Glucose Interference in Serum and Urine Samples with Various Creatinine Concentrations Measured by the Jaffe Kinetic Method

BACKGROUND: The effect of glucose interference on creatinine measurement by Jaffe kinetic method differs between serum and urine specimens. We investigated the effects of creatinine concentration and specimen dilution on glucose interference with urine creatinine measurement. METHODS: Leftover serum...

Descripción completa

Detalles Bibliográficos
Autores principales: Choosongsang, Pensiri, Bhornsrivathanyou, Naphatohn, Aiadsakun, Peechana, Choosongsang, Phattanapong, Bodhikul, Anucha, Yamsuwan, Yupawadee, Sriwimol, Wilaiwan, Soonthornpun, Supamai
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Communications and Publications Division (CPD) of the IFCC 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10131238/
https://www.ncbi.nlm.nih.gov/pubmed/37124654
_version_ 1785031134860541952
author Choosongsang, Pensiri
Bhornsrivathanyou, Naphatohn
Aiadsakun, Peechana
Choosongsang, Phattanapong
Bodhikul, Anucha
Yamsuwan, Yupawadee
Sriwimol, Wilaiwan
Soonthornpun, Supamai
author_facet Choosongsang, Pensiri
Bhornsrivathanyou, Naphatohn
Aiadsakun, Peechana
Choosongsang, Phattanapong
Bodhikul, Anucha
Yamsuwan, Yupawadee
Sriwimol, Wilaiwan
Soonthornpun, Supamai
author_sort Choosongsang, Pensiri
collection PubMed
description BACKGROUND: The effect of glucose interference on creatinine measurement by Jaffe kinetic method differs between serum and urine specimens. We investigated the effects of creatinine concentration and specimen dilution on glucose interference with urine creatinine measurement. METHODS: Leftover serum and urine specimens were collected and stored at -20°C until study. Serum specimens were mixed to make 5 glucose concentrations ranging from <5.6 to 27.8 mmol/L, each group consisting of 5 levels of creatinine concentration ranging from <45 to 354 μmol/L. Urine specimens were divided into 5 groups of creatinine concentration ranging from <1,769 to >7956 μmol/L, each sample was spiked with glucose powder to produce 5 aliquots with glucose concentrations ranging from 0 to 666 mmol/L. Urine samples were automatically diluted 1:20 before analysis. Percent interference of creatinine measurement by Jaffe kinetic method was calculated using enzymatic method as the reference. RESULTS: A total of 148 serum samples and 335 urine samples were analyzed. In serum, glucose interference with Jaffe creatinine measurement was found if creatinine concentrations were 177 μmol/L or less, corresponding to 3,540 μmol/L or less in urine specimens prior to 1:20 dilution. The degree of interference was greater when glucose concentration was higher or creatinine concentration was lower. CONCLUSIONS: When creatinine concentration and specimen dilution were considered, the effects of glucose interference on Jaffe creatinine measurement were similar in serum and urine specimens, and was found when creatinine concentrations in serum or diluted urine were 177 μmol/L or less.
format Online
Article
Text
id pubmed-10131238
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher The Communications and Publications Division (CPD) of the IFCC
record_format MEDLINE/PubMed
spelling pubmed-101312382023-04-27 Glucose Interference in Serum and Urine Samples with Various Creatinine Concentrations Measured by the Jaffe Kinetic Method Choosongsang, Pensiri Bhornsrivathanyou, Naphatohn Aiadsakun, Peechana Choosongsang, Phattanapong Bodhikul, Anucha Yamsuwan, Yupawadee Sriwimol, Wilaiwan Soonthornpun, Supamai EJIFCC Research Article BACKGROUND: The effect of glucose interference on creatinine measurement by Jaffe kinetic method differs between serum and urine specimens. We investigated the effects of creatinine concentration and specimen dilution on glucose interference with urine creatinine measurement. METHODS: Leftover serum and urine specimens were collected and stored at -20°C until study. Serum specimens were mixed to make 5 glucose concentrations ranging from <5.6 to 27.8 mmol/L, each group consisting of 5 levels of creatinine concentration ranging from <45 to 354 μmol/L. Urine specimens were divided into 5 groups of creatinine concentration ranging from <1,769 to >7956 μmol/L, each sample was spiked with glucose powder to produce 5 aliquots with glucose concentrations ranging from 0 to 666 mmol/L. Urine samples were automatically diluted 1:20 before analysis. Percent interference of creatinine measurement by Jaffe kinetic method was calculated using enzymatic method as the reference. RESULTS: A total of 148 serum samples and 335 urine samples were analyzed. In serum, glucose interference with Jaffe creatinine measurement was found if creatinine concentrations were 177 μmol/L or less, corresponding to 3,540 μmol/L or less in urine specimens prior to 1:20 dilution. The degree of interference was greater when glucose concentration was higher or creatinine concentration was lower. CONCLUSIONS: When creatinine concentration and specimen dilution were considered, the effects of glucose interference on Jaffe creatinine measurement were similar in serum and urine specimens, and was found when creatinine concentrations in serum or diluted urine were 177 μmol/L or less. The Communications and Publications Division (CPD) of the IFCC 2023-04-18 /pmc/articles/PMC10131238/ /pubmed/37124654 Text en Copyright © 2023 International Federation of Clinical Chemistry and Laboratory Medicine (IFCC). All rights reserved. https://creativecommons.org/licenses/by-nc/4.0/This is a Platinum Open Access Journal distributed under the terms of the Creative Commons Attribution Non-Commercial License which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Choosongsang, Pensiri
Bhornsrivathanyou, Naphatohn
Aiadsakun, Peechana
Choosongsang, Phattanapong
Bodhikul, Anucha
Yamsuwan, Yupawadee
Sriwimol, Wilaiwan
Soonthornpun, Supamai
Glucose Interference in Serum and Urine Samples with Various Creatinine Concentrations Measured by the Jaffe Kinetic Method
title Glucose Interference in Serum and Urine Samples with Various Creatinine Concentrations Measured by the Jaffe Kinetic Method
title_full Glucose Interference in Serum and Urine Samples with Various Creatinine Concentrations Measured by the Jaffe Kinetic Method
title_fullStr Glucose Interference in Serum and Urine Samples with Various Creatinine Concentrations Measured by the Jaffe Kinetic Method
title_full_unstemmed Glucose Interference in Serum and Urine Samples with Various Creatinine Concentrations Measured by the Jaffe Kinetic Method
title_short Glucose Interference in Serum and Urine Samples with Various Creatinine Concentrations Measured by the Jaffe Kinetic Method
title_sort glucose interference in serum and urine samples with various creatinine concentrations measured by the jaffe kinetic method
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10131238/
https://www.ncbi.nlm.nih.gov/pubmed/37124654
work_keys_str_mv AT choosongsangpensiri glucoseinterferenceinserumandurinesampleswithvariouscreatinineconcentrationsmeasuredbythejaffekineticmethod
AT bhornsrivathanyounaphatohn glucoseinterferenceinserumandurinesampleswithvariouscreatinineconcentrationsmeasuredbythejaffekineticmethod
AT aiadsakunpeechana glucoseinterferenceinserumandurinesampleswithvariouscreatinineconcentrationsmeasuredbythejaffekineticmethod
AT choosongsangphattanapong glucoseinterferenceinserumandurinesampleswithvariouscreatinineconcentrationsmeasuredbythejaffekineticmethod
AT bodhikulanucha glucoseinterferenceinserumandurinesampleswithvariouscreatinineconcentrationsmeasuredbythejaffekineticmethod
AT yamsuwanyupawadee glucoseinterferenceinserumandurinesampleswithvariouscreatinineconcentrationsmeasuredbythejaffekineticmethod
AT sriwimolwilaiwan glucoseinterferenceinserumandurinesampleswithvariouscreatinineconcentrationsmeasuredbythejaffekineticmethod
AT soonthornpunsupamai glucoseinterferenceinserumandurinesampleswithvariouscreatinineconcentrationsmeasuredbythejaffekineticmethod