Metasurfaces of capacitively loaded metallic rings for magnetic resonance imaging surface coils
This work investigates the use of a metasurface made up of a two-dimensional array of capacitively loaded metallic rings to enhance the signal-to-noise ratio of magnetic resonance imaging surface coils and to tailor the magnetic near-field radio frequency pattern of the coils. It is found that the s...
Autor principal: | |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Nature Publishing Group UK
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10126161/ https://www.ncbi.nlm.nih.gov/pubmed/36810367 http://dx.doi.org/10.1038/s41598-023-30185-y |
_version_ | 1785030178989146112 |
---|---|
author | Freire, Manuel J. |
author_facet | Freire, Manuel J. |
author_sort | Freire, Manuel J. |
collection | PubMed |
description | This work investigates the use of a metasurface made up of a two-dimensional array of capacitively loaded metallic rings to enhance the signal-to-noise ratio of magnetic resonance imaging surface coils and to tailor the magnetic near-field radio frequency pattern of the coils. It is found that the signal-to-noise ratio is increased if the coupling between the capacitively loaded metallic rings in the array is increased. The input resistance and the radiofrequency magnetic field of the metasurface loaded coil are numerically analyzed by means of an efficient algorithm termed the discrete model to determine the signal-to-noise ratio. Standing surface waves or magnetoinductive waves supported by the metasurface introduce resonances in the frequency dependence of the input resistance. The signal-to-noise ratio is found to be optimal at the frequency corresponding to a local minimum existing between these resonances.The discrete model is used in an optimization procedure to fit the structural parameters of a metasurface to enhance the signal-to-noise ratio at the frequency corresponding to this local minimum in the input resistance. It is found that the signal-to-noise ratio can be greatly improved if the mutual coupling between the capacitively loaded metallic rings of the array is made stronger by bringing them closer or by using rings of squared shape instead of circular. These conclusions derived from the numerical results provided by the discrete model are double-checked by means of numerical simulations provided by the commercial electromagnetic solver Simulia CST and by experimental results. Numerical results provided by CST are also shown to demonstrate that the surface impedance of the array of elements can be adjusted to provide a more homogeneous magnetic near-field radio frequency pattern that ultimately leads to a more uniform magnetic resonance image at a desired slice. This is achieved by preventing the reflection of propagating magnetoinductive waves at the edges of the array by matching the elements arranged at the edges of the array with capacitors of suitable value. |
format | Online Article Text |
id | pubmed-10126161 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Nature Publishing Group UK |
record_format | MEDLINE/PubMed |
spelling | pubmed-101261612023-04-26 Metasurfaces of capacitively loaded metallic rings for magnetic resonance imaging surface coils Freire, Manuel J. Sci Rep Article This work investigates the use of a metasurface made up of a two-dimensional array of capacitively loaded metallic rings to enhance the signal-to-noise ratio of magnetic resonance imaging surface coils and to tailor the magnetic near-field radio frequency pattern of the coils. It is found that the signal-to-noise ratio is increased if the coupling between the capacitively loaded metallic rings in the array is increased. The input resistance and the radiofrequency magnetic field of the metasurface loaded coil are numerically analyzed by means of an efficient algorithm termed the discrete model to determine the signal-to-noise ratio. Standing surface waves or magnetoinductive waves supported by the metasurface introduce resonances in the frequency dependence of the input resistance. The signal-to-noise ratio is found to be optimal at the frequency corresponding to a local minimum existing between these resonances.The discrete model is used in an optimization procedure to fit the structural parameters of a metasurface to enhance the signal-to-noise ratio at the frequency corresponding to this local minimum in the input resistance. It is found that the signal-to-noise ratio can be greatly improved if the mutual coupling between the capacitively loaded metallic rings of the array is made stronger by bringing them closer or by using rings of squared shape instead of circular. These conclusions derived from the numerical results provided by the discrete model are double-checked by means of numerical simulations provided by the commercial electromagnetic solver Simulia CST and by experimental results. Numerical results provided by CST are also shown to demonstrate that the surface impedance of the array of elements can be adjusted to provide a more homogeneous magnetic near-field radio frequency pattern that ultimately leads to a more uniform magnetic resonance image at a desired slice. This is achieved by preventing the reflection of propagating magnetoinductive waves at the edges of the array by matching the elements arranged at the edges of the array with capacitors of suitable value. Nature Publishing Group UK 2023-02-21 /pmc/articles/PMC10126161/ /pubmed/36810367 http://dx.doi.org/10.1038/s41598-023-30185-y Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Article Freire, Manuel J. Metasurfaces of capacitively loaded metallic rings for magnetic resonance imaging surface coils |
title | Metasurfaces of capacitively loaded metallic rings for magnetic resonance imaging surface coils |
title_full | Metasurfaces of capacitively loaded metallic rings for magnetic resonance imaging surface coils |
title_fullStr | Metasurfaces of capacitively loaded metallic rings for magnetic resonance imaging surface coils |
title_full_unstemmed | Metasurfaces of capacitively loaded metallic rings for magnetic resonance imaging surface coils |
title_short | Metasurfaces of capacitively loaded metallic rings for magnetic resonance imaging surface coils |
title_sort | metasurfaces of capacitively loaded metallic rings for magnetic resonance imaging surface coils |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10126161/ https://www.ncbi.nlm.nih.gov/pubmed/36810367 http://dx.doi.org/10.1038/s41598-023-30185-y |
work_keys_str_mv | AT freiremanuelj metasurfacesofcapacitivelyloadedmetallicringsformagneticresonanceimagingsurfacecoils |