Breakage of a re-activated PRECICE® nail: A case report

INTRODUCTION AND IMPORTANCE: Distraction osteogenesis using motorized nails have shown promising results. These high-priced nails are regularly removed following successful lengthening sessions. The internal lengthening mechanism inside such nails is presumably intact and perhaps capable of running...

Descripción completa

Detalles Bibliográficos
Autor principal: Alrabai, Hamza M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10130627/
https://www.ncbi.nlm.nih.gov/pubmed/37060767
http://dx.doi.org/10.1016/j.ijscr.2023.108182
_version_ 1785030998222700544
author Alrabai, Hamza M.
author_facet Alrabai, Hamza M.
author_sort Alrabai, Hamza M.
collection PubMed
description INTRODUCTION AND IMPORTANCE: Distraction osteogenesis using motorized nails have shown promising results. These high-priced nails are regularly removed following successful lengthening sessions. The internal lengthening mechanism inside such nails is presumably intact and perhaps capable of running additional lengthening. Motorized nails are relatively weaker than conventional nails. This case report presented a broken re-activated PRECICE nail which has not been previously reported. PRESENTATION OF CASE: A 21-year-old male patient with right short tibia. He has completed an uneventful lengthening session using tibial PRECICE® nail. A residual length discrepancy persisted despite 4 cm length gain. Attempt of re-lengthening was conducted utilizing the pre-implanted PRECICE® nail. The re-lengthening session run smoothly throughout distraction phase. Late in the consolidation phase, partial nail breakage occurred. The patient acknowledged intermittent gentle weightbearing against the physician's recommendations. The broken nail was removed on emergency basis. Cast immobilization with non-weightbearing continued until achievement of full consolidation clinically and radiographically. CLINICAL DISCUSSION: Most of commercially available lengthening motorized nails are considered less rigid compared to conventional nails and hence more susceptible to deformation and metal failure. However, nail breakage has been reported in both conventional and lengthening nails. Re-activation of PRECICE nail for additional lengthening session could carry higher risk of breakage. CONCLUSION: Possibility of nail breakage may be higher in re-activated PRECICE® nails relative to newly implanted PRECICE® nails. More studies are required to evaluate behavior of re-activated lengthening nails. Defective lengthening nails must not be re-activated. The patient compliance is essential for successful re-activation of lengthening nails.
format Online
Article
Text
id pubmed-10130627
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-101306272023-04-27 Breakage of a re-activated PRECICE® nail: A case report Alrabai, Hamza M. Int J Surg Case Rep Case Report INTRODUCTION AND IMPORTANCE: Distraction osteogenesis using motorized nails have shown promising results. These high-priced nails are regularly removed following successful lengthening sessions. The internal lengthening mechanism inside such nails is presumably intact and perhaps capable of running additional lengthening. Motorized nails are relatively weaker than conventional nails. This case report presented a broken re-activated PRECICE nail which has not been previously reported. PRESENTATION OF CASE: A 21-year-old male patient with right short tibia. He has completed an uneventful lengthening session using tibial PRECICE® nail. A residual length discrepancy persisted despite 4 cm length gain. Attempt of re-lengthening was conducted utilizing the pre-implanted PRECICE® nail. The re-lengthening session run smoothly throughout distraction phase. Late in the consolidation phase, partial nail breakage occurred. The patient acknowledged intermittent gentle weightbearing against the physician's recommendations. The broken nail was removed on emergency basis. Cast immobilization with non-weightbearing continued until achievement of full consolidation clinically and radiographically. CLINICAL DISCUSSION: Most of commercially available lengthening motorized nails are considered less rigid compared to conventional nails and hence more susceptible to deformation and metal failure. However, nail breakage has been reported in both conventional and lengthening nails. Re-activation of PRECICE nail for additional lengthening session could carry higher risk of breakage. CONCLUSION: Possibility of nail breakage may be higher in re-activated PRECICE® nails relative to newly implanted PRECICE® nails. More studies are required to evaluate behavior of re-activated lengthening nails. Defective lengthening nails must not be re-activated. The patient compliance is essential for successful re-activation of lengthening nails. Elsevier 2023-04-12 /pmc/articles/PMC10130627/ /pubmed/37060767 http://dx.doi.org/10.1016/j.ijscr.2023.108182 Text en © 2023 The Author(s) https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Case Report
Alrabai, Hamza M.
Breakage of a re-activated PRECICE® nail: A case report
title Breakage of a re-activated PRECICE® nail: A case report
title_full Breakage of a re-activated PRECICE® nail: A case report
title_fullStr Breakage of a re-activated PRECICE® nail: A case report
title_full_unstemmed Breakage of a re-activated PRECICE® nail: A case report
title_short Breakage of a re-activated PRECICE® nail: A case report
title_sort breakage of a re-activated precice® nail: a case report
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10130627/
https://www.ncbi.nlm.nih.gov/pubmed/37060767
http://dx.doi.org/10.1016/j.ijscr.2023.108182
work_keys_str_mv AT alrabaihamzam breakageofareactivatedprecicenailacasereport