A case of two connected stents deployed during iStent inject W surgery

BACKGROUND: We report a case with two connected stents ejected simultaneously during an iStent inject W surgery, a modified second-generation iStent Trabecular Micro-Bypass System. CASE PRESENTATION: A 57-year-old woman with primary open-angle glaucoma underwent a combined cataract and iStent inject...

Descripción completa

Detalles Bibliográficos
Autores principales: Shimada, Ayaka, Ichioka, Sho, Ishida, Akiko, Kaidzu, Sachiko, Tanito, Masaki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10170729/
https://www.ncbi.nlm.nih.gov/pubmed/37165322
http://dx.doi.org/10.1186/s12886-023-02951-z
_version_ 1785039283011190784
author Shimada, Ayaka
Ichioka, Sho
Ishida, Akiko
Kaidzu, Sachiko
Tanito, Masaki
author_facet Shimada, Ayaka
Ichioka, Sho
Ishida, Akiko
Kaidzu, Sachiko
Tanito, Masaki
author_sort Shimada, Ayaka
collection PubMed
description BACKGROUND: We report a case with two connected stents ejected simultaneously during an iStent inject W surgery, a modified second-generation iStent Trabecular Micro-Bypass System. CASE PRESENTATION: A 57-year-old woman with primary open-angle glaucoma underwent a combined cataract and iStent inject W surgery in her left eye. After the trabecular meshwork/Schlemm’s canal was pierced by the trocar of injector, the delivery button was pressed a first time, but the stent was not ejected. After the button was pressed a second time, connected two stents were ejected. After removing the dislocated stents from the anterior chamber, two stents were implanted into the desired places using another injector. Except for mild hyphema, no postoperative complication occurred. Stereomicroscopic observation showed that the two stents were connected by a broken trocar shaft. An X-ray showed that the trocar shaft was broken at the part referred to as the “sprayed trocar”. Scanning electron microscopy showed that the surface features of the broken trocar and trocar tip represented tensile failure. CONCLUSIONS: Although rare, considering that the damage was seen at the structurally weak part (i.e., sprayed trocar), the same phenomenon can happen. For patient safety, surgeons are recommended to inspect the device when the deployment of either the first or second stent is unsuccessful during the iStent inject surgery. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12886-023-02951-z.
format Online
Article
Text
id pubmed-10170729
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-101707292023-05-11 A case of two connected stents deployed during iStent inject W surgery Shimada, Ayaka Ichioka, Sho Ishida, Akiko Kaidzu, Sachiko Tanito, Masaki BMC Ophthalmol Case Report BACKGROUND: We report a case with two connected stents ejected simultaneously during an iStent inject W surgery, a modified second-generation iStent Trabecular Micro-Bypass System. CASE PRESENTATION: A 57-year-old woman with primary open-angle glaucoma underwent a combined cataract and iStent inject W surgery in her left eye. After the trabecular meshwork/Schlemm’s canal was pierced by the trocar of injector, the delivery button was pressed a first time, but the stent was not ejected. After the button was pressed a second time, connected two stents were ejected. After removing the dislocated stents from the anterior chamber, two stents were implanted into the desired places using another injector. Except for mild hyphema, no postoperative complication occurred. Stereomicroscopic observation showed that the two stents were connected by a broken trocar shaft. An X-ray showed that the trocar shaft was broken at the part referred to as the “sprayed trocar”. Scanning electron microscopy showed that the surface features of the broken trocar and trocar tip represented tensile failure. CONCLUSIONS: Although rare, considering that the damage was seen at the structurally weak part (i.e., sprayed trocar), the same phenomenon can happen. For patient safety, surgeons are recommended to inspect the device when the deployment of either the first or second stent is unsuccessful during the iStent inject surgery. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12886-023-02951-z. BioMed Central 2023-05-10 /pmc/articles/PMC10170729/ /pubmed/37165322 http://dx.doi.org/10.1186/s12886-023-02951-z Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Case Report
Shimada, Ayaka
Ichioka, Sho
Ishida, Akiko
Kaidzu, Sachiko
Tanito, Masaki
A case of two connected stents deployed during iStent inject W surgery
title A case of two connected stents deployed during iStent inject W surgery
title_full A case of two connected stents deployed during iStent inject W surgery
title_fullStr A case of two connected stents deployed during iStent inject W surgery
title_full_unstemmed A case of two connected stents deployed during iStent inject W surgery
title_short A case of two connected stents deployed during iStent inject W surgery
title_sort case of two connected stents deployed during istent inject w surgery
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10170729/
https://www.ncbi.nlm.nih.gov/pubmed/37165322
http://dx.doi.org/10.1186/s12886-023-02951-z
work_keys_str_mv AT shimadaayaka acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT ichiokasho acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT ishidaakiko acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT kaidzusachiko acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT tanitomasaki acaseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT shimadaayaka caseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT ichiokasho caseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT ishidaakiko caseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT kaidzusachiko caseoftwoconnectedstentsdeployedduringistentinjectwsurgery
AT tanitomasaki caseoftwoconnectedstentsdeployedduringistentinjectwsurgery