Technical note: Consistency of IAEA's TRS‐483 and AAPM's extended TG‐51 protocols for clinical reference dosimetry of the CyberKnife M6 machine

BACKGROUND: While IAEA's TRS‐483 code of practice is adapted for the calibration of CyberKnife machines, AAPM's TG‐51 is still the protocol recommended by the manufacturer for their calibration. The differences between both protocols could lead to differences in absorbed dose to water duri...

Descripción completa

Detalles Bibliográficos
Autores principales: Duchaine, Jasmine, Markel, Daniel, Ley, Jean‐Luc, Béliveau‐Nadeau, Dominique, Zerouali, Karim, Doucet, Robert, Bouchard, Hugo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10161123/
https://www.ncbi.nlm.nih.gov/pubmed/36995902
http://dx.doi.org/10.1002/acm2.13976
Descripción
Sumario:BACKGROUND: While IAEA's TRS‐483 code of practice is adapted for the calibration of CyberKnife machines, AAPM's TG‐51 is still the protocol recommended by the manufacturer for their calibration. The differences between both protocols could lead to differences in absorbed dose to water during the calibration process. PURPOSE: The aims of this work are to evaluate the difference resulting from the application of TG‐51 (including the manufacturer's adaptations) and TRS‐483 in terms of absorbed dose to water for a CyberKnife M6, and to evaluate the consistency of TRS‐483. METHODS: Measurements are performed on a CyberKnife M6 unit under machine‐specific reference conditions using a calibrated Exradin A12 ionization chamber. Monte Carlo (MC) simulations are performed to estimate [Formula: see text] and [Formula: see text] using a fully modeled detector and an optimized CyberKnife M6 beam model. The latter is also estimated experimentally. Differences between the adapted TG‐51 and TRS‐483 protocols are identified and their impact is quantified. RESULTS: When using an in‐house experimentally‐evaluated volume averaging correction factor, a difference of 0.11% in terms of absorbed dose to water per monitor unit is observed when applying both protocols. This disparity is solely associated to the difference in beam quality correction factor. If a generic volume averaging correction factor is used during the application of TRS‐483, the difference in calibration increases to 0.14%. In both cases, the disparity is not statistically significant according to TRS‐483's reported uncertainties on their beam quality correction factor (i.e., 1%). MC results lead to [Formula: see text] and [Formula: see text]. Results illustrate that the generic beam quality correction factor provided in the TRS‐483 might be overestimated by 0.36% compared to our specific model and that this overestimation could be due to the volume averaging component. CONCLUSIONS: For clinical reference dosimetry of the CyberKnife M6, the application of TRS‐483 is found to be consistent with TG‐51.