Cargando…
Agreement of somatic and renal near-infrared spectroscopy with reference blood samples during a controlled hypoxia sequence: a healthy volunteer study
Purpose: O3® Regional Oximetry (Masimo Corporation, California, USA) is validated for cerebral oximetry. We aimed to assess agreement of somatic and renal near-infrared spectroscopy with reference blood samples. Methods: O3 sensors were placed bilaterally on the quadriceps and flank of 26 healthy vo...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Netherlands
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10175462/ https://www.ncbi.nlm.nih.gov/pubmed/36463540 http://dx.doi.org/10.1007/s10877-022-00944-9 |
_version_ | 1785040217486393344 |
---|---|
author | De Keijzer, Ilonka N. Massari, Dario Niezen, Caren K. Bokkers, Reinoud P.H. Vos, Jaap Jan Scheeren, Thomas W.L. |
author_facet | De Keijzer, Ilonka N. Massari, Dario Niezen, Caren K. Bokkers, Reinoud P.H. Vos, Jaap Jan Scheeren, Thomas W.L. |
author_sort | De Keijzer, Ilonka N. |
collection | PubMed |
description | Purpose: O3® Regional Oximetry (Masimo Corporation, California, USA) is validated for cerebral oximetry. We aimed to assess agreement of somatic and renal near-infrared spectroscopy with reference blood samples. Methods: O3 sensors were placed bilaterally on the quadriceps and flank of 26 healthy volunteers. A stepped, controlled hypoxia sequence was performed by adding a mixture of nitrogen and room air to the breathing circuit. O3-derived oxygen saturation values were obtained at baseline and at six decremental saturation levels (5% steps). Blood samples (radial artery, iliac vein (somatic reference) and renal vein) were obtained at each step. Reference values were calculated as: 0.7 × venous saturation + 0.3 × arterial saturation. The agreement between O3-derived values with blood reference values was assessed by calculating root-mean-square error accuracy and Bland-Altman plots. Results: The root-mean-square error accuracy was 6.0% between quadriceps oxygen saturation and somatic reference values. The mean bias was 0.8%, with limits of agreement from -7.7 to 9.3%. These were 5.1% and 0.6% (-8.3 to 9.5%) for flank oxygen saturation and somatic reference values, respectively, and 7.7% and -4.9% (-15.0 to 5.2%) for flank oxygen saturation and renal reference values. The kidney depth was 3.1 ± 0.9 cm below the skin. Conclusion: O3 regional oximetry can be used on the quadriceps and flank to monitor somatic saturation, yet has a saturation-level dependent bias. O3-derived values obtained at the flank underestimated renal reference values. Additionally, it is unlikely that the flank sensors did directly measure renal tissue. Trial registration: Clinicaltrials.gov (NCT04584788): registered October 6th, 2020. |
format | Online Article Text |
id | pubmed-10175462 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Springer Netherlands |
record_format | MEDLINE/PubMed |
spelling | pubmed-101754622023-05-13 Agreement of somatic and renal near-infrared spectroscopy with reference blood samples during a controlled hypoxia sequence: a healthy volunteer study De Keijzer, Ilonka N. Massari, Dario Niezen, Caren K. Bokkers, Reinoud P.H. Vos, Jaap Jan Scheeren, Thomas W.L. J Clin Monit Comput Original Research Purpose: O3® Regional Oximetry (Masimo Corporation, California, USA) is validated for cerebral oximetry. We aimed to assess agreement of somatic and renal near-infrared spectroscopy with reference blood samples. Methods: O3 sensors were placed bilaterally on the quadriceps and flank of 26 healthy volunteers. A stepped, controlled hypoxia sequence was performed by adding a mixture of nitrogen and room air to the breathing circuit. O3-derived oxygen saturation values were obtained at baseline and at six decremental saturation levels (5% steps). Blood samples (radial artery, iliac vein (somatic reference) and renal vein) were obtained at each step. Reference values were calculated as: 0.7 × venous saturation + 0.3 × arterial saturation. The agreement between O3-derived values with blood reference values was assessed by calculating root-mean-square error accuracy and Bland-Altman plots. Results: The root-mean-square error accuracy was 6.0% between quadriceps oxygen saturation and somatic reference values. The mean bias was 0.8%, with limits of agreement from -7.7 to 9.3%. These were 5.1% and 0.6% (-8.3 to 9.5%) for flank oxygen saturation and somatic reference values, respectively, and 7.7% and -4.9% (-15.0 to 5.2%) for flank oxygen saturation and renal reference values. The kidney depth was 3.1 ± 0.9 cm below the skin. Conclusion: O3 regional oximetry can be used on the quadriceps and flank to monitor somatic saturation, yet has a saturation-level dependent bias. O3-derived values obtained at the flank underestimated renal reference values. Additionally, it is unlikely that the flank sensors did directly measure renal tissue. Trial registration: Clinicaltrials.gov (NCT04584788): registered October 6th, 2020. Springer Netherlands 2022-12-04 2023 /pmc/articles/PMC10175462/ /pubmed/36463540 http://dx.doi.org/10.1007/s10877-022-00944-9 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Original Research De Keijzer, Ilonka N. Massari, Dario Niezen, Caren K. Bokkers, Reinoud P.H. Vos, Jaap Jan Scheeren, Thomas W.L. Agreement of somatic and renal near-infrared spectroscopy with reference blood samples during a controlled hypoxia sequence: a healthy volunteer study |
title | Agreement of somatic and renal near-infrared spectroscopy with reference blood samples during a controlled hypoxia sequence: a healthy volunteer study |
title_full | Agreement of somatic and renal near-infrared spectroscopy with reference blood samples during a controlled hypoxia sequence: a healthy volunteer study |
title_fullStr | Agreement of somatic and renal near-infrared spectroscopy with reference blood samples during a controlled hypoxia sequence: a healthy volunteer study |
title_full_unstemmed | Agreement of somatic and renal near-infrared spectroscopy with reference blood samples during a controlled hypoxia sequence: a healthy volunteer study |
title_short | Agreement of somatic and renal near-infrared spectroscopy with reference blood samples during a controlled hypoxia sequence: a healthy volunteer study |
title_sort | agreement of somatic and renal near-infrared spectroscopy with reference blood samples during a controlled hypoxia sequence: a healthy volunteer study |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10175462/ https://www.ncbi.nlm.nih.gov/pubmed/36463540 http://dx.doi.org/10.1007/s10877-022-00944-9 |
work_keys_str_mv | AT dekeijzerilonkan agreementofsomaticandrenalnearinfraredspectroscopywithreferencebloodsamplesduringacontrolledhypoxiasequenceahealthyvolunteerstudy AT massaridario agreementofsomaticandrenalnearinfraredspectroscopywithreferencebloodsamplesduringacontrolledhypoxiasequenceahealthyvolunteerstudy AT niezencarenk agreementofsomaticandrenalnearinfraredspectroscopywithreferencebloodsamplesduringacontrolledhypoxiasequenceahealthyvolunteerstudy AT bokkersreinoudph agreementofsomaticandrenalnearinfraredspectroscopywithreferencebloodsamplesduringacontrolledhypoxiasequenceahealthyvolunteerstudy AT vosjaapjan agreementofsomaticandrenalnearinfraredspectroscopywithreferencebloodsamplesduringacontrolledhypoxiasequenceahealthyvolunteerstudy AT scheerenthomaswl agreementofsomaticandrenalnearinfraredspectroscopywithreferencebloodsamplesduringacontrolledhypoxiasequenceahealthyvolunteerstudy |