Cargando…

An Electrochemical o-Phthalaldehyde Sensor Using a Modified Disposable Screen-Printed Electrode with Polyacrylate Hydrogel for Concentration Verification of Clinical Disinfectant

The study proposes an o-phthalaldehyde (OPA) sensor for rapid and reliable detection of OPA in healthcare disinfection practices, based on a hydrogel-modified screen-printed carbon electrode strip. The hydrogel film, which contains glycine and N-acetylcysteine, reacts with OPA to produce a reductive...

Descripción completa

Detalles Bibliográficos
Autores principales: Chen, Richie L. C., Hsieh, Bo-Chuan, Lin, Jia-Sin, Cheng, Tzong-Jih
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10136433/
https://www.ncbi.nlm.nih.gov/pubmed/37185560
http://dx.doi.org/10.3390/bios13040485
Descripción
Sumario:The study proposes an o-phthalaldehyde (OPA) sensor for rapid and reliable detection of OPA in healthcare disinfection practices, based on a hydrogel-modified screen-printed carbon electrode strip. The hydrogel film, which contains glycine and N-acetylcysteine, reacts with OPA to produce a reductive isoindole derivative. The derivative is then oxidized for OPA determination using cyclic voltammetry. The proposed sensor achieves an optimal detection time of 20–30 s and requires only a small analyte volume of 5 µL. It exhibits good precision (10%) and sensitivity (3.3 μA/cm(2) mM) in a phosphate-buffered solution (pH 7.6), with excellent linearity (R(2) > 0.97) and precision (<3%) in the detection range (0.2–0.6%) required for clinical OPA solutions. Moreover, the sensor demonstrates good concentration verification of Cidex-OPA disinfection in healthcare institutes, with high sensitivity (18.28 μA/cm(2) mM) and precision around the minimum effective concentration (0.3%). Overall, the proposed sensor offers a promising and practical solution for accurate and reliable OPA detection in clinical disinfection practices.