://www.ncbi.nlm.nih.gov/pubmed/37038130
http://dx.doi.org/10.1186/s12880-023-00996-6
_version_ 1785021790152556544
author Heikkinen, Anssi-Kalle
Rissanen, Valtteri
Aarnisalo, Antti A.
Nyman, Kristofer
Sinkkonen, Saku T.
Koivisto, Juha
author_facet Heikkinen, Anssi-Kalle
Rissanen, Valtteri
Aarnisalo, Antti A.
Nyman, Kristofer
Sinkkonen, Saku T.
Koivisto, Juha
author_sort Heikkinen, Anssi-Kalle
collection PubMed
description BACKGROUND: Multi slice computed tomography (MSCT) is the most common used method in middle ear imaging. However, MSCT lacks the ability to distinguish the ossicular chain microstructures in detail resulting in poorer diagnostic outcomes. Novel cone beam computed tomography (CBCT) devices’ image resolution is, on the other hand, better than MSCT resolution. The aim of this study was to optimize imaging parameters of a novel full body CBCT device to obtain optimal contrast to noise ratio (CNR) with low effective dose, and to optimize its clinical usability. METHODS: Imaging of five anonymous excised human cadaver temporal bones, the acquisition of the effective doses and the CNR measurements were performed for images acquired on using Planmed XFI® full body CBCT device (Planmed Oy, Helsinki, Finland) with a voxel size of 75 µm. All images acquired from the specimens using 10 different imaging protocols varying from their tube current exposure time product (mAs) and tube voltage (kVp) were analyzed for eight anatomical landmarks and evaluated by three evaluators. RESULTS: With the exception of protocol with 90 kVp 100 mAs, all other protocols used are competent to image the finest structures. With a moderate effective dose (86.5 µSv), protocol with 90 kV 450 mAs was chosen the best protocol used in this study. A significant correlation between CNR and clinical image quality of the protocols was observed in linear regression model. Using the optimized imaging parameters, we were able to distinguish even the most delicate middle ear structures in 2D images and produce accurate 3D reconstructions. CONCLUSIONS: In this ex vivo experiment, the new Planmed XFI® full body CBCT device produced excellent 2D resolution and easily created 3D reconstructions in middle ear imaging with moderate effective doses. This device would be suitable for middle ear diagnostics and for e.g., preoperative planning. Furthermore, the results of this study can be used to optimize the effective dose by selecting appropriate exposure parameters depending on the diagnostic task.
format Online
Article
Text
id pubmed-10084678
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-100846782023-04-11 Assessment of subjective image quality, contrast to noise ratio and modulation transfer function in the middle ear using a novel full body cone beam computed tomography device Heikkinen, Anssi-Kalle Rissanen, Valtteri Aarnisalo, Antti A. Nyman, Kristofer Sinkkonen, Saku T. Koivisto, Juha BMC Med Imaging Research BACKGROUND: Multi slice computed tomography (MSCT) is the most common used method in middle ear imaging. However, MSCT lacks the ability to distinguish the ossicular chain microstructures in detail resulting in poorer diagnostic outcomes. Novel cone beam computed tomography (CBCT) devices’ image resolution is, on the other hand, better than MSCT resolution. The aim of this study was to optimize imaging parameters of a novel full body CBCT device to obtain optimal contrast to noise ratio (CNR) with low effective dose, and to optimize its clinical usability. METHODS: Imaging of five anonymous excised human cadaver temporal bones, the acquisition of the effective doses and the CNR measurements were performed for images acquired on using Planmed XFI® full body CBCT device (Planmed Oy, Helsinki, Finland) with a voxel size of 75 µm. All images acquired from the specimens using 10 different imaging protocols varying from their tube current exposure time product (mAs) and tube voltage (kVp) were analyzed for eight anatomical landmarks and evaluated by three evaluators. RESULTS: With the exception of protocol with 90 kVp 100 mAs, all other protocols used are competent to image the finest structures. With a moderate effective dose (86.5 µSv), protocol with 90 kV 450 mAs was chosen the best protocol used in this study. A significant correlation between CNR and clinical image quality of the protocols was observed in linear regression model. Using the optimized imaging parameters, we were able to distinguish even the most delicate middle ear structures in 2D images and produce accurate 3D reconstructions. CONCLUSIONS: In this ex vivo experiment, the new Planmed XFI® full body CBCT device produced excellent 2D resolution and easily created 3D reconstructions in middle ear imaging with moderate effective doses. This device would be suitable for middle ear diagnostics and for e.g., preoperative planning. Furthermore, the results of this study can be used to optimize the effective dose by selecting appropriate exposure parameters depending on the diagnostic task. BioMed Central 2023-04-10 /pmc/articles/PMC10084678/ /pubmed/37038130 http://dx.doi.org/10.1186/s12880-023-00996-6 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Heikkinen, Anssi-Kalle
Rissanen, Valtteri
Aarnisalo, Antti A.
Nyman, Kristofer
Sinkkonen, Saku T.
Koivisto, Juha
Assessment of subjective image quality, contrast to noise ratio and modulation transfer function in the middle ear using a novel full body cone beam computed tomography device
title Assessment of subjective image quality, contrast to noise ratio and modulation transfer function in the middle ear using a novel full body cone beam computed tomography device
title_full Assessment of subjective image quality, contrast to noise ratio and modulation transfer function in the middle ear using a novel full body cone beam computed tomography device
title_fullStr Assessment of subjective image quality, contrast to noise ratio and modulation transfer function in the middle ear using a novel full body cone beam computed tomography device
title_full_unstemmed Assessment of subjective image quality, contrast to noise ratio and modulation transfer function in the middle ear using a novel full body cone beam computed tomography device
title_short Assessment of subjective image quality, contrast to noise ratio and modulation transfer function in the middle ear using a novel full body cone beam computed tomography device
title_sort assessment of subjective image quality, contrast to noise ratio and modulation transfer function in the middle ear using a novel full body cone beam computed tomography device
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10084678/
https://www.ncbi.nlm.nih.gov/pubmed/37038130
http://dx.doi.org/10.1186/s12880-023-00996-6
work_keys_str_mv AT heikkinenanssikalle assessmentofsubjectiveimagequalitycontrasttonoiseratioandmodulationtransferfunctioninthemiddleearusinganovelfullbodyconebeamcomputedtomographydevice
AT rissanenvaltteri assessmentofsubjectiveimagequalitycontrasttonoiseratioandmodulationtransferfunctioninthemiddleearusinganovelfullbodyconebeamcomputedtomographydevice
AT aarnisaloanttia assessmentofsubjectiveimagequalitycontrasttonoiseratioandmodulationtransferfunctioninthemiddleearusinganovelfullbodyconebeamcomputedtomographydevice
AT nymankristofer assessmentofsubjectiveimagequalitycontrasttonoiseratioandmodulationtransferfunctioninthemiddleearusinganovelfullbodyconebeamcomputedtomographydevice
AT sinkkonensakut assessmentofsubjectiveimagequalitycontrasttonoiseratioandmodulationtransferfunctioninthemiddleearusinganovelfullbodyconebeamcomputedtomographydevice
AT koivistojuha assessmentofsubjectiveimagequalitycontrasttonoiseratioandmodulationtransferfunctioninthemiddleearusinganovelfullbodyconebeamcomputedtomographydevice