Stretched and fractured Neuroform Atlas(®) stent during a stent‑assisted coil embolization: A case report
The Neuroform Atlas(®) stent is one of the most recently developed stents for coil embolization, with advancements in a lower-profile delivery system, enhanced trackability, smaller cell size, and increased wall conformability. Because of these advantages, the Neuroform Atlas(®) stent shows high tec...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
D.A. Spandidos
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10119665/ https://www.ncbi.nlm.nih.gov/pubmed/37090084 http://dx.doi.org/10.3892/etm.2023.11906 |
_version_ | 1785029057517191168 |
---|---|
author | Ko, Hak Cheol Shin, Hee Sup |
author_facet | Ko, Hak Cheol Shin, Hee Sup |
author_sort | Ko, Hak Cheol |
collection | PubMed |
description | The Neuroform Atlas(®) stent is one of the most recently developed stents for coil embolization, with advancements in a lower-profile delivery system, enhanced trackability, smaller cell size, and increased wall conformability. Because of these advantages, the Neuroform Atlas(®) stent shows high technical success with few procedure-related complications. However, the present study reported a rare complication of a stretched and partially fractured Neuroform Atlas(®) stent due to unexpected partial withdrawal of microcatheter during deployment for coil embolization of an intracranial aneurysm. The measured length of the stent was ~30 mm, which was greater than the normal length (21 mm). An additional stent was inserted into the distal part of the deployed stent to stabilize the damaged stent and remodel the aneurysm neck. This complication was considered to potentially result from the combination of several factors, including: Curved vessel; open-cell stent; unexpected microcatheter withdrawal during stent deployment; and hooking of the aneurysm selecting microcatheter with stent strut. Understanding the stent design and careful manipulation while avoiding unexpected withdrawal of the microcatheter could prevent this complication. |
format | Online Article Text |
id | pubmed-10119665 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | D.A. Spandidos |
record_format | MEDLINE/PubMed |
spelling | pubmed-101196652023-04-22 Stretched and fractured Neuroform Atlas(®) stent during a stent‑assisted coil embolization: A case report Ko, Hak Cheol Shin, Hee Sup Exp Ther Med Case Report The Neuroform Atlas(®) stent is one of the most recently developed stents for coil embolization, with advancements in a lower-profile delivery system, enhanced trackability, smaller cell size, and increased wall conformability. Because of these advantages, the Neuroform Atlas(®) stent shows high technical success with few procedure-related complications. However, the present study reported a rare complication of a stretched and partially fractured Neuroform Atlas(®) stent due to unexpected partial withdrawal of microcatheter during deployment for coil embolization of an intracranial aneurysm. The measured length of the stent was ~30 mm, which was greater than the normal length (21 mm). An additional stent was inserted into the distal part of the deployed stent to stabilize the damaged stent and remodel the aneurysm neck. This complication was considered to potentially result from the combination of several factors, including: Curved vessel; open-cell stent; unexpected microcatheter withdrawal during stent deployment; and hooking of the aneurysm selecting microcatheter with stent strut. Understanding the stent design and careful manipulation while avoiding unexpected withdrawal of the microcatheter could prevent this complication. D.A. Spandidos 2023-03-22 /pmc/articles/PMC10119665/ /pubmed/37090084 http://dx.doi.org/10.3892/etm.2023.11906 Text en Copyright: © Cheol Ko et al. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs License (https://creativecommons.org/licenses/by-nc-nd/4.0/) , which permits use and distribution in any medium, provided the original work is properly cited, the use is non-commercial and no modifications or adaptations are made. |
spellingShingle | Case Report Ko, Hak Cheol Shin, Hee Sup Stretched and fractured Neuroform Atlas(®) stent during a stent‑assisted coil embolization: A case report |
title | Stretched and fractured Neuroform Atlas(®) stent during a stent‑assisted coil embolization: A case report |
title_full | Stretched and fractured Neuroform Atlas(®) stent during a stent‑assisted coil embolization: A case report |
title_fullStr | Stretched and fractured Neuroform Atlas(®) stent during a stent‑assisted coil embolization: A case report |
title_full_unstemmed | Stretched and fractured Neuroform Atlas(®) stent during a stent‑assisted coil embolization: A case report |
title_short | Stretched and fractured Neuroform Atlas(®) stent during a stent‑assisted coil embolization: A case report |
title_sort | stretched and fractured neuroform atlas(®) stent during a stent‑assisted coil embolization: a case report |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10119665/ https://www.ncbi.nlm.nih.gov/pubmed/37090084 http://dx.doi.org/10.3892/etm.2023.11906 |
work_keys_str_mv | AT kohakcheol stretchedandfracturedneuroformatlasstentduringastentassistedcoilembolizationacasereport AT shinheesup stretchedandfracturedneuroformatlasstentduringastentassistedcoilembolizationacasereport |