Steam-sterilized and degradable fused filament fabrication-printed polylactide/polyhydroxyalkanoate surgical guides for dental implants: Are they accurate enough for static navigation?

Three-dimensional (3D) printing is a rapidly evolving field and has gained increasing importance in the medical sector. However, the increasing usage of printing materials is accompanied by more wastages. With a rising awareness of the environmental impact of the medical sector, the development of h...

Descripción completa

Detalles Bibliográficos
Autores principales: Gielisch, Matthias, Heimes, Diana, Thiem, Daniel G. E., Boesing, C., Krumpholtz, M., Al-Nawas, B., Kämmerer, Peer W.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Whioce Publishing Pte. Ltd. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10090533/
https://www.ncbi.nlm.nih.gov/pubmed/37065653
http://dx.doi.org/10.18063/ijb.v9i2.655
_version_ 1785022979828088832
author Gielisch, Matthias
Heimes, Diana
Thiem, Daniel G. E.
Boesing, C.
Krumpholtz, M.
Al-Nawas, B.
Kämmerer, Peer W.
author_facet Gielisch, Matthias
Heimes, Diana
Thiem, Daniel G. E.
Boesing, C.
Krumpholtz, M.
Al-Nawas, B.
Kämmerer, Peer W.
author_sort Gielisch, Matthias
collection PubMed
description Three-dimensional (3D) printing is a rapidly evolving field and has gained increasing importance in the medical sector. However, the increasing usage of printing materials is accompanied by more wastages. With a rising awareness of the environmental impact of the medical sector, the development of highly accurate and biodegradable materials is of great interest. This study aims to compare the accuracy of polylactide/polyhydroxyalkanoate (PLA/PHA) surgical guides printed by fused filament fabrication and material jetted guides of MED610 in fully guided dental implant placement before and after steam sterilization. Five guides were tested in this study and each was either printed with PLA/PHA or MED610 and either steam-sterilized or not. After implant insertion in a 3D-printed upper jaw model, the divergence between planned and achieved implant position was calculated by digital superimposition. Angular deviation and 3D deviation at the base and the apex were determined. Non-sterilized PLA/PHA guides showed an angle deviation of 0.38 ± 0.53° compared to 2.88 ± 0.75° in sterile guides (P > 0.001), an offset of 0.49 ± 0.21 mm and 0.94 ± 0.23 mm (P < 0.05), and an offset at the apex of 0.50 ± 0.23 mm before and 1.04 ± 0.19 mm after steam sterilization (P < 0.025). No statistically significant difference could be shown for angle deviation or 3D offset at both locations for guides printed with MED610. PLA/PHA printing material showed significant deviations in angle and 3D accuracy after sterilization. However, the reached accuracy level is comparable to levels reached with materials already used in clinical routine and therefore, PLA/PHA surgical guide is a convenient and green alternative.
format Online
Article
Text
id pubmed-10090533
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Whioce Publishing Pte. Ltd.
record_format MEDLINE/PubMed
spelling pubmed-100905332023-04-13 Steam-sterilized and degradable fused filament fabrication-printed polylactide/polyhydroxyalkanoate surgical guides for dental implants: Are they accurate enough for static navigation? Gielisch, Matthias Heimes, Diana Thiem, Daniel G. E. Boesing, C. Krumpholtz, M. Al-Nawas, B. Kämmerer, Peer W. Int J Bioprint Research Article Three-dimensional (3D) printing is a rapidly evolving field and has gained increasing importance in the medical sector. However, the increasing usage of printing materials is accompanied by more wastages. With a rising awareness of the environmental impact of the medical sector, the development of highly accurate and biodegradable materials is of great interest. This study aims to compare the accuracy of polylactide/polyhydroxyalkanoate (PLA/PHA) surgical guides printed by fused filament fabrication and material jetted guides of MED610 in fully guided dental implant placement before and after steam sterilization. Five guides were tested in this study and each was either printed with PLA/PHA or MED610 and either steam-sterilized or not. After implant insertion in a 3D-printed upper jaw model, the divergence between planned and achieved implant position was calculated by digital superimposition. Angular deviation and 3D deviation at the base and the apex were determined. Non-sterilized PLA/PHA guides showed an angle deviation of 0.38 ± 0.53° compared to 2.88 ± 0.75° in sterile guides (P > 0.001), an offset of 0.49 ± 0.21 mm and 0.94 ± 0.23 mm (P < 0.05), and an offset at the apex of 0.50 ± 0.23 mm before and 1.04 ± 0.19 mm after steam sterilization (P < 0.025). No statistically significant difference could be shown for angle deviation or 3D offset at both locations for guides printed with MED610. PLA/PHA printing material showed significant deviations in angle and 3D accuracy after sterilization. However, the reached accuracy level is comparable to levels reached with materials already used in clinical routine and therefore, PLA/PHA surgical guide is a convenient and green alternative. Whioce Publishing Pte. Ltd. 2022-12-23 /pmc/articles/PMC10090533/ /pubmed/37065653 http://dx.doi.org/10.18063/ijb.v9i2.655 Text en Copyright: © 2022 Author(s). https://creativecommons.org/licenses/by-nc/4.0/This is an Open-Access article distributed under the terms of the Creative Commons Attribution-Noncommercial License, permitting all noncommercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Gielisch, Matthias
Heimes, Diana
Thiem, Daniel G. E.
Boesing, C.
Krumpholtz, M.
Al-Nawas, B.
Kämmerer, Peer W.
Steam-sterilized and degradable fused filament fabrication-printed polylactide/polyhydroxyalkanoate surgical guides for dental implants: Are they accurate enough for static navigation?
title Steam-sterilized and degradable fused filament fabrication-printed polylactide/polyhydroxyalkanoate surgical guides for dental implants: Are they accurate enough for static navigation?
title_full Steam-sterilized and degradable fused filament fabrication-printed polylactide/polyhydroxyalkanoate surgical guides for dental implants: Are they accurate enough for static navigation?
title_fullStr Steam-sterilized and degradable fused filament fabrication-printed polylactide/polyhydroxyalkanoate surgical guides for dental implants: Are they accurate enough for static navigation?
title_full_unstemmed Steam-sterilized and degradable fused filament fabrication-printed polylactide/polyhydroxyalkanoate surgical guides for dental implants: Are they accurate enough for static navigation?
title_short Steam-sterilized and degradable fused filament fabrication-printed polylactide/polyhydroxyalkanoate surgical guides for dental implants: Are they accurate enough for static navigation?
title_sort steam-sterilized and degradable fused filament fabrication-printed polylactide/polyhydroxyalkanoate surgical guides for dental implants: are they accurate enough for static navigation?
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10090533/
https://www.ncbi.nlm.nih.gov/pubmed/37065653
http://dx.doi.org/10.18063/ijb.v9i2.655
work_keys_str_mv AT gielischmatthias steamsterilizedanddegradablefusedfilamentfabricationprintedpolylactidepolyhydroxyalkanoatesurgicalguidesfordentalimplantsaretheyaccurateenoughforstaticnavigation
AT heimesdiana steamsterilizedanddegradablefusedfilamentfabricationprintedpolylactidepolyhydroxyalkanoatesurgicalguidesfordentalimplantsaretheyaccurateenoughforstaticnavigation
AT thiemdanielge steamsterilizedanddegradablefusedfilamentfabricationprintedpolylactidepolyhydroxyalkanoatesurgicalguidesfordentalimplantsaretheyaccurateenoughforstaticnavigation
AT boesingc steamsterilizedanddegradablefusedfilamentfabricationprintedpolylactidepolyhydroxyalkanoatesurgicalguidesfordentalimplantsaretheyaccurateenoughforstaticnavigation
AT krumpholtzm steamsterilizedanddegradablefusedfilamentfabricationprintedpolylactidepolyhydroxyalkanoatesurgicalguidesfordentalimplantsaretheyaccurateenoughforstaticnavigation
AT alnawasb steamsterilizedanddegradablefusedfilamentfabricationprintedpolylactidepolyhydroxyalkanoatesurgicalguidesfordentalimplantsaretheyaccurateenoughforstaticnavigation
AT kammererpeerw steamsterilizedanddegradablefusedfilamentfabricationprintedpolylactidepolyhydroxyalkanoatesurgicalguidesfordentalimplantsaretheyaccurateenoughforstaticnavigation