P035 Is the Grael oximetry averaging time interchangeable with a Masimo pulse oximeter algorithm in polysomnography?

Compumedics recording software (Grael V2) for polysomnography (PSG) calculates SpO2 values using a 3-heartbeat long averaging window. This is derived from the ECG and thus introduces variability in the averaging time that is dependent on the heart rate. Little is known about the effect this has on t...

Descripción completa

Detalles Bibliográficos
Autores principales: Eritaia, J, Suthers, B
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10109304/
http://dx.doi.org/10.1093/sleepadvances/zpab014.083
_version_ 1785027034624294912
author Eritaia, J
Suthers, B
author_facet Eritaia, J
Suthers, B
author_sort Eritaia, J
collection PubMed
description Compumedics recording software (Grael V2) for polysomnography (PSG) calculates SpO2 values using a 3-heartbeat long averaging window. This is derived from the ECG and thus introduces variability in the averaging time that is dependent on the heart rate. Little is known about the effect this has on the common oximetry metrics used in PSG interpretation. This study explorer the interchangeability of the Grael V2 inbuilt 3-beat averaging algorithm with a short averaging window of 2 - 4 seconds using a Masimo Radical 7 pulse oximeter during a PSG. SpO2 data were collected from 2 oximeter probes (Grael and Radical 7) both attached to a patient’s fingers. After SpO2 artifacts were removed, the following SpO2 parameters from each oximeter were generated: mean sleep SpO2, oxygen desaturation index (ODI) using 2%, 3% and 4% drop in SpO2 in sleep, total sleep time (TST) with SpO2 < 90% and < 80% as well as time spent < SpO2 88% in minutes. 88 sleep studies were included in the data collection. For ODI2%, 3% and 4%, bias (95% limits of agreement) values were -0.75 events/hr (9.99 to -11.49 events/hr), -0.74 events/hr (10.00 - -11.49 events/hr) and -0.20 events/hr (8.45 - -8.86 events/hr) respectively. There was no significant difference between measurements except for the mean sleep SpO2 values, p < .001. Although no bias found between measurements, there was poor agreement between the algorithms as demonstrated by the wide 95% limits of agreement suggesting that the two oximeter devices are not interchangeable.
format Online
Article
Text
id pubmed-10109304
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-101093042023-05-15 P035 Is the Grael oximetry averaging time interchangeable with a Masimo pulse oximeter algorithm in polysomnography? Eritaia, J Suthers, B Sleep Adv Poster Presentations Compumedics recording software (Grael V2) for polysomnography (PSG) calculates SpO2 values using a 3-heartbeat long averaging window. This is derived from the ECG and thus introduces variability in the averaging time that is dependent on the heart rate. Little is known about the effect this has on the common oximetry metrics used in PSG interpretation. This study explorer the interchangeability of the Grael V2 inbuilt 3-beat averaging algorithm with a short averaging window of 2 - 4 seconds using a Masimo Radical 7 pulse oximeter during a PSG. SpO2 data were collected from 2 oximeter probes (Grael and Radical 7) both attached to a patient’s fingers. After SpO2 artifacts were removed, the following SpO2 parameters from each oximeter were generated: mean sleep SpO2, oxygen desaturation index (ODI) using 2%, 3% and 4% drop in SpO2 in sleep, total sleep time (TST) with SpO2 < 90% and < 80% as well as time spent < SpO2 88% in minutes. 88 sleep studies were included in the data collection. For ODI2%, 3% and 4%, bias (95% limits of agreement) values were -0.75 events/hr (9.99 to -11.49 events/hr), -0.74 events/hr (10.00 - -11.49 events/hr) and -0.20 events/hr (8.45 - -8.86 events/hr) respectively. There was no significant difference between measurements except for the mean sleep SpO2 values, p < .001. Although no bias found between measurements, there was poor agreement between the algorithms as demonstrated by the wide 95% limits of agreement suggesting that the two oximeter devices are not interchangeable. Oxford University Press 2021-10-07 /pmc/articles/PMC10109304/ http://dx.doi.org/10.1093/sleepadvances/zpab014.083 Text en © The Author(s) 2021. Published by Oxford University Press on behalf of Sleep Research Society. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs licence (https://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial reproduction and distribution of the work, in any medium, provided the original work is not altered or transformed in any way, and that the work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle Poster Presentations
Eritaia, J
Suthers, B
P035 Is the Grael oximetry averaging time interchangeable with a Masimo pulse oximeter algorithm in polysomnography?
title P035 Is the Grael oximetry averaging time interchangeable with a Masimo pulse oximeter algorithm in polysomnography?
title_full P035 Is the Grael oximetry averaging time interchangeable with a Masimo pulse oximeter algorithm in polysomnography?
title_fullStr P035 Is the Grael oximetry averaging time interchangeable with a Masimo pulse oximeter algorithm in polysomnography?
title_full_unstemmed P035 Is the Grael oximetry averaging time interchangeable with a Masimo pulse oximeter algorithm in polysomnography?
title_short P035 Is the Grael oximetry averaging time interchangeable with a Masimo pulse oximeter algorithm in polysomnography?
title_sort p035 is the grael oximetry averaging time interchangeable with a masimo pulse oximeter algorithm in polysomnography?
topic Poster Presentations
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10109304/
http://dx.doi.org/10.1093/sleepadvances/zpab014.083
work_keys_str_mv AT eritaiaj p035isthegraeloximetryaveragingtimeinterchangeablewithamasimopulseoximeteralgorithminpolysomnography
AT suthersb p035isthegraeloximetryaveragingtimeinterchangeablewithamasimopulseoximeteralgorithminpolysomnography