Cargando…
Catalytic Biosensors Operating under Quasi‐Equilibrium Conditions for Mitigating the Changes in Substrate Diffusion
Despite the success of continuous glucose measuring systems operating through the skin for about 14 days, long‐term implantable biosensors are facing challenges caused by the foreign‐body reaction. We present a conceptually new strategy using catalytic enzyme‐based biosensors based on a measuring se...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10099152/ https://www.ncbi.nlm.nih.gov/pubmed/36253337 http://dx.doi.org/10.1002/anie.202211559 |
Sumario: | Despite the success of continuous glucose measuring systems operating through the skin for about 14 days, long‐term implantable biosensors are facing challenges caused by the foreign‐body reaction. We present a conceptually new strategy using catalytic enzyme‐based biosensors based on a measuring sequence leading to minimum disturbance of the substrate equilibrium concentration by controlling the sensor between “on” and “off” state combined with short potentiometric data acquisition. It is required that the enzyme activity can be completely switched off and no parasitic side reactions allow substrate turnover. This is achieved by using an O(2)‐independent FAD‐dependent glucose dehydrogenase embedded within a crosslinked redox polymer. A short measuring interval allows the glucose concentration equilibrium to be restored quickly which enables the biosensor to operate under quasi‐equilibrium conditions. |
---|