Cortical suspensory button fixation has superior biomechanical properties to knotless anchor suture in anterior cruciate ligament repair: a biomechanical study

The purpose of our biomechanical study was to assess load-to-failure, stiffness, gap formation following cyclic loading, and the failure mechanism for anterior cruciate ligament (ACL) repair comparing the cortical suspensory button and knotless anchor suture. Eight Thiel’s embalmed paired cadaveric...

Descripción completa

Detalles Bibliográficos
Autores principales: Itthipanichpong, Thun, Thamrongskulsiri, Napatpong, Tangpornprasert, Pairat, Virulsri, Chanyaphan, Limskul, Danaithep, Kuptniratsaikul, Somsak, Tanpowpong, Thanathep
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10172190/
https://www.ncbi.nlm.nih.gov/pubmed/37165044
http://dx.doi.org/10.1038/s41598-023-34766-9
_version_ 1785039570447892480
author Itthipanichpong, Thun
Thamrongskulsiri, Napatpong
Tangpornprasert, Pairat
Virulsri, Chanyaphan
Limskul, Danaithep
Kuptniratsaikul, Somsak
Tanpowpong, Thanathep
author_facet Itthipanichpong, Thun
Thamrongskulsiri, Napatpong
Tangpornprasert, Pairat
Virulsri, Chanyaphan
Limskul, Danaithep
Kuptniratsaikul, Somsak
Tanpowpong, Thanathep
author_sort Itthipanichpong, Thun
collection PubMed
description The purpose of our biomechanical study was to assess load-to-failure, stiffness, gap formation following cyclic loading, and the failure mechanism for anterior cruciate ligament (ACL) repair comparing the cortical suspensory button and knotless anchor suture. Eight Thiel’s embalmed paired cadaveric knees from four cadavers were dissected. The specimens were assigned to undergo ACL repair either with cortical suspensory button or with knotless anchor suture. The Instron machine replicates cyclic loading and then determines the gap formation. Traction was applied until failure. The load-to-failure, stiffness, and modes of failure in both groups were recorded. The load-to-failure, stiffness, and gap formation were compared between the two groups using the student's t-test. The mean load-to-failure in the cortical suspensory button group was significantly higher than the knotless anchor suture group (212.96 ± 54.57 vs 44.57 ± 20.80, p value < 0.01). No statistically significant difference was found regarding gap formation following cyclic loading and stiffness between the cortical suspensory button group and the knotless anchor suture group. This biomechanical study showed a higher load-to-failure for the ACL repair with cortical suspensory button compared to ACL repair with knotless anchor suture, while no statistically significant difference was found regarding the gap formation following cyclic loading and the stiffness. The load-to-failure in both cortical suspensory button and knotless anchor suture are below regular daily activity load. Thus, an internal brace or external support is recommended during rehabilitation.
format Online
Article
Text
id pubmed-10172190
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Nature Publishing Group UK
record_format MEDLINE/PubMed
spelling pubmed-101721902023-05-12 Cortical suspensory button fixation has superior biomechanical properties to knotless anchor suture in anterior cruciate ligament repair: a biomechanical study Itthipanichpong, Thun Thamrongskulsiri, Napatpong Tangpornprasert, Pairat Virulsri, Chanyaphan Limskul, Danaithep Kuptniratsaikul, Somsak Tanpowpong, Thanathep Sci Rep Article The purpose of our biomechanical study was to assess load-to-failure, stiffness, gap formation following cyclic loading, and the failure mechanism for anterior cruciate ligament (ACL) repair comparing the cortical suspensory button and knotless anchor suture. Eight Thiel’s embalmed paired cadaveric knees from four cadavers were dissected. The specimens were assigned to undergo ACL repair either with cortical suspensory button or with knotless anchor suture. The Instron machine replicates cyclic loading and then determines the gap formation. Traction was applied until failure. The load-to-failure, stiffness, and modes of failure in both groups were recorded. The load-to-failure, stiffness, and gap formation were compared between the two groups using the student's t-test. The mean load-to-failure in the cortical suspensory button group was significantly higher than the knotless anchor suture group (212.96 ± 54.57 vs 44.57 ± 20.80, p value < 0.01). No statistically significant difference was found regarding gap formation following cyclic loading and stiffness between the cortical suspensory button group and the knotless anchor suture group. This biomechanical study showed a higher load-to-failure for the ACL repair with cortical suspensory button compared to ACL repair with knotless anchor suture, while no statistically significant difference was found regarding the gap formation following cyclic loading and the stiffness. The load-to-failure in both cortical suspensory button and knotless anchor suture are below regular daily activity load. Thus, an internal brace or external support is recommended during rehabilitation. Nature Publishing Group UK 2023-05-10 /pmc/articles/PMC10172190/ /pubmed/37165044 http://dx.doi.org/10.1038/s41598-023-34766-9 Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Article
Itthipanichpong, Thun
Thamrongskulsiri, Napatpong
Tangpornprasert, Pairat
Virulsri, Chanyaphan
Limskul, Danaithep
Kuptniratsaikul, Somsak
Tanpowpong, Thanathep
Cortical suspensory button fixation has superior biomechanical properties to knotless anchor suture in anterior cruciate ligament repair: a biomechanical study
title Cortical suspensory button fixation has superior biomechanical properties to knotless anchor suture in anterior cruciate ligament repair: a biomechanical study
title_full Cortical suspensory button fixation has superior biomechanical properties to knotless anchor suture in anterior cruciate ligament repair: a biomechanical study
title_fullStr Cortical suspensory button fixation has superior biomechanical properties to knotless anchor suture in anterior cruciate ligament repair: a biomechanical study
title_full_unstemmed Cortical suspensory button fixation has superior biomechanical properties to knotless anchor suture in anterior cruciate ligament repair: a biomechanical study
title_short Cortical suspensory button fixation has superior biomechanical properties to knotless anchor suture in anterior cruciate ligament repair: a biomechanical study
title_sort cortical suspensory button fixation has superior biomechanical properties to knotless anchor suture in anterior cruciate ligament repair: a biomechanical study
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10172190/
https://www.ncbi.nlm.nih.gov/pubmed/37165044
http://dx.doi.org/10.1038/s41598-023-34766-9
work_keys_str_mv AT itthipanichpongthun corticalsuspensorybuttonfixationhassuperiorbiomechanicalpropertiestoknotlessanchorsutureinanteriorcruciateligamentrepairabiomechanicalstudy
AT thamrongskulsirinapatpong corticalsuspensorybuttonfixationhassuperiorbiomechanicalpropertiestoknotlessanchorsutureinanteriorcruciateligamentrepairabiomechanicalstudy
AT tangpornprasertpairat corticalsuspensorybuttonfixationhassuperiorbiomechanicalpropertiestoknotlessanchorsutureinanteriorcruciateligamentrepairabiomechanicalstudy
AT virulsrichanyaphan corticalsuspensorybuttonfixationhassuperiorbiomechanicalpropertiestoknotlessanchorsutureinanteriorcruciateligamentrepairabiomechanicalstudy
AT limskuldanaithep corticalsuspensorybuttonfixationhassuperiorbiomechanicalpropertiestoknotlessanchorsutureinanteriorcruciateligamentrepairabiomechanicalstudy
AT kuptniratsaikulsomsak corticalsuspensorybuttonfixationhassuperiorbiomechanicalpropertiestoknotlessanchorsutureinanteriorcruciateligamentrepairabiomechanicalstudy
AT tanpowpongthanathep corticalsuspensorybuttonfixationhassuperiorbiomechanicalpropertiestoknotlessanchorsutureinanteriorcruciateligamentrepairabiomechanicalstudy