Robot-Assisted Epiretinal Membrane Peeling: A Prospective Assessment of Pre- and Intra-Operative Times and of Surgeons’ Subjective Perceptions

Purpose: The Preceyes Surgical System (PSS) is a robotic assistive device that may enhance surgical precision. This study assessed pre- and intra-operative times and surgeons’ perceptions of robot-assisted epiretinal membrane peeling (RA-MP). Methods: We analyzed the time requirement of three main t...

Descripción completa

Detalles Bibliográficos
Autores principales: Turgut, Ferhat, Somfai, Gábor Márk, Heussen, Florian M., Eberle, Alexander, de Smet, Marc D., Becker, Matthias D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10141075/
https://www.ncbi.nlm.nih.gov/pubmed/37109104
http://dx.doi.org/10.3390/jcm12082768
Descripción
Sumario:Purpose: The Preceyes Surgical System (PSS) is a robotic assistive device that may enhance surgical precision. This study assessed pre- and intra-operative times and surgeons’ perceptions of robot-assisted epiretinal membrane peeling (RA-MP). Methods: We analyzed the time requirement of three main tasks: the preparation of the PSS (I), patient preparation (II), and surgery (III). Following surgery, the surgeons were asked questions about their experience. Results: RA-MP was performed in nine eyes of nine patients. Task I required an average time of 12.3 min, initially taking 15 min but decreasing to 6 min in the last surgery. Task II showed a mean time of 47.2 (range of 36–65) min. Task III had a mean time of 72.4 (range of 57–100) min. A mean time of 27.9 (range of 9–46) min was necessary for RA-MP. The responses to the questionnaire revealed a trend towards increasing ease and reduced stress as familiarity with the PSS increased. Conclusions: A substantial reduction in pre- and intra-operative times, decreasing to a total of 115 min, was demonstrated. RA-MP was positively anticipated by the surgeons and led to no hand or arm strain while being more complex than manual MP.