Repeated Measurements Are Necessary for Evaluating Accurate Diurnal Rhythm Using a Self-Intraocular Pressure Measurement Device

Purpose: To investigate how many tests need to be performed to adequately assess intraocular pressure (IOP) diurnal change using a self-measuring rebound tonometer among glaucoma patients. Subjects and Methods: Adult patients with primary open-angle glaucoma were included. IOP was measured in the mo...

Descripción completa

Detalles Bibliográficos
Autores principales: Shigemoto, Yumi, Hasebe, Yuka, Kitamura, Kazuyoshi, Fukuda, Yoshiko, Sakamoto, Masako, Matsubara, Mio, Minaguchi, Shinya, Kashiwagi, Kenji
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10094971/
https://www.ncbi.nlm.nih.gov/pubmed/37048542
http://dx.doi.org/10.3390/jcm12072460
Descripción
Sumario:Purpose: To investigate how many tests need to be performed to adequately assess intraocular pressure (IOP) diurnal change using a self-measuring rebound tonometer among glaucoma patients. Subjects and Methods: Adult patients with primary open-angle glaucoma were included. IOP was measured in the morning (6 AM to 9 AM), afternoon (12 PM to 3 PM), and at night (6 PM to 9 PM) for seven consecutive days. Twenty-four (7 males and 17 females, mean age 59.5 ± 11.0 years) patients who successfully measured IOP at least three times per day during the correct time periods for four days were subjected to analysis. Results: The IOP rhythm was significantly greater on the first day of measurement (6.6 ± 3.6 mmHg) than that averaged during subsequent days (4.4 ± 2.2 mmHg). The time of the highest and lowest IOP measurements on the first day of IOP measurement and during the entire measurement period coincided in 72.9% and 64.6% of cases, respectively. The concordance rate of the highest IOP time between the whole measurement period and each measurement day was less than 60%. Conclusion: The diurnal IOP rhythm measured by the patients themselves was not consistent, and multiple days of measurements may be necessary to correctly assess diurnal IOP rhythm.