P074 Three models of pulse oximeters are within specifications when assessed using a benchtop pulse oximeter simulator

To assess possible SpO2 accuracy differences between 3 different (oximeter & software) testing techniques. Hypothesised that differences in SpO2 would be within the oximeter manufacturer’s specification. A synthetic digit, OX-1 OxSim® Optical SpO2 Pulse Oximeter Simulator (Pronk Technologies Inc...

Descripción completa

Detalles Bibliográficos
Autores principales: McDermott, E, Cuesta, R, Nguy, R, VanBraak, E, Naughton, M, Roebuck, T
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10109057/
http://dx.doi.org/10.1093/sleepadvances/zpac029.144
_version_ 1785026976077053952
author McDermott, E
Cuesta, R
Nguy, R
VanBraak, E
Naughton, M
Roebuck, T
author_facet McDermott, E
Cuesta, R
Nguy, R
VanBraak, E
Naughton, M
Roebuck, T
author_sort McDermott, E
collection PubMed
description To assess possible SpO2 accuracy differences between 3 different (oximeter & software) testing techniques. Hypothesised that differences in SpO2 would be within the oximeter manufacturer’s specification. A synthetic digit, OX-1 OxSim® Optical SpO2 Pulse Oximeter Simulator (Pronk Technologies Inc., Sun Valley, CA) can provide simulated SpO2 & HR values : 85% (80BPM), 98% (80BPM), 98% (140BPM), 99% (80BPM) at low perfusion and 95% (40BPM) low HR. Simulated SpO2 delivered for 5 minutes, average calculated from signal recorded with corresponding download software: Rad-5 (Masimo, Irvine, CA, USA, 2sec averaging time) - Visi download (Stowood Scientific, Oxford, UK), the Rad-7 (Masimo, Irvine, CA, USA, 2-4sec averaging time) - PSG 4 (Compumedics, Melbourne, Australia), Wrist-Ox2® (Nonin Medical Inc, Plymouth, MN, USA, 3sec averaging time) - Noxturnal software (Nox Medical Inc, Reykjavik, Iceland). Four devices for each model Rad-5, Rad-7 and Nox Wrist-Ox2® were examined at 5 SpO2 & HR settings. The SpO2 downloaded from each device was compared with the OxSim® values. Oximeter reading v OxSim® : at 85%, Rad-5 (y = -0.146x + 14.375, R² = 0.9627 bias 0.606), Rad-7 (y = -0.0897x + 8.737, R² = 0.5295 bias 0.295), Wrist-Ox2® (y = -0.0161x + 1.2261, R² = 0.0481 bias -0.281). At all other comparison values, including low perfusion and low HR, all values fell within +/- 2% SpO2. Small differences in the measured SpO2 for the 3 oximeter models & paired software were noted, however all devices gave SpO2 values within 2% of simulated SpO2 values as per manufacturer specifications.
format Online
Article
Text
id pubmed-10109057
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-101090572023-05-15 P074 Three models of pulse oximeters are within specifications when assessed using a benchtop pulse oximeter simulator McDermott, E Cuesta, R Nguy, R VanBraak, E Naughton, M Roebuck, T Sleep Adv Poster Presentations To assess possible SpO2 accuracy differences between 3 different (oximeter & software) testing techniques. Hypothesised that differences in SpO2 would be within the oximeter manufacturer’s specification. A synthetic digit, OX-1 OxSim® Optical SpO2 Pulse Oximeter Simulator (Pronk Technologies Inc., Sun Valley, CA) can provide simulated SpO2 & HR values : 85% (80BPM), 98% (80BPM), 98% (140BPM), 99% (80BPM) at low perfusion and 95% (40BPM) low HR. Simulated SpO2 delivered for 5 minutes, average calculated from signal recorded with corresponding download software: Rad-5 (Masimo, Irvine, CA, USA, 2sec averaging time) - Visi download (Stowood Scientific, Oxford, UK), the Rad-7 (Masimo, Irvine, CA, USA, 2-4sec averaging time) - PSG 4 (Compumedics, Melbourne, Australia), Wrist-Ox2® (Nonin Medical Inc, Plymouth, MN, USA, 3sec averaging time) - Noxturnal software (Nox Medical Inc, Reykjavik, Iceland). Four devices for each model Rad-5, Rad-7 and Nox Wrist-Ox2® were examined at 5 SpO2 & HR settings. The SpO2 downloaded from each device was compared with the OxSim® values. Oximeter reading v OxSim® : at 85%, Rad-5 (y = -0.146x + 14.375, R² = 0.9627 bias 0.606), Rad-7 (y = -0.0897x + 8.737, R² = 0.5295 bias 0.295), Wrist-Ox2® (y = -0.0161x + 1.2261, R² = 0.0481 bias -0.281). At all other comparison values, including low perfusion and low HR, all values fell within +/- 2% SpO2. Small differences in the measured SpO2 for the 3 oximeter models & paired software were noted, however all devices gave SpO2 values within 2% of simulated SpO2 values as per manufacturer specifications. Oxford University Press 2022-11-09 /pmc/articles/PMC10109057/ http://dx.doi.org/10.1093/sleepadvances/zpac029.144 Text en © The Author(s) 2022. Published by Oxford University Press on behalf of Sleep Research Society. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs licence (https://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial reproduction and distribution of the work, in any medium, provided the original work is not altered or transformed in any way, and that the work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle Poster Presentations
McDermott, E
Cuesta, R
Nguy, R
VanBraak, E
Naughton, M
Roebuck, T
P074 Three models of pulse oximeters are within specifications when assessed using a benchtop pulse oximeter simulator
title P074 Three models of pulse oximeters are within specifications when assessed using a benchtop pulse oximeter simulator
title_full P074 Three models of pulse oximeters are within specifications when assessed using a benchtop pulse oximeter simulator
title_fullStr P074 Three models of pulse oximeters are within specifications when assessed using a benchtop pulse oximeter simulator
title_full_unstemmed P074 Three models of pulse oximeters are within specifications when assessed using a benchtop pulse oximeter simulator
title_short P074 Three models of pulse oximeters are within specifications when assessed using a benchtop pulse oximeter simulator
title_sort p074 three models of pulse oximeters are within specifications when assessed using a benchtop pulse oximeter simulator
topic Poster Presentations
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10109057/
http://dx.doi.org/10.1093/sleepadvances/zpac029.144
work_keys_str_mv AT mcdermotte p074threemodelsofpulseoximetersarewithinspecificationswhenassessedusingabenchtoppulseoximetersimulator
AT cuestar p074threemodelsofpulseoximetersarewithinspecificationswhenassessedusingabenchtoppulseoximetersimulator
AT nguyr p074threemodelsofpulseoximetersarewithinspecificationswhenassessedusingabenchtoppulseoximetersimulator
AT vanbraake p074threemodelsofpulseoximetersarewithinspecificationswhenassessedusingabenchtoppulseoximetersimulator
AT naughtonm p074threemodelsofpulseoximetersarewithinspecificationswhenassessedusingabenchtoppulseoximetersimulator
AT roebuckt p074threemodelsofpulseoximetersarewithinspecificationswhenassessedusingabenchtoppulseoximetersimulator