High Fidelity Pressure Wires Provide Accurate Validation of Non-Invasive Central Blood Pressure and Pulse Wave Velocity Measurements

Central blood pressure (cBP) is known to be a better predictor of the damage caused by hypertension in comparison with peripheral blood pressure. During cardiac catheterization, we measured cBP in the ascending aorta with a fluid-filled guiding catheter (FF) in 75 patients and with a high-fidelity m...

Descripción completa

Detalles Bibliográficos
Autores principales: Scalia, Alessandro, Ghafari, Chadi, Navarre, Wivine, Delmotte, Philippe, Phillips, Rob, Carlier, Stéphane
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10135723/
https://www.ncbi.nlm.nih.gov/pubmed/37189852
http://dx.doi.org/10.3390/biomedicines11041235
_version_ 1785032047340814336
author Scalia, Alessandro
Ghafari, Chadi
Navarre, Wivine
Delmotte, Philippe
Phillips, Rob
Carlier, Stéphane
author_facet Scalia, Alessandro
Ghafari, Chadi
Navarre, Wivine
Delmotte, Philippe
Phillips, Rob
Carlier, Stéphane
author_sort Scalia, Alessandro
collection PubMed
description Central blood pressure (cBP) is known to be a better predictor of the damage caused by hypertension in comparison with peripheral blood pressure. During cardiac catheterization, we measured cBP in the ascending aorta with a fluid-filled guiding catheter (FF) in 75 patients and with a high-fidelity micromanometer tipped wire (FFR) in 20 patients. The wire was withdrawn into the brachial artery and aorto-brachial pulse wave velocity (abPWV) was calculated from the length of the pullback and the time delay between the ascending aorta and the brachial artery pulse waves by gating to the R-wave of the ECG for both measurements. In 23 patients, a cuff was inflated around the calf and an aorta-tibial pulse wave velocity (atPWV) was calculated from the distance between the cuff around the leg and the axillary notch and the time delay between the ascending aorta and the tibial pulse waves. Brachial BP was measured non-invasively and cBP was estimated using a new suprasystolic oscillometric technology. The mean differences between invasively measured cBP by FFR and non-invasive estimation were −0.4 ± 5.7 mmHg and by FF 5.4 ± 9.4 mmHg in 52 patients. Diastolic and mean cBP were both overestimated by oscillometry, with mean differences of −8.9 ± 5.5 mmHg and −6.4 ± 5.1 mmHg compared with the FFR and −10.6 ± 6.3 mmHg and −5.9 ± 6.2 mmHg with the FF. Non-invasive systolic cBP compared accurately with the high-fidelity FFR measurements, demonstrating a low bias (≤5 mmHg) and high precision (SD ≤ 8 mmHg). These criteria were not met when using the FF measurements. Invasively derived average Ao-brachial abPWV was 7.0 ± 1.4 m/s and that of Ao-tibial atPWV was 9.1 ± 1.8 m/s. Non-invasively estimated PWV based on the reflected wave transit time did not correlate with abPWV or with atPWV. In conclusion, we demonstrate the advantages of a novel method of validation for non-invasive cBP monitoring devices using acknowledged gold standard FFR wire transducers and the possibility to easily measure PWV during coronary angiography with the impact of cardiovascular risk factors.
format Online
Article
Text
id pubmed-10135723
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-101357232023-04-28 High Fidelity Pressure Wires Provide Accurate Validation of Non-Invasive Central Blood Pressure and Pulse Wave Velocity Measurements Scalia, Alessandro Ghafari, Chadi Navarre, Wivine Delmotte, Philippe Phillips, Rob Carlier, Stéphane Biomedicines Article Central blood pressure (cBP) is known to be a better predictor of the damage caused by hypertension in comparison with peripheral blood pressure. During cardiac catheterization, we measured cBP in the ascending aorta with a fluid-filled guiding catheter (FF) in 75 patients and with a high-fidelity micromanometer tipped wire (FFR) in 20 patients. The wire was withdrawn into the brachial artery and aorto-brachial pulse wave velocity (abPWV) was calculated from the length of the pullback and the time delay between the ascending aorta and the brachial artery pulse waves by gating to the R-wave of the ECG for both measurements. In 23 patients, a cuff was inflated around the calf and an aorta-tibial pulse wave velocity (atPWV) was calculated from the distance between the cuff around the leg and the axillary notch and the time delay between the ascending aorta and the tibial pulse waves. Brachial BP was measured non-invasively and cBP was estimated using a new suprasystolic oscillometric technology. The mean differences between invasively measured cBP by FFR and non-invasive estimation were −0.4 ± 5.7 mmHg and by FF 5.4 ± 9.4 mmHg in 52 patients. Diastolic and mean cBP were both overestimated by oscillometry, with mean differences of −8.9 ± 5.5 mmHg and −6.4 ± 5.1 mmHg compared with the FFR and −10.6 ± 6.3 mmHg and −5.9 ± 6.2 mmHg with the FF. Non-invasive systolic cBP compared accurately with the high-fidelity FFR measurements, demonstrating a low bias (≤5 mmHg) and high precision (SD ≤ 8 mmHg). These criteria were not met when using the FF measurements. Invasively derived average Ao-brachial abPWV was 7.0 ± 1.4 m/s and that of Ao-tibial atPWV was 9.1 ± 1.8 m/s. Non-invasively estimated PWV based on the reflected wave transit time did not correlate with abPWV or with atPWV. In conclusion, we demonstrate the advantages of a novel method of validation for non-invasive cBP monitoring devices using acknowledged gold standard FFR wire transducers and the possibility to easily measure PWV during coronary angiography with the impact of cardiovascular risk factors. MDPI 2023-04-21 /pmc/articles/PMC10135723/ /pubmed/37189852 http://dx.doi.org/10.3390/biomedicines11041235 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Scalia, Alessandro
Ghafari, Chadi
Navarre, Wivine
Delmotte, Philippe
Phillips, Rob
Carlier, Stéphane
High Fidelity Pressure Wires Provide Accurate Validation of Non-Invasive Central Blood Pressure and Pulse Wave Velocity Measurements
title High Fidelity Pressure Wires Provide Accurate Validation of Non-Invasive Central Blood Pressure and Pulse Wave Velocity Measurements
title_full High Fidelity Pressure Wires Provide Accurate Validation of Non-Invasive Central Blood Pressure and Pulse Wave Velocity Measurements
title_fullStr High Fidelity Pressure Wires Provide Accurate Validation of Non-Invasive Central Blood Pressure and Pulse Wave Velocity Measurements
title_full_unstemmed High Fidelity Pressure Wires Provide Accurate Validation of Non-Invasive Central Blood Pressure and Pulse Wave Velocity Measurements
title_short High Fidelity Pressure Wires Provide Accurate Validation of Non-Invasive Central Blood Pressure and Pulse Wave Velocity Measurements
title_sort high fidelity pressure wires provide accurate validation of non-invasive central blood pressure and pulse wave velocity measurements
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10135723/
https://www.ncbi.nlm.nih.gov/pubmed/37189852
http://dx.doi.org/10.3390/biomedicines11041235
work_keys_str_mv AT scaliaalessandro highfidelitypressurewiresprovideaccuratevalidationofnoninvasivecentralbloodpressureandpulsewavevelocitymeasurements
AT ghafarichadi highfidelitypressurewiresprovideaccuratevalidationofnoninvasivecentralbloodpressureandpulsewavevelocitymeasurements
AT navarrewivine highfidelitypressurewiresprovideaccuratevalidationofnoninvasivecentralbloodpressureandpulsewavevelocitymeasurements
AT delmottephilippe highfidelitypressurewiresprovideaccuratevalidationofnoninvasivecentralbloodpressureandpulsewavevelocitymeasurements
AT phillipsrob highfidelitypressurewiresprovideaccuratevalidationofnoninvasivecentralbloodpressureandpulsewavevelocitymeasurements
AT carlierstephane highfidelitypressurewiresprovideaccuratevalidationofnoninvasivecentralbloodpressureandpulsewavevelocitymeasurements