328 Rapid Screening Tool for Identifying Acute Myocardial Injury: An Exploratory Study Evaluating the Ability of a Novel, Noninvasive Device to Detect Cardiac Troponin

OBJECTIVES/GOALS: Tropsensor, a noninvasive portable device using infrared spectroscopy, delivers a troponin res ult within five minutes, significantly quicker than standard of care (SOC) assays. This pilot study assesses the correlation of the Tropsensor and high sensitivity cardiac troponin (hs-cT...

Descripción completa

Detalles Bibliográficos
Autores principales: McDonald, Sam, Katz, Jessie, Titus, Jitto, Biswal, Siddharth, Burman, Atandra, Agusala, Kartik, Diercks, Deborah, Vigen, Rebecca
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cambridge University Press 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10129733/
http://dx.doi.org/10.1017/cts.2023.376
_version_ 1785030815701270528
author McDonald, Sam
Katz, Jessie
Titus, Jitto
Biswal, Siddharth
Burman, Atandra
Agusala, Kartik
Diercks, Deborah
Vigen, Rebecca
author_facet McDonald, Sam
Katz, Jessie
Titus, Jitto
Biswal, Siddharth
Burman, Atandra
Agusala, Kartik
Diercks, Deborah
Vigen, Rebecca
author_sort McDonald, Sam
collection PubMed
description OBJECTIVES/GOALS: Tropsensor, a noninvasive portable device using infrared spectroscopy, delivers a troponin res ult within five minutes, significantly quicker than standard of care (SOC) assays. This pilot study assesses the correlation of the Tropsensor and high sensitivity cardiac troponin (hs-cTnI) assay results. METHODS/STUDY POPULATION: Patients undergoing cTnI testing with the Abbott Architect STAT (Abbott Laboratories) hs-cTnI assay were recruited at a quaternary-care emergency department (ED). The Tropsensor was applied to the underside of the patients’ wrist within 5 minutes of the SOC blood draw for 5 minutes. The results of the hs-cTnI assays were compared with the raw output of the Tropsensor device to assess the relationship using both Spearman’s and nonlinear logarithmic measures of correlation. Patient demographic data was extracted from the EHR to supplement the data collected for this study. RESULTS/ANTICIPATED RESULTS: 58 patients were recruited with a mean age of 60 years (60% male, 40% female). Due to connection error, 8 patients’data did not get captured by the device. Additionally, due to noise related to suboptimal device contact with the wrist, 24 patients’data (41%) were rejected. Of the 26 patients with usable data, 9 patients had a troponin above the 99th percentile threshold. A nonlinear correlation of 0.64 and Spearman’s rho of 0.59 were observed between the SOC hs-cTnI assay and Tropsensor optical data. DISCUSSION/SIGNIFICANCE: The Tropsensor exhibits 64.4% correlation to the SOC hs-cTnI assay. While further evaluation is needed, this exploratory study provides insight into the potential of a transdermal optical device to be used as a screening tool for AMI in an ED triage situation.
format Online
Article
Text
id pubmed-10129733
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Cambridge University Press
record_format MEDLINE/PubMed
spelling pubmed-101297332023-04-26 328 Rapid Screening Tool for Identifying Acute Myocardial Injury: An Exploratory Study Evaluating the Ability of a Novel, Noninvasive Device to Detect Cardiac Troponin McDonald, Sam Katz, Jessie Titus, Jitto Biswal, Siddharth Burman, Atandra Agusala, Kartik Diercks, Deborah Vigen, Rebecca J Clin Transl Sci Precision Medicine/Health OBJECTIVES/GOALS: Tropsensor, a noninvasive portable device using infrared spectroscopy, delivers a troponin res ult within five minutes, significantly quicker than standard of care (SOC) assays. This pilot study assesses the correlation of the Tropsensor and high sensitivity cardiac troponin (hs-cTnI) assay results. METHODS/STUDY POPULATION: Patients undergoing cTnI testing with the Abbott Architect STAT (Abbott Laboratories) hs-cTnI assay were recruited at a quaternary-care emergency department (ED). The Tropsensor was applied to the underside of the patients’ wrist within 5 minutes of the SOC blood draw for 5 minutes. The results of the hs-cTnI assays were compared with the raw output of the Tropsensor device to assess the relationship using both Spearman’s and nonlinear logarithmic measures of correlation. Patient demographic data was extracted from the EHR to supplement the data collected for this study. RESULTS/ANTICIPATED RESULTS: 58 patients were recruited with a mean age of 60 years (60% male, 40% female). Due to connection error, 8 patients’data did not get captured by the device. Additionally, due to noise related to suboptimal device contact with the wrist, 24 patients’data (41%) were rejected. Of the 26 patients with usable data, 9 patients had a troponin above the 99th percentile threshold. A nonlinear correlation of 0.64 and Spearman’s rho of 0.59 were observed between the SOC hs-cTnI assay and Tropsensor optical data. DISCUSSION/SIGNIFICANCE: The Tropsensor exhibits 64.4% correlation to the SOC hs-cTnI assay. While further evaluation is needed, this exploratory study provides insight into the potential of a transdermal optical device to be used as a screening tool for AMI in an ED triage situation. Cambridge University Press 2023-04-24 /pmc/articles/PMC10129733/ http://dx.doi.org/10.1017/cts.2023.376 Text en © The Association for Clinical and Translational Science 2023 https://creativecommons.org/licenses/by-nc-nd/4.0/This is an Open Access article, distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives licence (https://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is unaltered and is properly cited. The written permission of Cambridge University Press must be obtained for commercial re-use or in order to create a derivative work.
spellingShingle Precision Medicine/Health
McDonald, Sam
Katz, Jessie
Titus, Jitto
Biswal, Siddharth
Burman, Atandra
Agusala, Kartik
Diercks, Deborah
Vigen, Rebecca
328 Rapid Screening Tool for Identifying Acute Myocardial Injury: An Exploratory Study Evaluating the Ability of a Novel, Noninvasive Device to Detect Cardiac Troponin
title 328 Rapid Screening Tool for Identifying Acute Myocardial Injury: An Exploratory Study Evaluating the Ability of a Novel, Noninvasive Device to Detect Cardiac Troponin
title_full 328 Rapid Screening Tool for Identifying Acute Myocardial Injury: An Exploratory Study Evaluating the Ability of a Novel, Noninvasive Device to Detect Cardiac Troponin
title_fullStr 328 Rapid Screening Tool for Identifying Acute Myocardial Injury: An Exploratory Study Evaluating the Ability of a Novel, Noninvasive Device to Detect Cardiac Troponin
title_full_unstemmed 328 Rapid Screening Tool for Identifying Acute Myocardial Injury: An Exploratory Study Evaluating the Ability of a Novel, Noninvasive Device to Detect Cardiac Troponin
title_short 328 Rapid Screening Tool for Identifying Acute Myocardial Injury: An Exploratory Study Evaluating the Ability of a Novel, Noninvasive Device to Detect Cardiac Troponin
title_sort 328 rapid screening tool for identifying acute myocardial injury: an exploratory study evaluating the ability of a novel, noninvasive device to detect cardiac troponin
topic Precision Medicine/Health
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10129733/
http://dx.doi.org/10.1017/cts.2023.376
work_keys_str_mv AT mcdonaldsam 328rapidscreeningtoolforidentifyingacutemyocardialinjuryanexploratorystudyevaluatingtheabilityofanovelnoninvasivedevicetodetectcardiactroponin
AT katzjessie 328rapidscreeningtoolforidentifyingacutemyocardialinjuryanexploratorystudyevaluatingtheabilityofanovelnoninvasivedevicetodetectcardiactroponin
AT titusjitto 328rapidscreeningtoolforidentifyingacutemyocardialinjuryanexploratorystudyevaluatingtheabilityofanovelnoninvasivedevicetodetectcardiactroponin
AT biswalsiddharth 328rapidscreeningtoolforidentifyingacutemyocardialinjuryanexploratorystudyevaluatingtheabilityofanovelnoninvasivedevicetodetectcardiactroponin
AT burmanatandra 328rapidscreeningtoolforidentifyingacutemyocardialinjuryanexploratorystudyevaluatingtheabilityofanovelnoninvasivedevicetodetectcardiactroponin
AT agusalakartik 328rapidscreeningtoolforidentifyingacutemyocardialinjuryanexploratorystudyevaluatingtheabilityofanovelnoninvasivedevicetodetectcardiactroponin
AT diercksdeborah 328rapidscreeningtoolforidentifyingacutemyocardialinjuryanexploratorystudyevaluatingtheabilityofanovelnoninvasivedevicetodetectcardiactroponin
AT vigenrebecca 328rapidscreeningtoolforidentifyingacutemyocardialinjuryanexploratorystudyevaluatingtheabilityofanovelnoninvasivedevicetodetectcardiactroponin