First experimental measurements of 2D microdosimetry maps in proton therapy
BACKGROUND: Empirical data in proton therapy indicate that relative biological effectiveness (RBE) is not constant, and it is directly related to the linear energy transfer (LET). The experimental assessment of LET with high resolution would be a powerful tool for minimizing the LET hot spots in int...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10087596/ https://www.ncbi.nlm.nih.gov/pubmed/36066129 http://dx.doi.org/10.1002/mp.15945 |
_version_ | 1785022385143939072 |
---|---|
author | Guardiola, Consuelo Bachiller‐Perea, Diana Kole, Emmanuel M. Mate Fleta, Celeste Quirion, David De Marzi, Ludovic Gómez, Faustino |
author_facet | Guardiola, Consuelo Bachiller‐Perea, Diana Kole, Emmanuel M. Mate Fleta, Celeste Quirion, David De Marzi, Ludovic Gómez, Faustino |
author_sort | Guardiola, Consuelo |
collection | PubMed |
description | BACKGROUND: Empirical data in proton therapy indicate that relative biological effectiveness (RBE) is not constant, and it is directly related to the linear energy transfer (LET). The experimental assessment of LET with high resolution would be a powerful tool for minimizing the LET hot spots in intensity‐modulated proton therapy, RBE‐ or LET‐guided evaluation and optimization to achieve biologically optimized proton plans, verifying the theoretical predictions of variable proton RBE models, and so on. This could impact clinical outcomes by reducing toxicities in organs at risk. PURPOSE: The present work shows the first 2D LET maps obtained at a proton therapy facility using the double scattering delivery mode in clinical conditions by means of new silicon 3D‐cylindrical microdetectors. METHODS: The device consists of a matrix of 121 independent silicon‐based detectors that have 3D‐cylindrical electrodes of 25‐µm diameter and 20‐µm depth, resulting each one of them in a well‐defined micrometric radiation sensitive volume etched inside the silicon. They have been specifically designed for a hadron therapy, improving the performance of current silicon‐based microdosimeters. Microdosimetry spectra were obtained at different positions of the Bragg curve by using a water‐equivalent phantom along an 89‐MeV pristine proton beam generated in the Y1 proton passive scattering beamline of the Orsay Proton Therapy Centre (Institut Curie, France). RESULTS: Microdosimetry 2D‐maps showing the variation of the lineal energy with depth in the three dimensions were obtained in situ during irradiation at clinical fluence rates (∼10(8) s(−1) cm(−2)) for the first time with a spatial resolution of 200 µm, the highest achieved in the transverse plane so far. The experimental results were cross‐checked with Monte Carlo simulations and a good agreement between the spectra shapes was found. The experimental frequency‐mean lineal energy values in silicon were 1.858 ± 0.019 keV µm(−1) at the entrance, 2.61 ± 0.03 keV µm(−1) at the proximal distance, 4.97 ± 0.05 keV µm(−1) close to the Bragg peak, and 8.6 ± 0.1 keV µm(−1) at the distal edge. They are in good agreement with the expected trends in the literature in clinical proton beams. CONCLUSIONS: We present the first 2D microdosimetry maps obtained in situ during irradiation at clinical fluence rates in proton therapy. Our results show that the arrays of 3D‐cylindrical microdetectors are a reliable microdosimeter to evaluate LET maps not only in the longitudinal axis of the beam, but also in the transverse plane allowing for LET characterization in three dimensions. This work is a proof of principle showing the capacity of our system to deliver LET 2D maps. This kind of experimental data is needed to validate variable proton RBE models and to optimize LET‐guided plans. |
format | Online Article Text |
id | pubmed-10087596 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-100875962023-04-12 First experimental measurements of 2D microdosimetry maps in proton therapy Guardiola, Consuelo Bachiller‐Perea, Diana Kole, Emmanuel M. Mate Fleta, Celeste Quirion, David De Marzi, Ludovic Gómez, Faustino Med Phys COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY BACKGROUND: Empirical data in proton therapy indicate that relative biological effectiveness (RBE) is not constant, and it is directly related to the linear energy transfer (LET). The experimental assessment of LET with high resolution would be a powerful tool for minimizing the LET hot spots in intensity‐modulated proton therapy, RBE‐ or LET‐guided evaluation and optimization to achieve biologically optimized proton plans, verifying the theoretical predictions of variable proton RBE models, and so on. This could impact clinical outcomes by reducing toxicities in organs at risk. PURPOSE: The present work shows the first 2D LET maps obtained at a proton therapy facility using the double scattering delivery mode in clinical conditions by means of new silicon 3D‐cylindrical microdetectors. METHODS: The device consists of a matrix of 121 independent silicon‐based detectors that have 3D‐cylindrical electrodes of 25‐µm diameter and 20‐µm depth, resulting each one of them in a well‐defined micrometric radiation sensitive volume etched inside the silicon. They have been specifically designed for a hadron therapy, improving the performance of current silicon‐based microdosimeters. Microdosimetry spectra were obtained at different positions of the Bragg curve by using a water‐equivalent phantom along an 89‐MeV pristine proton beam generated in the Y1 proton passive scattering beamline of the Orsay Proton Therapy Centre (Institut Curie, France). RESULTS: Microdosimetry 2D‐maps showing the variation of the lineal energy with depth in the three dimensions were obtained in situ during irradiation at clinical fluence rates (∼10(8) s(−1) cm(−2)) for the first time with a spatial resolution of 200 µm, the highest achieved in the transverse plane so far. The experimental results were cross‐checked with Monte Carlo simulations and a good agreement between the spectra shapes was found. The experimental frequency‐mean lineal energy values in silicon were 1.858 ± 0.019 keV µm(−1) at the entrance, 2.61 ± 0.03 keV µm(−1) at the proximal distance, 4.97 ± 0.05 keV µm(−1) close to the Bragg peak, and 8.6 ± 0.1 keV µm(−1) at the distal edge. They are in good agreement with the expected trends in the literature in clinical proton beams. CONCLUSIONS: We present the first 2D microdosimetry maps obtained in situ during irradiation at clinical fluence rates in proton therapy. Our results show that the arrays of 3D‐cylindrical microdetectors are a reliable microdosimeter to evaluate LET maps not only in the longitudinal axis of the beam, but also in the transverse plane allowing for LET characterization in three dimensions. This work is a proof of principle showing the capacity of our system to deliver LET 2D maps. This kind of experimental data is needed to validate variable proton RBE models and to optimize LET‐guided plans. John Wiley and Sons Inc. 2022-09-19 2023-01 /pmc/articles/PMC10087596/ /pubmed/36066129 http://dx.doi.org/10.1002/mp.15945 Text en © 2022 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ (https://creativecommons.org/licenses/by-nc-nd/4.0/) License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made. |
spellingShingle | COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY Guardiola, Consuelo Bachiller‐Perea, Diana Kole, Emmanuel M. Mate Fleta, Celeste Quirion, David De Marzi, Ludovic Gómez, Faustino First experimental measurements of 2D microdosimetry maps in proton therapy |
title | First experimental measurements of 2D microdosimetry maps in proton therapy |
title_full | First experimental measurements of 2D microdosimetry maps in proton therapy |
title_fullStr | First experimental measurements of 2D microdosimetry maps in proton therapy |
title_full_unstemmed | First experimental measurements of 2D microdosimetry maps in proton therapy |
title_short | First experimental measurements of 2D microdosimetry maps in proton therapy |
title_sort | first experimental measurements of 2d microdosimetry maps in proton therapy |
topic | COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10087596/ https://www.ncbi.nlm.nih.gov/pubmed/36066129 http://dx.doi.org/10.1002/mp.15945 |
work_keys_str_mv | AT guardiolaconsuelo firstexperimentalmeasurementsof2dmicrodosimetrymapsinprotontherapy AT bachillerpereadiana firstexperimentalmeasurementsof2dmicrodosimetrymapsinprotontherapy AT koleemmanuelmmate firstexperimentalmeasurementsof2dmicrodosimetrymapsinprotontherapy AT fletaceleste firstexperimentalmeasurementsof2dmicrodosimetrymapsinprotontherapy AT quiriondavid firstexperimentalmeasurementsof2dmicrodosimetrymapsinprotontherapy AT demarziludovic firstexperimentalmeasurementsof2dmicrodosimetrymapsinprotontherapy AT gomezfaustino firstexperimentalmeasurementsof2dmicrodosimetrymapsinprotontherapy |