Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT

BACKGROUND: In this study, stent appearance in a novel silicon-based photon-counting computed tomography (Si-PCCT) prototype was compared with a conventional energy-integrating detector CT (EIDCT) system. METHODS: An ex vivo phantom was created, consisting of a 2% agar-water mixture, in which human-...

Descripción completa

Detalles Bibliográficos
Autores principales: Verelst, Emma, Buls, Nico, De Mey, Johan, Nieboer, Koenraad Hans, Vandenbergh, Frans, Crotty, Dominic, Deak, Paul, Sundvall, Albert, Holmin, Staffan, De Smet, Aron, Provyn, Steven, Van Gompel, Gert
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Vienna 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10130245/
https://www.ncbi.nlm.nih.gov/pubmed/37097376
http://dx.doi.org/10.1186/s41747-023-00333-0
_version_ 1785030924426018816
author Verelst, Emma
Buls, Nico
De Mey, Johan
Nieboer, Koenraad Hans
Vandenbergh, Frans
Crotty, Dominic
Deak, Paul
Sundvall, Albert
Holmin, Staffan
De Smet, Aron
Provyn, Steven
Van Gompel, Gert
author_facet Verelst, Emma
Buls, Nico
De Mey, Johan
Nieboer, Koenraad Hans
Vandenbergh, Frans
Crotty, Dominic
Deak, Paul
Sundvall, Albert
Holmin, Staffan
De Smet, Aron
Provyn, Steven
Van Gompel, Gert
author_sort Verelst, Emma
collection PubMed
description BACKGROUND: In this study, stent appearance in a novel silicon-based photon-counting computed tomography (Si-PCCT) prototype was compared with a conventional energy-integrating detector CT (EIDCT) system. METHODS: An ex vivo phantom was created, consisting of a 2% agar-water mixture, in which human-resected and stented arteries were individually embedded. Using similar technique parameters, helical scan data was acquired using a novel prototype Si-PCCT and a conventional EIDCT system at a volumetric CT dose index (CTDI(vol)) of 9 mGy. Reconstructions were made at 50(2) and 150(2) mm(2) field-of-views (FOVs) using a bone kernel and adaptive statistical iterative reconstruction with 0% blending. Using a 5-point Likert scale, reader evaluations were performed on stent appearance, blooming and inter-stent visibility. Quantitative image analysis was performed on stent diameter accuracy, blooming and inter-stent distinction. Qualitative and quantitative differences between Si-PCCT and EIDCT systems were tested with a Wilcoxon signed-rank test and a paired samples t-test, respectively. Inter- and intra-reader agreement was assessed using the intraclass correlation coefficient (ICC). RESULTS: Qualitatively, Si-PCCT images were rated higher than EIDCT images at 150-mm FOV, based on stent appearance (p = 0.026) and blooming (p = 0.015), with a moderate inter- (ICC = 0.50) and intra-reader (ICC = 0.60) agreement. Quantitatively, Si-PCCT yielded more accurate diameter measurements (p = 0.001), reduced blooming (p < 0.001) and improved inter-stent distinction (p < 0.001). Similar trends were observed for the images reconstructed at 50-mm FOV. CONCLUSIONS: When compared to EIDCT, the improved spatial resolution of Si-PCCT yields enhanced stent appearance, more accurate diameter measurements, reduced blooming and improved inter-stent distinction. KEY POINTS: • This study evaluated stent appearance in a novel silicon-based photon-counting computed tomography (Si-PCCT) prototype. • Compared to standard CT, Si-PCCT resulted in more accurate stent diameter measurements. • Si-PCCT also reduced blooming artefacts and improved inter-stent visibility.
format Online
Article
Text
id pubmed-10130245
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Springer Vienna
record_format MEDLINE/PubMed
spelling pubmed-101302452023-04-27 Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT Verelst, Emma Buls, Nico De Mey, Johan Nieboer, Koenraad Hans Vandenbergh, Frans Crotty, Dominic Deak, Paul Sundvall, Albert Holmin, Staffan De Smet, Aron Provyn, Steven Van Gompel, Gert Eur Radiol Exp Original Article BACKGROUND: In this study, stent appearance in a novel silicon-based photon-counting computed tomography (Si-PCCT) prototype was compared with a conventional energy-integrating detector CT (EIDCT) system. METHODS: An ex vivo phantom was created, consisting of a 2% agar-water mixture, in which human-resected and stented arteries were individually embedded. Using similar technique parameters, helical scan data was acquired using a novel prototype Si-PCCT and a conventional EIDCT system at a volumetric CT dose index (CTDI(vol)) of 9 mGy. Reconstructions were made at 50(2) and 150(2) mm(2) field-of-views (FOVs) using a bone kernel and adaptive statistical iterative reconstruction with 0% blending. Using a 5-point Likert scale, reader evaluations were performed on stent appearance, blooming and inter-stent visibility. Quantitative image analysis was performed on stent diameter accuracy, blooming and inter-stent distinction. Qualitative and quantitative differences between Si-PCCT and EIDCT systems were tested with a Wilcoxon signed-rank test and a paired samples t-test, respectively. Inter- and intra-reader agreement was assessed using the intraclass correlation coefficient (ICC). RESULTS: Qualitatively, Si-PCCT images were rated higher than EIDCT images at 150-mm FOV, based on stent appearance (p = 0.026) and blooming (p = 0.015), with a moderate inter- (ICC = 0.50) and intra-reader (ICC = 0.60) agreement. Quantitatively, Si-PCCT yielded more accurate diameter measurements (p = 0.001), reduced blooming (p < 0.001) and improved inter-stent distinction (p < 0.001). Similar trends were observed for the images reconstructed at 50-mm FOV. CONCLUSIONS: When compared to EIDCT, the improved spatial resolution of Si-PCCT yields enhanced stent appearance, more accurate diameter measurements, reduced blooming and improved inter-stent distinction. KEY POINTS: • This study evaluated stent appearance in a novel silicon-based photon-counting computed tomography (Si-PCCT) prototype. • Compared to standard CT, Si-PCCT resulted in more accurate stent diameter measurements. • Si-PCCT also reduced blooming artefacts and improved inter-stent visibility. Springer Vienna 2023-04-25 /pmc/articles/PMC10130245/ /pubmed/37097376 http://dx.doi.org/10.1186/s41747-023-00333-0 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Article
Verelst, Emma
Buls, Nico
De Mey, Johan
Nieboer, Koenraad Hans
Vandenbergh, Frans
Crotty, Dominic
Deak, Paul
Sundvall, Albert
Holmin, Staffan
De Smet, Aron
Provyn, Steven
Van Gompel, Gert
Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT
title Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT
title_full Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT
title_fullStr Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT
title_full_unstemmed Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT
title_short Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT
title_sort stent appearance in a novel silicon-based photon-counting ct prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating ct
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10130245/
https://www.ncbi.nlm.nih.gov/pubmed/37097376
http://dx.doi.org/10.1186/s41747-023-00333-0
work_keys_str_mv AT verelstemma stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct
AT bulsnico stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct
AT demeyjohan stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct
AT nieboerkoenraadhans stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct
AT vandenberghfrans stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct
AT crottydominic stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct
AT deakpaul stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct
AT sundvallalbert stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct
AT holminstaffan stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct
AT desmetaron stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct
AT provynsteven stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct
AT vangompelgert stentappearanceinanovelsiliconbasedphotoncountingctprototypeexvivophantomstudyinheadtoheadcomparisonwithconventionalenergyintegratingct