Cargando…

Contamination of pulse oximeter probes before and after decontamination in two intensive care units

BACKGROUND: The internal surfaces of pulse oximeter probes may be overlooked as hot spots for pathogenic microorganisms in an intensive care unit (ICU), thereby contributing to the high incidence of hospital-acquired infections. OBJECTIVES: To determine the growth and identification of microorganism...

Descripción completa

Detalles Bibliográficos
Autores principales: Desai, F, Scribante, J, Perrie, H, Fourtounas, M
Formato: Online Artículo Texto
Lenguaje:English
Publicado: South African Medical Association 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10109541/
https://www.ncbi.nlm.nih.gov/pubmed/37077522
http://dx.doi.org/10.7196/SAJCC.2019.v35i2.394
_version_ 1785027088679436288
author Desai, F
Scribante, J
Perrie, H
Fourtounas, M
author_facet Desai, F
Scribante, J
Perrie, H
Fourtounas, M
author_sort Desai, F
collection PubMed
description BACKGROUND: The internal surfaces of pulse oximeter probes may be overlooked as hot spots for pathogenic microorganisms in an intensive care unit (ICU), thereby contributing to the high incidence of hospital-acquired infections. OBJECTIVES: To determine the growth and identification of microorganisms on pulse oximeter probes in the multidisciplinary ICU (MICU) at Charlotte Maxeke Johannesburg Academic Hospital and the burns ICU (BICU) at Chris Hani Baragwanath Academic Hospital, before and after decontamination. METHODS: This was a cross-sectional, comparative and contextual study, using purposive sampling. Data were collected from the internal surfaces of 34 pulse oximeter probes in a MICU and BICU. Each pulse oximeter probe was swabbed before and after decontamination. The endemic microorganism profile for the two ICUs was obtained from a laboratory database. RESULTS: Internal surfaces of 31 (91%; 95% confidence interval (CI) 0.76 - 0.98) pulse oximeter probes were contaminated with 9 different pathogenic microorganisms pre decontamination. Acinetobacter baumannii, Klebsiella pneumoniaei and Pseudomonas aeruginosa were endemic to both ICUs, and were the most-frequently isolated microorganisms. Staphylococcus aureus was the most common microorganism endemic to both ICUs, isolated on the internal surfaces of only 2 pulse oximeter probes. Of the internal surfaces of pulse oximeter probes, 6 (18%; 95% CI 0.07 - 0.35) remained contaminated post decontamination, with a microorganism growth reduction of 80% (p=0.0001). CONCLUSION: The internal surfaces of pulse oximeter probes may serve as hot spots for an array of pathogens with the potential to cause infection and outbreaks in ICUs. Decontamination of the internal surfaces of pulse oximeter probes should be emphasised. CONTRIBUTIONS OF THE STUDY: This study identifies internal surfaces of pulse oximeter probes as reservoirs for infection in intensive care units (ICUs), particularly in a burns ICU setting, pinpointing one of many sources of hospital-acquired infections within ICU cubicles. The study emphasises the need to clean the internal surfaces of pulse oximeter probes (whether visibly soiled or not) prior to disinfection.
format Online
Article
Text
id pubmed-10109541
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher South African Medical Association
record_format MEDLINE/PubMed
spelling pubmed-101095412023-04-18 Contamination of pulse oximeter probes before and after decontamination in two intensive care units Desai, F Scribante, J Perrie, H Fourtounas, M South Afr J Crit Care Research BACKGROUND: The internal surfaces of pulse oximeter probes may be overlooked as hot spots for pathogenic microorganisms in an intensive care unit (ICU), thereby contributing to the high incidence of hospital-acquired infections. OBJECTIVES: To determine the growth and identification of microorganisms on pulse oximeter probes in the multidisciplinary ICU (MICU) at Charlotte Maxeke Johannesburg Academic Hospital and the burns ICU (BICU) at Chris Hani Baragwanath Academic Hospital, before and after decontamination. METHODS: This was a cross-sectional, comparative and contextual study, using purposive sampling. Data were collected from the internal surfaces of 34 pulse oximeter probes in a MICU and BICU. Each pulse oximeter probe was swabbed before and after decontamination. The endemic microorganism profile for the two ICUs was obtained from a laboratory database. RESULTS: Internal surfaces of 31 (91%; 95% confidence interval (CI) 0.76 - 0.98) pulse oximeter probes were contaminated with 9 different pathogenic microorganisms pre decontamination. Acinetobacter baumannii, Klebsiella pneumoniaei and Pseudomonas aeruginosa were endemic to both ICUs, and were the most-frequently isolated microorganisms. Staphylococcus aureus was the most common microorganism endemic to both ICUs, isolated on the internal surfaces of only 2 pulse oximeter probes. Of the internal surfaces of pulse oximeter probes, 6 (18%; 95% CI 0.07 - 0.35) remained contaminated post decontamination, with a microorganism growth reduction of 80% (p=0.0001). CONCLUSION: The internal surfaces of pulse oximeter probes may serve as hot spots for an array of pathogens with the potential to cause infection and outbreaks in ICUs. Decontamination of the internal surfaces of pulse oximeter probes should be emphasised. CONTRIBUTIONS OF THE STUDY: This study identifies internal surfaces of pulse oximeter probes as reservoirs for infection in intensive care units (ICUs), particularly in a burns ICU setting, pinpointing one of many sources of hospital-acquired infections within ICU cubicles. The study emphasises the need to clean the internal surfaces of pulse oximeter probes (whether visibly soiled or not) prior to disinfection. South African Medical Association 2019-11-01 /pmc/articles/PMC10109541/ /pubmed/37077522 http://dx.doi.org/10.7196/SAJCC.2019.v35i2.394 Text en https://creativecommons.org/licenses/by-nc/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution - NonCommercial Works License (CC BY-NC 4.0) (https://creativecommons.org/licenses/by-nc/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research
Desai, F
Scribante, J
Perrie, H
Fourtounas, M
Contamination of pulse oximeter probes before and after decontamination in two intensive care units
title Contamination of pulse oximeter probes before and after decontamination in two intensive care units
title_full Contamination of pulse oximeter probes before and after decontamination in two intensive care units
title_fullStr Contamination of pulse oximeter probes before and after decontamination in two intensive care units
title_full_unstemmed Contamination of pulse oximeter probes before and after decontamination in two intensive care units
title_short Contamination of pulse oximeter probes before and after decontamination in two intensive care units
title_sort contamination of pulse oximeter probes before and after decontamination in two intensive care units
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10109541/
https://www.ncbi.nlm.nih.gov/pubmed/37077522
http://dx.doi.org/10.7196/SAJCC.2019.v35i2.394
work_keys_str_mv AT desaif contaminationofpulseoximeterprobesbeforeandafterdecontaminationintwointensivecareunits
AT scribantej contaminationofpulseoximeterprobesbeforeandafterdecontaminationintwointensivecareunits
AT perrieh contaminationofpulseoximeterprobesbeforeandafterdecontaminationintwointensivecareunits
AT fourtounasm contaminationofpulseoximeterprobesbeforeandafterdecontaminationintwointensivecareunits