Validation of the HPLC Analytical Method for the Determination of Chemical and Radiochemical Purity of Ga-68-DOTATATE
PURPOSE OF THE STUDY: Ga-68-DOTA-peptides targeting somatostatin receptors have been assessed as a valuable tool in neuroendocrine tumors imaging using positron emission tomography (PET). A new selective and sensitive high-pressure liquid chromatography (HPLC) method was developed for determining ch...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10171759/ https://www.ncbi.nlm.nih.gov/pubmed/37180199 http://dx.doi.org/10.4103/ijnm.ijnm_11_22 |
_version_ | 1785039491512139776 |
---|---|
author | Sammartano, Antonino Migliari, Silvia Serreli, Giulio Scarlattei, Maura Baldari, Giorgio Ruffini, Livia |
author_facet | Sammartano, Antonino Migliari, Silvia Serreli, Giulio Scarlattei, Maura Baldari, Giorgio Ruffini, Livia |
author_sort | Sammartano, Antonino |
collection | PubMed |
description | PURPOSE OF THE STUDY: Ga-68-DOTA-peptides targeting somatostatin receptors have been assessed as a valuable tool in neuroendocrine tumors imaging using positron emission tomography (PET). A new selective and sensitive high-pressure liquid chromatography (HPLC) method was developed for determining chemical and radiochemical purity of Ga-68-DOTATATE (PET) tracer. The identification of peaks was achieved on a symmetry C18 column 3 μm 120Å (3.0 mm × 150 mm spherical particles) using (A) water + 0.1% trifluoroacetic acid (TFA) and (B) acetonitrile + 0.1% TFA, as the mobile phases at a flow rate of 0.600 mL/min and monitored at 220 nm. The run time was 16 min. MATERIALS AND METHODS: The method was validated to fulfill International Conference on Harmonization requirements and EDQM guidelines, and it included specificity, linearity, limit of detection (LOD), limit of quantification (LOQ), accuracy, and precision. RESULTS: The calibration curve was linear over the concentration range from 0.5 to 3 μg/ml, with a correlation coefficient (r2) equal to 0.999, average coefficient of variation (CV%) 2%, and average bias% did not deviate more than 5% for all concentrations. The LOD and LOQ for DOTATATE were 0.5 and 0.1 μg/mL, respectively. The method was considered precise, obtaining coefficients of variation between 0.22% and 0.52% for intraday and 0.20% and 0.61% for interday precision. Accuracy of method was confirmed with average bias% that did not deviate more than 5% for all concentrations. CONCLUSION: All results were acceptable and this confirmed that the method is suitable for its intended use in routine quality control of Ga-68-DOTATATE to guarantee the high quality of the finished product before release. |
format | Online Article Text |
id | pubmed-10171759 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-101717592023-05-11 Validation of the HPLC Analytical Method for the Determination of Chemical and Radiochemical Purity of Ga-68-DOTATATE Sammartano, Antonino Migliari, Silvia Serreli, Giulio Scarlattei, Maura Baldari, Giorgio Ruffini, Livia Indian J Nucl Med Original Article PURPOSE OF THE STUDY: Ga-68-DOTA-peptides targeting somatostatin receptors have been assessed as a valuable tool in neuroendocrine tumors imaging using positron emission tomography (PET). A new selective and sensitive high-pressure liquid chromatography (HPLC) method was developed for determining chemical and radiochemical purity of Ga-68-DOTATATE (PET) tracer. The identification of peaks was achieved on a symmetry C18 column 3 μm 120Å (3.0 mm × 150 mm spherical particles) using (A) water + 0.1% trifluoroacetic acid (TFA) and (B) acetonitrile + 0.1% TFA, as the mobile phases at a flow rate of 0.600 mL/min and monitored at 220 nm. The run time was 16 min. MATERIALS AND METHODS: The method was validated to fulfill International Conference on Harmonization requirements and EDQM guidelines, and it included specificity, linearity, limit of detection (LOD), limit of quantification (LOQ), accuracy, and precision. RESULTS: The calibration curve was linear over the concentration range from 0.5 to 3 μg/ml, with a correlation coefficient (r2) equal to 0.999, average coefficient of variation (CV%) 2%, and average bias% did not deviate more than 5% for all concentrations. The LOD and LOQ for DOTATATE were 0.5 and 0.1 μg/mL, respectively. The method was considered precise, obtaining coefficients of variation between 0.22% and 0.52% for intraday and 0.20% and 0.61% for interday precision. Accuracy of method was confirmed with average bias% that did not deviate more than 5% for all concentrations. CONCLUSION: All results were acceptable and this confirmed that the method is suitable for its intended use in routine quality control of Ga-68-DOTATATE to guarantee the high quality of the finished product before release. Wolters Kluwer - Medknow 2023 2023-02-24 /pmc/articles/PMC10171759/ /pubmed/37180199 http://dx.doi.org/10.4103/ijnm.ijnm_11_22 Text en Copyright: © 2023 Indian Journal of Nuclear Medicine https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Sammartano, Antonino Migliari, Silvia Serreli, Giulio Scarlattei, Maura Baldari, Giorgio Ruffini, Livia Validation of the HPLC Analytical Method for the Determination of Chemical and Radiochemical Purity of Ga-68-DOTATATE |
title | Validation of the HPLC Analytical Method for the Determination of Chemical and Radiochemical Purity of Ga-68-DOTATATE |
title_full | Validation of the HPLC Analytical Method for the Determination of Chemical and Radiochemical Purity of Ga-68-DOTATATE |
title_fullStr | Validation of the HPLC Analytical Method for the Determination of Chemical and Radiochemical Purity of Ga-68-DOTATATE |
title_full_unstemmed | Validation of the HPLC Analytical Method for the Determination of Chemical and Radiochemical Purity of Ga-68-DOTATATE |
title_short | Validation of the HPLC Analytical Method for the Determination of Chemical and Radiochemical Purity of Ga-68-DOTATATE |
title_sort | validation of the hplc analytical method for the determination of chemical and radiochemical purity of ga-68-dotatate |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10171759/ https://www.ncbi.nlm.nih.gov/pubmed/37180199 http://dx.doi.org/10.4103/ijnm.ijnm_11_22 |
work_keys_str_mv | AT sammartanoantonino validationofthehplcanalyticalmethodforthedeterminationofchemicalandradiochemicalpurityofga68dotatate AT migliarisilvia validationofthehplcanalyticalmethodforthedeterminationofchemicalandradiochemicalpurityofga68dotatate AT serreligiulio validationofthehplcanalyticalmethodforthedeterminationofchemicalandradiochemicalpurityofga68dotatate AT scarlatteimaura validationofthehplcanalyticalmethodforthedeterminationofchemicalandradiochemicalpurityofga68dotatate AT baldarigiorgio validationofthehplcanalyticalmethodforthedeterminationofchemicalandradiochemicalpurityofga68dotatate AT ruffinilivia validationofthehplcanalyticalmethodforthedeterminationofchemicalandradiochemicalpurityofga68dotatate |