Can We Completely Trust in Automated Software for Preoperative Planning of Shoulder Arthroplasty? Software Update May Modify Glenoid Version, Glenoid Inclination and Humeral Head Subluxation Values

Background: The purpose of this study was to evaluate the impact of software updating on measurements of the glenoid inclination and version, along with humeral head subluxation performed by an automated 3D planning program. The hypothesis was that the software update could significantly modify the...

Descripción completa

Detalles Bibliográficos
Autores principales: Garofalo, Raffaele, Fontanarosa, Alberto, Castagna, Alessandro, Lassandro, Nunzio, Del Buono, Angelo, De Crescenzo, Angelo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10094783/
https://www.ncbi.nlm.nih.gov/pubmed/37048703
http://dx.doi.org/10.3390/jcm12072620
_version_ 1785023924119011328
author Garofalo, Raffaele
Fontanarosa, Alberto
Castagna, Alessandro
Lassandro, Nunzio
Del Buono, Angelo
De Crescenzo, Angelo
author_facet Garofalo, Raffaele
Fontanarosa, Alberto
Castagna, Alessandro
Lassandro, Nunzio
Del Buono, Angelo
De Crescenzo, Angelo
author_sort Garofalo, Raffaele
collection PubMed
description Background: The purpose of this study was to evaluate the impact of software updating on measurements of the glenoid inclination and version, along with humeral head subluxation performed by an automated 3D planning program. The hypothesis was that the software update could significantly modify the values of the glenoid inclination and version, as well as of the humeral head subluxation. Methods: A comprehensive pool of 76 shoulder computed tomography (CT) scans of patients who underwent total shoulder arthroplasty (TSA) or reverse total shoulder arthroplasty (RTSA) were analyzed with the automated program Blueprint in 2018 and again in 2020 after a software update. Results: A statistically significant difference of 8.1 ± 8.2 and 5.4 ± 7.8 (mean difference of −2.8 ± 5.0, p < 0.001) was indeed reached when comparing the mean glenoid inclination achieved with Blueprint 2018 and Blueprint 2020, respectively. The glenoid version, as well as the humeral head subluxation evaluations, were not significantly different between the two software versions, with mean values being −9.4 ± 8.9 and −9.0 ± 7.4 and 60.1 ± 12.6 and 61.8 ± 12.0, respectively (p = 0.708 and p = 0.115, respectively). In 22% of CT scans, the software update determined a variation of the glenoid inclination of more than 5° or 10°. Conclusion: The present study shows the software update of an automated preoperative planning program may significantly modify the values of glenoid inclination. Even though without a significant difference, variations were also found for the glenoid version and humeral head subluxation. Accordingly, these results should further advise surgeons to carefully and critically evaluate data acquired with automated software.
format Online
Article
Text
id pubmed-10094783
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-100947832023-04-13 Can We Completely Trust in Automated Software for Preoperative Planning of Shoulder Arthroplasty? Software Update May Modify Glenoid Version, Glenoid Inclination and Humeral Head Subluxation Values Garofalo, Raffaele Fontanarosa, Alberto Castagna, Alessandro Lassandro, Nunzio Del Buono, Angelo De Crescenzo, Angelo J Clin Med Article Background: The purpose of this study was to evaluate the impact of software updating on measurements of the glenoid inclination and version, along with humeral head subluxation performed by an automated 3D planning program. The hypothesis was that the software update could significantly modify the values of the glenoid inclination and version, as well as of the humeral head subluxation. Methods: A comprehensive pool of 76 shoulder computed tomography (CT) scans of patients who underwent total shoulder arthroplasty (TSA) or reverse total shoulder arthroplasty (RTSA) were analyzed with the automated program Blueprint in 2018 and again in 2020 after a software update. Results: A statistically significant difference of 8.1 ± 8.2 and 5.4 ± 7.8 (mean difference of −2.8 ± 5.0, p < 0.001) was indeed reached when comparing the mean glenoid inclination achieved with Blueprint 2018 and Blueprint 2020, respectively. The glenoid version, as well as the humeral head subluxation evaluations, were not significantly different between the two software versions, with mean values being −9.4 ± 8.9 and −9.0 ± 7.4 and 60.1 ± 12.6 and 61.8 ± 12.0, respectively (p = 0.708 and p = 0.115, respectively). In 22% of CT scans, the software update determined a variation of the glenoid inclination of more than 5° or 10°. Conclusion: The present study shows the software update of an automated preoperative planning program may significantly modify the values of glenoid inclination. Even though without a significant difference, variations were also found for the glenoid version and humeral head subluxation. Accordingly, these results should further advise surgeons to carefully and critically evaluate data acquired with automated software. MDPI 2023-03-30 /pmc/articles/PMC10094783/ /pubmed/37048703 http://dx.doi.org/10.3390/jcm12072620 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Garofalo, Raffaele
Fontanarosa, Alberto
Castagna, Alessandro
Lassandro, Nunzio
Del Buono, Angelo
De Crescenzo, Angelo
Can We Completely Trust in Automated Software for Preoperative Planning of Shoulder Arthroplasty? Software Update May Modify Glenoid Version, Glenoid Inclination and Humeral Head Subluxation Values
title Can We Completely Trust in Automated Software for Preoperative Planning of Shoulder Arthroplasty? Software Update May Modify Glenoid Version, Glenoid Inclination and Humeral Head Subluxation Values
title_full Can We Completely Trust in Automated Software for Preoperative Planning of Shoulder Arthroplasty? Software Update May Modify Glenoid Version, Glenoid Inclination and Humeral Head Subluxation Values
title_fullStr Can We Completely Trust in Automated Software for Preoperative Planning of Shoulder Arthroplasty? Software Update May Modify Glenoid Version, Glenoid Inclination and Humeral Head Subluxation Values
title_full_unstemmed Can We Completely Trust in Automated Software for Preoperative Planning of Shoulder Arthroplasty? Software Update May Modify Glenoid Version, Glenoid Inclination and Humeral Head Subluxation Values
title_short Can We Completely Trust in Automated Software for Preoperative Planning of Shoulder Arthroplasty? Software Update May Modify Glenoid Version, Glenoid Inclination and Humeral Head Subluxation Values
title_sort can we completely trust in automated software for preoperative planning of shoulder arthroplasty? software update may modify glenoid version, glenoid inclination and humeral head subluxation values
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10094783/
https://www.ncbi.nlm.nih.gov/pubmed/37048703
http://dx.doi.org/10.3390/jcm12072620
work_keys_str_mv AT garofaloraffaele canwecompletelytrustinautomatedsoftwareforpreoperativeplanningofshoulderarthroplastysoftwareupdatemaymodifyglenoidversionglenoidinclinationandhumeralheadsubluxationvalues
AT fontanarosaalberto canwecompletelytrustinautomatedsoftwareforpreoperativeplanningofshoulderarthroplastysoftwareupdatemaymodifyglenoidversionglenoidinclinationandhumeralheadsubluxationvalues
AT castagnaalessandro canwecompletelytrustinautomatedsoftwareforpreoperativeplanningofshoulderarthroplastysoftwareupdatemaymodifyglenoidversionglenoidinclinationandhumeralheadsubluxationvalues
AT lassandronunzio canwecompletelytrustinautomatedsoftwareforpreoperativeplanningofshoulderarthroplastysoftwareupdatemaymodifyglenoidversionglenoidinclinationandhumeralheadsubluxationvalues
AT delbuonoangelo canwecompletelytrustinautomatedsoftwareforpreoperativeplanningofshoulderarthroplastysoftwareupdatemaymodifyglenoidversionglenoidinclinationandhumeralheadsubluxationvalues
AT decrescenzoangelo canwecompletelytrustinautomatedsoftwareforpreoperativeplanningofshoulderarthroplastysoftwareupdatemaymodifyglenoidversionglenoidinclinationandhumeralheadsubluxationvalues