Cargando…
The Complication of a Retained Broken Distal Interlocking Screw Within a Cephalomedullary Nail During Conversion Hip Arthroplasty: A Case Report
In patients who undergo femoral fracture fixation with a cephalomedullary nail, the breakage of one or more of the distal interlocking screws is a well-described phenomenon. The presence of a broken interlocking screw in patients who require the removal of their cephalomedullary nail presents a uniq...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Cureus
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10156128/ https://www.ncbi.nlm.nih.gov/pubmed/37153273 http://dx.doi.org/10.7759/cureus.37075 |
_version_ | 1785036475698511872 |
---|---|
author | Panchal, Olivia Schultz, Matthew J Lutz, Rex W Smith, Eric B Deirmengian, Gregory K |
author_facet | Panchal, Olivia Schultz, Matthew J Lutz, Rex W Smith, Eric B Deirmengian, Gregory K |
author_sort | Panchal, Olivia |
collection | PubMed |
description | In patients who undergo femoral fracture fixation with a cephalomedullary nail, the breakage of one or more of the distal interlocking screws is a well-described phenomenon. The presence of a broken interlocking screw in patients who require the removal of their cephalomedullary nail presents a unique challenge. The broken interlocking screw may be retrieved, or the screw may be retained if it is not engaged within the nail and the nail can safely be removed while leaving the broken screw fragment behind. We report a hip conversion arthroplasty case with a broken interlocking screw where the nail was removed with ease and the broken screw was assumed to have been left behind. Cerclage wires were placed for an apparent proximal femoral fracture. Postoperative X-rays demonstrated a large lucency tracking from the prior location of the distal interlocking screw to the calcar region. This finding made it evident that the broken screw had been retained in the nail and was dragged up the femur upon nail removal, causing a large gouge spanning the entire femur. |
format | Online Article Text |
id | pubmed-10156128 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Cureus |
record_format | MEDLINE/PubMed |
spelling | pubmed-101561282023-05-04 The Complication of a Retained Broken Distal Interlocking Screw Within a Cephalomedullary Nail During Conversion Hip Arthroplasty: A Case Report Panchal, Olivia Schultz, Matthew J Lutz, Rex W Smith, Eric B Deirmengian, Gregory K Cureus Orthopedics In patients who undergo femoral fracture fixation with a cephalomedullary nail, the breakage of one or more of the distal interlocking screws is a well-described phenomenon. The presence of a broken interlocking screw in patients who require the removal of their cephalomedullary nail presents a unique challenge. The broken interlocking screw may be retrieved, or the screw may be retained if it is not engaged within the nail and the nail can safely be removed while leaving the broken screw fragment behind. We report a hip conversion arthroplasty case with a broken interlocking screw where the nail was removed with ease and the broken screw was assumed to have been left behind. Cerclage wires were placed for an apparent proximal femoral fracture. Postoperative X-rays demonstrated a large lucency tracking from the prior location of the distal interlocking screw to the calcar region. This finding made it evident that the broken screw had been retained in the nail and was dragged up the femur upon nail removal, causing a large gouge spanning the entire femur. Cureus 2023-04-03 /pmc/articles/PMC10156128/ /pubmed/37153273 http://dx.doi.org/10.7759/cureus.37075 Text en Copyright © 2023, Panchal et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Orthopedics Panchal, Olivia Schultz, Matthew J Lutz, Rex W Smith, Eric B Deirmengian, Gregory K The Complication of a Retained Broken Distal Interlocking Screw Within a Cephalomedullary Nail During Conversion Hip Arthroplasty: A Case Report |
title | The Complication of a Retained Broken Distal Interlocking Screw Within a Cephalomedullary Nail During Conversion Hip Arthroplasty: A Case Report |
title_full | The Complication of a Retained Broken Distal Interlocking Screw Within a Cephalomedullary Nail During Conversion Hip Arthroplasty: A Case Report |
title_fullStr | The Complication of a Retained Broken Distal Interlocking Screw Within a Cephalomedullary Nail During Conversion Hip Arthroplasty: A Case Report |
title_full_unstemmed | The Complication of a Retained Broken Distal Interlocking Screw Within a Cephalomedullary Nail During Conversion Hip Arthroplasty: A Case Report |
title_short | The Complication of a Retained Broken Distal Interlocking Screw Within a Cephalomedullary Nail During Conversion Hip Arthroplasty: A Case Report |
title_sort | complication of a retained broken distal interlocking screw within a cephalomedullary nail during conversion hip arthroplasty: a case report |
topic | Orthopedics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10156128/ https://www.ncbi.nlm.nih.gov/pubmed/37153273 http://dx.doi.org/10.7759/cureus.37075 |
work_keys_str_mv | AT panchalolivia thecomplicationofaretainedbrokendistalinterlockingscrewwithinacephalomedullarynailduringconversionhiparthroplastyacasereport AT schultzmatthewj thecomplicationofaretainedbrokendistalinterlockingscrewwithinacephalomedullarynailduringconversionhiparthroplastyacasereport AT lutzrexw thecomplicationofaretainedbrokendistalinterlockingscrewwithinacephalomedullarynailduringconversionhiparthroplastyacasereport AT smithericb thecomplicationofaretainedbrokendistalinterlockingscrewwithinacephalomedullarynailduringconversionhiparthroplastyacasereport AT deirmengiangregoryk thecomplicationofaretainedbrokendistalinterlockingscrewwithinacephalomedullarynailduringconversionhiparthroplastyacasereport AT panchalolivia complicationofaretainedbrokendistalinterlockingscrewwithinacephalomedullarynailduringconversionhiparthroplastyacasereport AT schultzmatthewj complicationofaretainedbrokendistalinterlockingscrewwithinacephalomedullarynailduringconversionhiparthroplastyacasereport AT lutzrexw complicationofaretainedbrokendistalinterlockingscrewwithinacephalomedullarynailduringconversionhiparthroplastyacasereport AT smithericb complicationofaretainedbrokendistalinterlockingscrewwithinacephalomedullarynailduringconversionhiparthroplastyacasereport AT deirmengiangregoryk complicationofaretainedbrokendistalinterlockingscrewwithinacephalomedullarynailduringconversionhiparthroplastyacasereport |