Clinical observation of tear film breakup time with a novel smartphone-attachable technology
PURPOSE: To demonstrate the practicability of a portable instrument in assessing tear film breakup time (TFBUT): a lens attachment for smartphones (LAS). METHODS: By applying LAS in combination with the iPhone 12 pro, and a recordable slit-lamp microscope, we obtained TFBUT videos from 58 volunteers...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10170453/ https://www.ncbi.nlm.nih.gov/pubmed/37165312 http://dx.doi.org/10.1186/s12886-023-02932-2 |
_version_ | 1785039230312906752 |
---|---|
author | Zhang, Xiaoran Shen, Jiawei Kang, Zhen Chen, Andrew Cao, Qian Xue, Chunyan |
author_facet | Zhang, Xiaoran Shen, Jiawei Kang, Zhen Chen, Andrew Cao, Qian Xue, Chunyan |
author_sort | Zhang, Xiaoran |
collection | PubMed |
description | PURPOSE: To demonstrate the practicability of a portable instrument in assessing tear film breakup time (TFBUT): a lens attachment for smartphones (LAS). METHODS: By applying LAS in combination with the iPhone 12 pro, and a recordable slit-lamp microscope, we obtained TFBUT videos from 58 volunteers. The comparison between the conventional slit-lamp microscope and LAS by an experienced ophthalmologist. Moreover, we also invited an ophthalmic postgraduate student and an undergraduate student with no clinical experience to assess TFBUT in those videos. The inter-observer reliability was assessed using intraclass correlation coefficients (ICC). RESULTS: The TFBUT of 116 eyes in total was recorded. Reliability indexes were adequate. The Spearman’s correlation analysis and the intraclass correlation coefficient suggested a strong correlation between the 2 modalities (Right eye: Spearman’s r = 0.929, 95% confidence interval (CI) = 0.847–0.963, ICC = 0.978, p < 0.001; Left eye: Spearman’s r = 0.931, 95% CI = 0.866–0.964, ICC = 0.985, p < 0.001;). Between instruments, the majority of TFBUT measurements showed good agreement on Bland Altman plot. A high concordance was observed in TFBUT, when assessed by an ophthalmologist and an ophthalmic postgraduate student (Left eye: LAS ICC = 0.951, p < 0.001; Left eye: slit-lamp microscope ICC = 0.944, p < 0.001). CONCLUSIONS: Compared with the conventional slit-lamp microscope, the LAS has sufficient validity for evaluating TFBUT in clinics or at home. |
format | Online Article Text |
id | pubmed-10170453 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-101704532023-05-11 Clinical observation of tear film breakup time with a novel smartphone-attachable technology Zhang, Xiaoran Shen, Jiawei Kang, Zhen Chen, Andrew Cao, Qian Xue, Chunyan BMC Ophthalmol Research PURPOSE: To demonstrate the practicability of a portable instrument in assessing tear film breakup time (TFBUT): a lens attachment for smartphones (LAS). METHODS: By applying LAS in combination with the iPhone 12 pro, and a recordable slit-lamp microscope, we obtained TFBUT videos from 58 volunteers. The comparison between the conventional slit-lamp microscope and LAS by an experienced ophthalmologist. Moreover, we also invited an ophthalmic postgraduate student and an undergraduate student with no clinical experience to assess TFBUT in those videos. The inter-observer reliability was assessed using intraclass correlation coefficients (ICC). RESULTS: The TFBUT of 116 eyes in total was recorded. Reliability indexes were adequate. The Spearman’s correlation analysis and the intraclass correlation coefficient suggested a strong correlation between the 2 modalities (Right eye: Spearman’s r = 0.929, 95% confidence interval (CI) = 0.847–0.963, ICC = 0.978, p < 0.001; Left eye: Spearman’s r = 0.931, 95% CI = 0.866–0.964, ICC = 0.985, p < 0.001;). Between instruments, the majority of TFBUT measurements showed good agreement on Bland Altman plot. A high concordance was observed in TFBUT, when assessed by an ophthalmologist and an ophthalmic postgraduate student (Left eye: LAS ICC = 0.951, p < 0.001; Left eye: slit-lamp microscope ICC = 0.944, p < 0.001). CONCLUSIONS: Compared with the conventional slit-lamp microscope, the LAS has sufficient validity for evaluating TFBUT in clinics or at home. BioMed Central 2023-05-10 /pmc/articles/PMC10170453/ /pubmed/37165312 http://dx.doi.org/10.1186/s12886-023-02932-2 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Zhang, Xiaoran Shen, Jiawei Kang, Zhen Chen, Andrew Cao, Qian Xue, Chunyan Clinical observation of tear film breakup time with a novel smartphone-attachable technology |
title | Clinical observation of tear film breakup time with a novel smartphone-attachable technology |
title_full | Clinical observation of tear film breakup time with a novel smartphone-attachable technology |
title_fullStr | Clinical observation of tear film breakup time with a novel smartphone-attachable technology |
title_full_unstemmed | Clinical observation of tear film breakup time with a novel smartphone-attachable technology |
title_short | Clinical observation of tear film breakup time with a novel smartphone-attachable technology |
title_sort | clinical observation of tear film breakup time with a novel smartphone-attachable technology |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10170453/ https://www.ncbi.nlm.nih.gov/pubmed/37165312 http://dx.doi.org/10.1186/s12886-023-02932-2 |
work_keys_str_mv | AT zhangxiaoran clinicalobservationoftearfilmbreakuptimewithanovelsmartphoneattachabletechnology AT shenjiawei clinicalobservationoftearfilmbreakuptimewithanovelsmartphoneattachabletechnology AT kangzhen clinicalobservationoftearfilmbreakuptimewithanovelsmartphoneattachabletechnology AT chenandrew clinicalobservationoftearfilmbreakuptimewithanovelsmartphoneattachabletechnology AT caoqian clinicalobservationoftearfilmbreakuptimewithanovelsmartphoneattachabletechnology AT xuechunyan clinicalobservationoftearfilmbreakuptimewithanovelsmartphoneattachabletechnology |