A case of two connected stents deployed during iStent inject W surgery
BACKGROUND: We report a case with two connected stents ejected simultaneously during an iStent inject W surgery, a modified second-generation iStent Trabecular Micro-Bypass System. CASE PRESENTATION: A 57-year-old woman with primary open-angle glaucoma underwent a combined cataract and iStent inject...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10170729/ https://www.ncbi.nlm.nih.gov/pubmed/37165322 http://dx.doi.org/10.1186/s12886-023-02951-z |
_version_ | 1785039283011190784 |
---|---|
author | Shimada, Ayaka Ichioka, Sho Ishida, Akiko Kaidzu, Sachiko Tanito, Masaki |
author_facet | Shimada, Ayaka Ichioka, Sho Ishida, Akiko Kaidzu, Sachiko Tanito, Masaki |
author_sort | Shimada, Ayaka |
collection | PubMed |
description | BACKGROUND: We report a case with two connected stents ejected simultaneously during an iStent inject W surgery, a modified second-generation iStent Trabecular Micro-Bypass System. CASE PRESENTATION: A 57-year-old woman with primary open-angle glaucoma underwent a combined cataract and iStent inject W surgery in her left eye. After the trabecular meshwork/Schlemm’s canal was pierced by the trocar of injector, the delivery button was pressed a first time, but the stent was not ejected. After the button was pressed a second time, connected two stents were ejected. After removing the dislocated stents from the anterior chamber, two stents were implanted into the desired places using another injector. Except for mild hyphema, no postoperative complication occurred. Stereomicroscopic observation showed that the two stents were connected by a broken trocar shaft. An X-ray showed that the trocar shaft was broken at the part referred to as the “sprayed trocar”. Scanning electron microscopy showed that the surface features of the broken trocar and trocar tip represented tensile failure. CONCLUSIONS: Although rare, considering that the damage was seen at the structurally weak part (i.e., sprayed trocar), the same phenomenon can happen. For patient safety, surgeons are recommended to inspect the device when the deployment of either the first or second stent is unsuccessful during the iStent inject surgery. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12886-023-02951-z. |
format | Online Article Text |
id | pubmed-10170729 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-101707292023-05-11 A case of two connected stents deployed during iStent inject W surgery Shimada, Ayaka Ichioka, Sho Ishida, Akiko Kaidzu, Sachiko Tanito, Masaki BMC Ophthalmol Case Report BACKGROUND: We report a case with two connected stents ejected simultaneously during an iStent inject W surgery, a modified second-generation iStent Trabecular Micro-Bypass System. CASE PRESENTATION: A 57-year-old woman with primary open-angle glaucoma underwent a combined cataract and iStent inject W surgery in her left eye. After the trabecular meshwork/Schlemm’s canal was pierced by the trocar of injector, the delivery button was pressed a first time, but the stent was not ejected. After the button was pressed a second time, connected two stents were ejected. After removing the dislocated stents from the anterior chamber, two stents were implanted into the desired places using another injector. Except for mild hyphema, no postoperative complication occurred. Stereomicroscopic observation showed that the two stents were connected by a broken trocar shaft. An X-ray showed that the trocar shaft was broken at the part referred to as the “sprayed trocar”. Scanning electron microscopy showed that the surface features of the broken trocar and trocar tip represented tensile failure. CONCLUSIONS: Although rare, considering that the damage was seen at the structurally weak part (i.e., sprayed trocar), the same phenomenon can happen. For patient safety, surgeons are recommended to inspect the device when the deployment of either the first or second stent is unsuccessful during the iStent inject surgery. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12886-023-02951-z. BioMed Central 2023-05-10 /pmc/articles/PMC10170729/ /pubmed/37165322 http://dx.doi.org/10.1186/s12886-023-02951-z Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Case Report Shimada, Ayaka Ichioka, Sho Ishida, Akiko Kaidzu, Sachiko Tanito, Masaki A case of two connected stents deployed during iStent inject W surgery |
title | A case of two connected stents deployed during iStent inject W surgery |
title_full | A case of two connected stents deployed during iStent inject W surgery |
title_fullStr | A case of two connected stents deployed during iStent inject W surgery |
title_full_unstemmed | A case of two connected stents deployed during iStent inject W surgery |
title_short | A case of two connected stents deployed during iStent inject W surgery |
title_sort | case of two connected stents deployed during istent inject w surgery |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10170729/ https://www.ncbi.nlm.nih.gov/pubmed/37165322 http://dx.doi.org/10.1186/s12886-023-02951-z |
work_keys_str_mv | AT shimadaayaka acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery AT ichiokasho acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery AT ishidaakiko acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery AT kaidzusachiko acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery AT tanitomasaki acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery AT shimadaayaka caseoftwoconnectedstentsdeployedduringistentinjectwsurgery AT ichiokasho caseoftwoconnectedstentsdeployedduringistentinjectwsurgery AT ishidaakiko caseoftwoconnectedstentsdeployedduringistentinjectwsurgery AT kaidzusachiko caseoftwoconnectedstentsdeployedduringistentinjectwsurgery AT tanitomasaki caseoftwoconnectedstentsdeployedduringistentinjectwsurgery |