Cargando…
The use of a wearable cardioverter defibrillator during postmastectomy radiotherapy as an alternative for relocation of the pulse generator
PURPOSE: We describe a case in which a breast cancer patient with an implantable cardioverter-defibrillator (ICD) was treated with postmastectomy followed by axillary radiation therapy. The ICD was partly within the target volume and according to the treatment planning system would receive a maximum...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10164824/ https://www.ncbi.nlm.nih.gov/pubmed/37168315 http://dx.doi.org/10.1016/j.adro.2023.101225 |
Sumario: | PURPOSE: We describe a case in which a breast cancer patient with an implantable cardioverter-defibrillator (ICD) was treated with postmastectomy followed by axillary radiation therapy. The ICD was partly within the target volume and according to the treatment planning system would receive a maximum dose of about 36 Gy. METHODS AND MATERIALS: Postoperatively, the patient had a wound infection needing surgical intervention and 3 months of antibiotics. Therefore, it was decided not to relocate the ICD but instead to switch off the shock function of the ICD for the entire radiation therapy course. RESULTS: The ICD was successfully substituted with a wearable cardioverter defibrillator (WCD), which the patient wore during the 3-week radiation therapy period until 2 weeks after the end of treatment. The ICD function was monitored from the start of the treatment until 6 months after treatment. CONCLUSIONS: In situations in which an ICD needs to be switched off semipermanently, a WCD can be used as alternative. The patient in this report tolerated the WCD well. Despite the high dose received by the ICD, the device and its battery continued to function normally. |
---|