Evaluation and comparison of ocular biometric parameters obtained with Tomey OA-2000 in silicone oil-filled aphakic eyes
PURPOSE: To evaluate a new non-contact instrument (OA-2000) measuring the ocular biometry parameters of silicone oil (SO)-filled aphakic eyes, as compared with IOLMaster 700. METHODS: Forty SO-filled aphakic eyes of 40 patients were enrolled in this cross-sectional clinical trial. The axial length (...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2023
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10189968/ https://www.ncbi.nlm.nih.gov/pubmed/37194016 http://dx.doi.org/10.1186/s12886-023-02962-w |
_version_ | 1785043193292652544 |
---|---|
author | Xiong, Yongqun Lin, Yongdong Zhao, Zifeng Wang, Hongxi Zhang, Guihua |
author_facet | Xiong, Yongqun Lin, Yongdong Zhao, Zifeng Wang, Hongxi Zhang, Guihua |
author_sort | Xiong, Yongqun |
collection | PubMed |
description | PURPOSE: To evaluate a new non-contact instrument (OA-2000) measuring the ocular biometry parameters of silicone oil (SO)-filled aphakic eyes, as compared with IOLMaster 700. METHODS: Forty SO-filled aphakic eyes of 40 patients were enrolled in this cross-sectional clinical trial. The axial length (AL), central corneal thickness (CCT), keratometry ((flattest keratometry) Kf and (steep keratometry, 90° apart from Kf) Ks), and axis of the Kf (Ax1) were measured with OA-2000 and IOLMaster 700. The coefficient of variation (CoV) was calculated to assess the repeatability. The correlation was evaluated by the Pearson coefficient. Bland-Altman analysis and paired t test were used to analyze the agreements and differences of parameters measured by the two devices, respectively. RESULTS: The mean AL obtained with the OA-2000 was 23.57 ± 0.93 mm (range: 21.50 to 25.68 mm), and that obtained with the IOLMaster 700 was 23.69 ± 0.94 mm (range: 21.85 to 25.86 mm), resulting in a mean offset of 0.124 ± 0.125 mm (p < 0.001). The mean offset of CCT measured by OA-2000 and IOLMaster 700 was 14.6 ± 7.5 μm (p < 0.001). However, the Kf, Ks and Ax1 values from the two devices were comparable (p > 0.05). All the measured parameters of the two devices showed strong linear correlations (all r ≥ 0.966). The Bland-Altman analysis showed a narrow 95% limits of agreement (LoA) of Kf, Ks and AL, but 95%LoA of CCT and Ax1 was wide, which were − 29.3 ~ 0.1 μm and-25.9 ~ 30.7°respectively. The CoVs of the biometric parameters obtained with OA-2000 were lower than 1%. CONCLUSION: In SO-filled aphakic eyes, the ocular parameters (including AL, Kf, Ks, Ax1, and CCT) measured by the OA-2000 and IOLMaster 700 had a good correlation. Two devices had an excellent agreement on ocular biometric measurements of Kf, Ks and AL. The OA-2000 provided excellent repeatability of ocular parameters in SO-filled aphakic eyes. |
format | Online Article Text |
id | pubmed-10189968 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2023 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-101899682023-05-18 Evaluation and comparison of ocular biometric parameters obtained with Tomey OA-2000 in silicone oil-filled aphakic eyes Xiong, Yongqun Lin, Yongdong Zhao, Zifeng Wang, Hongxi Zhang, Guihua BMC Ophthalmol Research PURPOSE: To evaluate a new non-contact instrument (OA-2000) measuring the ocular biometry parameters of silicone oil (SO)-filled aphakic eyes, as compared with IOLMaster 700. METHODS: Forty SO-filled aphakic eyes of 40 patients were enrolled in this cross-sectional clinical trial. The axial length (AL), central corneal thickness (CCT), keratometry ((flattest keratometry) Kf and (steep keratometry, 90° apart from Kf) Ks), and axis of the Kf (Ax1) were measured with OA-2000 and IOLMaster 700. The coefficient of variation (CoV) was calculated to assess the repeatability. The correlation was evaluated by the Pearson coefficient. Bland-Altman analysis and paired t test were used to analyze the agreements and differences of parameters measured by the two devices, respectively. RESULTS: The mean AL obtained with the OA-2000 was 23.57 ± 0.93 mm (range: 21.50 to 25.68 mm), and that obtained with the IOLMaster 700 was 23.69 ± 0.94 mm (range: 21.85 to 25.86 mm), resulting in a mean offset of 0.124 ± 0.125 mm (p < 0.001). The mean offset of CCT measured by OA-2000 and IOLMaster 700 was 14.6 ± 7.5 μm (p < 0.001). However, the Kf, Ks and Ax1 values from the two devices were comparable (p > 0.05). All the measured parameters of the two devices showed strong linear correlations (all r ≥ 0.966). The Bland-Altman analysis showed a narrow 95% limits of agreement (LoA) of Kf, Ks and AL, but 95%LoA of CCT and Ax1 was wide, which were − 29.3 ~ 0.1 μm and-25.9 ~ 30.7°respectively. The CoVs of the biometric parameters obtained with OA-2000 were lower than 1%. CONCLUSION: In SO-filled aphakic eyes, the ocular parameters (including AL, Kf, Ks, Ax1, and CCT) measured by the OA-2000 and IOLMaster 700 had a good correlation. Two devices had an excellent agreement on ocular biometric measurements of Kf, Ks and AL. The OA-2000 provided excellent repeatability of ocular parameters in SO-filled aphakic eyes. BioMed Central 2023-05-16 /pmc/articles/PMC10189968/ /pubmed/37194016 http://dx.doi.org/10.1186/s12886-023-02962-w Text en © The Author(s) 2023 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Xiong, Yongqun Lin, Yongdong Zhao, Zifeng Wang, Hongxi Zhang, Guihua Evaluation and comparison of ocular biometric parameters obtained with Tomey OA-2000 in silicone oil-filled aphakic eyes |
title | Evaluation and comparison of ocular biometric parameters obtained with Tomey OA-2000 in silicone oil-filled aphakic eyes |
title_full | Evaluation and comparison of ocular biometric parameters obtained with Tomey OA-2000 in silicone oil-filled aphakic eyes |
title_fullStr | Evaluation and comparison of ocular biometric parameters obtained with Tomey OA-2000 in silicone oil-filled aphakic eyes |
title_full_unstemmed | Evaluation and comparison of ocular biometric parameters obtained with Tomey OA-2000 in silicone oil-filled aphakic eyes |
title_short | Evaluation and comparison of ocular biometric parameters obtained with Tomey OA-2000 in silicone oil-filled aphakic eyes |
title_sort | evaluation and comparison of ocular biometric parameters obtained with tomey oa-2000 in silicone oil-filled aphakic eyes |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10189968/ https://www.ncbi.nlm.nih.gov/pubmed/37194016 http://dx.doi.org/10.1186/s12886-023-02962-w |
work_keys_str_mv | AT xiongyongqun evaluationandcomparisonofocularbiometricparametersobtainedwithtomeyoa2000insiliconeoilfilledaphakiceyes AT linyongdong evaluationandcomparisonofocularbiometricparametersobtainedwithtomeyoa2000insiliconeoilfilledaphakiceyes AT zhaozifeng evaluationandcomparisonofocularbiometricparametersobtainedwithtomeyoa2000insiliconeoilfilledaphakiceyes AT wanghongxi evaluationandcomparisonofocularbiometricparametersobtainedwithtomeyoa2000insiliconeoilfilledaphakiceyes AT zhangguihua evaluationandcomparisonofocularbiometricparametersobtainedwithtomeyoa2000insiliconeoilfilledaphakiceyes |