How Many Fenestrations Should I Make When Placing a Baerveldt Glaucoma Implant? A Laboratory Study

PURPOSE: This study investigates the effect of one versus two fenestrations on both fluid egress and opening pressure from a non-valved glaucoma implant. METHODS: In this laboratory study, we used an in vitro closed system comprised of ligated silicone tubing connected to a fluid reservoir and manom...

Descripción completa

Detalles Bibliográficos
Autores principales: C Yang, Michael, D Yang, Christopher, Y Lin, Ken
Formato: Online Artículo Texto
Lenguaje:English
Publicado: PUBLISHED BY KNOWLEDGE E 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10172796/
https://www.ncbi.nlm.nih.gov/pubmed/37181610
http://dx.doi.org/10.18502/jovr.v18i2.13181
Descripción
Sumario:PURPOSE: This study investigates the effect of one versus two fenestrations on both fluid egress and opening pressure from a non-valved glaucoma implant. METHODS: In this laboratory study, we used an in vitro closed system comprised of ligated silicone tubing connected to a fluid reservoir and manometer to simulate the tubing found in a Baerveldt glaucoma drainage implant. Fenestrations were created using an 8-0 Vicryl TG140-8 suture needle. Main outcome measures included volume of fluid egress and fenestration opening pressures, which were measured via micropipette and increasing pressure until fluid egress was observed. RESULTS: No significant difference was observed in fluid egress between tubing with one versus two fenestrations at pressures [Formula: see text] 40 mmHg. At 50 mmHg, a statistically significant difference was observed in fluid egress between tubing with one versus two fenestrations (P [Formula: see text] 0.05). The first fenestration opened at 10.5 [Formula: see text] 3.77 mmHg and the second fenestration opened at 28.83 [Formula: see text] 5.09 mmHg (average [Formula: see text] standard deviation). CONCLUSION: Our in vitro findings suggest there may exist a critical pressure [Formula: see text] 40 mmHg at which the second fenestration starts to play a significant role in fluid drainage. There may be no difference in the amount of fluid egress and effect on intraocular pressure between one or two tube fenestrations when preoperative intraocular pressure is [Formula: see text] 40 mmHg.