Blood transfusion decisions cannot be based on percutaneous haemoglobin measurements
BACKGROUND: Percutaneous measurement of haemoglobin (SpHb) has been an emerging technique in the past decades. It has been suggested that SpHb measurement could be used as a trend monitor and as a supportive tool for easier and faster transfusion decision-making. The aim of this study was to investi...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Termedia Publishing House
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10158438/ https://www.ncbi.nlm.nih.gov/pubmed/34006050 http://dx.doi.org/10.5114/ait.2021.105658 |
_version_ | 1785036929371209728 |
---|---|
author | Spronk, Angela de Jong, Renate Koopman, Seppe Hoeks, Sanne Stolker, Robert Jan |
author_facet | Spronk, Angela de Jong, Renate Koopman, Seppe Hoeks, Sanne Stolker, Robert Jan |
author_sort | Spronk, Angela |
collection | PubMed |
description | BACKGROUND: Percutaneous measurement of haemoglobin (SpHb) has been an emerging technique in the past decades. It has been suggested that SpHb measurement could be used as a trend monitor and as a supportive tool for easier and faster transfusion decision-making. The aim of this study was to investigate whether SpHb monitoring is a useful instrument in transfusion decision-making. METHODS: Patients scheduled for surgery with expected blood loss over 800 mL were included in the study. SpHb was measured using a Masimo Rainbow probe. Blood samples were drawn before and after surgery and, if clinically indicated, during surgery. Moreover, perfusion parameters were analysed, as well as transfusion triggers. RESULTS: Based on transfusion triggers 27.1% of patients would not have been transfused according to National Guidelines (14.5% transfused in error, 12.5% not transfused when indicated). Invasive haemoglobin (invasive Hb) and SpHb were obtained 266 times in 75 patients. The mean invasive Hb was 7.37 ± 1.34 mmol L(–1) and SpHb was 6.47 ± 0.81 mmol L(–1) (P < 0.001). Bland-Altman analysis corrected for multiple measurements revealed proportional bias of –4.05 + 0.72 Hb (least bias at Hb 5.62) CONCLUSIONS: The precision of the SpHb measurement exceeded the acceptable range of error. We concluded that SpHb measurement using the Rainbow device is too unreliable to be an acceptable alternative to invasive Hb measurement, or even as a trend monitor or decision support tool. |
format | Online Article Text |
id | pubmed-10158438 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Termedia Publishing House |
record_format | MEDLINE/PubMed |
spelling | pubmed-101584382023-05-17 Blood transfusion decisions cannot be based on percutaneous haemoglobin measurements Spronk, Angela de Jong, Renate Koopman, Seppe Hoeks, Sanne Stolker, Robert Jan Anaesthesiol Intensive Ther Original and Clinical Articles BACKGROUND: Percutaneous measurement of haemoglobin (SpHb) has been an emerging technique in the past decades. It has been suggested that SpHb measurement could be used as a trend monitor and as a supportive tool for easier and faster transfusion decision-making. The aim of this study was to investigate whether SpHb monitoring is a useful instrument in transfusion decision-making. METHODS: Patients scheduled for surgery with expected blood loss over 800 mL were included in the study. SpHb was measured using a Masimo Rainbow probe. Blood samples were drawn before and after surgery and, if clinically indicated, during surgery. Moreover, perfusion parameters were analysed, as well as transfusion triggers. RESULTS: Based on transfusion triggers 27.1% of patients would not have been transfused according to National Guidelines (14.5% transfused in error, 12.5% not transfused when indicated). Invasive haemoglobin (invasive Hb) and SpHb were obtained 266 times in 75 patients. The mean invasive Hb was 7.37 ± 1.34 mmol L(–1) and SpHb was 6.47 ± 0.81 mmol L(–1) (P < 0.001). Bland-Altman analysis corrected for multiple measurements revealed proportional bias of –4.05 + 0.72 Hb (least bias at Hb 5.62) CONCLUSIONS: The precision of the SpHb measurement exceeded the acceptable range of error. We concluded that SpHb measurement using the Rainbow device is too unreliable to be an acceptable alternative to invasive Hb measurement, or even as a trend monitor or decision support tool. Termedia Publishing House 2021-05-05 2021-06 /pmc/articles/PMC10158438/ /pubmed/34006050 http://dx.doi.org/10.5114/ait.2021.105658 Text en Copyright © Polish Society of Anaesthesiology and Intensive Therapy https://creativecommons.org/licenses/by-nc-sa/4.0/This is an Open Access journal, all articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/ (https://creativecommons.org/licenses/by-nc-sa/4.0/) ), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license. |
spellingShingle | Original and Clinical Articles Spronk, Angela de Jong, Renate Koopman, Seppe Hoeks, Sanne Stolker, Robert Jan Blood transfusion decisions cannot be based on percutaneous haemoglobin measurements |
title | Blood transfusion decisions cannot be based on percutaneous haemoglobin measurements |
title_full | Blood transfusion decisions cannot be based on percutaneous haemoglobin measurements |
title_fullStr | Blood transfusion decisions cannot be based on percutaneous haemoglobin measurements |
title_full_unstemmed | Blood transfusion decisions cannot be based on percutaneous haemoglobin measurements |
title_short | Blood transfusion decisions cannot be based on percutaneous haemoglobin measurements |
title_sort | blood transfusion decisions cannot be based on percutaneous haemoglobin measurements |
topic | Original and Clinical Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10158438/ https://www.ncbi.nlm.nih.gov/pubmed/34006050 http://dx.doi.org/10.5114/ait.2021.105658 |
work_keys_str_mv | AT spronkangela bloodtransfusiondecisionscannotbebasedonpercutaneoushaemoglobinmeasurements AT dejongrenate bloodtransfusiondecisionscannotbebasedonpercutaneoushaemoglobinmeasurements AT koopmanseppe bloodtransfusiondecisionscannotbebasedonpercutaneoushaemoglobinmeasurements AT hoekssanne bloodtransfusiondecisionscannotbebasedonpercutaneoushaemoglobinmeasurements AT stolkerrobertjan bloodtransfusiondecisionscannotbebasedonpercutaneoushaemoglobinmeasurements |