Cargando…

Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model()

OBJECTIVE: Little is known about the cardiovascular changes after TEVAR and regarding the impact on aortic stiffness for different stent graft generations specifically, following changes in device design. The present study evaluated the stent graft induced aortic stiffening of two generations of the...

Descripción completa

Detalles Bibliográficos
Autores principales: Mandigers, Tim J., Conti, Michele, Allievi, Sara, Dedola, Francesca, Bissacco, Daniele, Bianchi, Daniele, Marconi, Stefania, Domanin, Maurizio, Van Herwaarden, Joost A., Auricchio, Ferdinando, Trimarchi, Santi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10199196/
https://www.ncbi.nlm.nih.gov/pubmed/37213485
http://dx.doi.org/10.1016/j.ejvsvf.2023.04.001
_version_ 1785044879928197120
author Mandigers, Tim J.
Conti, Michele
Allievi, Sara
Dedola, Francesca
Bissacco, Daniele
Bianchi, Daniele
Marconi, Stefania
Domanin, Maurizio
Van Herwaarden, Joost A.
Auricchio, Ferdinando
Trimarchi, Santi
author_facet Mandigers, Tim J.
Conti, Michele
Allievi, Sara
Dedola, Francesca
Bissacco, Daniele
Bianchi, Daniele
Marconi, Stefania
Domanin, Maurizio
Van Herwaarden, Joost A.
Auricchio, Ferdinando
Trimarchi, Santi
author_sort Mandigers, Tim J.
collection PubMed
description OBJECTIVE: Little is known about the cardiovascular changes after TEVAR and regarding the impact on aortic stiffness for different stent graft generations specifically, following changes in device design. The present study evaluated the stent graft induced aortic stiffening of two generations of the Valiant thoracic aortic stent graft. METHODS: This was an ex vivo porcine investigation using an experimental mock circulatory loop. Thoracic aortas of young healthy pigs were harvested and connected to the mock circulatory loop. At a 60 bpm heart rate and stable mean arterial pressure, baseline aortic characteristics were obtained. Pulse wave velocity (PWV) was calculated before and after stent graft deployment. Paired and independent sample t tests or their non-parametric alternatives were performed to test for differences where appropriate. RESULTS: Twenty porcine thoracic aortas were divided into two equal subgroups, in which a Valiant Captivia or a Valiant Navion stent graft was deployed. Both stent grafts were similar in diameter and length. Baseline aortic characteristics did not differ between the subgroups. Mean arterial pressure values did not change after either stent graft, while pulse pressures increased statistically significantly after Captivia (mean 44 ± 10 mmHg to 51 ± 13 mmHg, p = .002) but not after Navion. Mean baseline PWV increased after both Captivia (4.4 ± 0.6 m/s to 4.8 ± 0.7 m/s, p = .007) and Navion (4.6 ± 0.7 m/s to 4.9 ± 0.7 m/s, p = .002). There was no statistically significant difference in the mean percentage increase in PWV for either subgroup (8 ± 4% vs. 6 ± 4%, p = .25). CONCLUSION: These experimental findings showed no statistically significant difference in the percentage increase of aortic PWV after either stent graft generation and confirm that TEVAR increases aortic PWV. As a surrogate for aortic stiffness, this calls for further improvements in future thoracic aortic stent graft designs regarding device compliance.
format Online
Article
Text
id pubmed-10199196
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-101991962023-05-21 Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model() Mandigers, Tim J. Conti, Michele Allievi, Sara Dedola, Francesca Bissacco, Daniele Bianchi, Daniele Marconi, Stefania Domanin, Maurizio Van Herwaarden, Joost A. Auricchio, Ferdinando Trimarchi, Santi EJVES Vasc Forum Original Research OBJECTIVE: Little is known about the cardiovascular changes after TEVAR and regarding the impact on aortic stiffness for different stent graft generations specifically, following changes in device design. The present study evaluated the stent graft induced aortic stiffening of two generations of the Valiant thoracic aortic stent graft. METHODS: This was an ex vivo porcine investigation using an experimental mock circulatory loop. Thoracic aortas of young healthy pigs were harvested and connected to the mock circulatory loop. At a 60 bpm heart rate and stable mean arterial pressure, baseline aortic characteristics were obtained. Pulse wave velocity (PWV) was calculated before and after stent graft deployment. Paired and independent sample t tests or their non-parametric alternatives were performed to test for differences where appropriate. RESULTS: Twenty porcine thoracic aortas were divided into two equal subgroups, in which a Valiant Captivia or a Valiant Navion stent graft was deployed. Both stent grafts were similar in diameter and length. Baseline aortic characteristics did not differ between the subgroups. Mean arterial pressure values did not change after either stent graft, while pulse pressures increased statistically significantly after Captivia (mean 44 ± 10 mmHg to 51 ± 13 mmHg, p = .002) but not after Navion. Mean baseline PWV increased after both Captivia (4.4 ± 0.6 m/s to 4.8 ± 0.7 m/s, p = .007) and Navion (4.6 ± 0.7 m/s to 4.9 ± 0.7 m/s, p = .002). There was no statistically significant difference in the mean percentage increase in PWV for either subgroup (8 ± 4% vs. 6 ± 4%, p = .25). CONCLUSION: These experimental findings showed no statistically significant difference in the percentage increase of aortic PWV after either stent graft generation and confirm that TEVAR increases aortic PWV. As a surrogate for aortic stiffness, this calls for further improvements in future thoracic aortic stent graft designs regarding device compliance. Elsevier 2023-04-12 /pmc/articles/PMC10199196/ /pubmed/37213485 http://dx.doi.org/10.1016/j.ejvsvf.2023.04.001 Text en © 2023 The Author(s) https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Original Research
Mandigers, Tim J.
Conti, Michele
Allievi, Sara
Dedola, Francesca
Bissacco, Daniele
Bianchi, Daniele
Marconi, Stefania
Domanin, Maurizio
Van Herwaarden, Joost A.
Auricchio, Ferdinando
Trimarchi, Santi
Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model()
title Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model()
title_full Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model()
title_fullStr Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model()
title_full_unstemmed Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model()
title_short Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model()
title_sort comparison of two generations of thoracic aortic stent grafts and their impact on aortic stiffness in an ex vivo porcine model()
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10199196/
https://www.ncbi.nlm.nih.gov/pubmed/37213485
http://dx.doi.org/10.1016/j.ejvsvf.2023.04.001
work_keys_str_mv AT mandigerstimj comparisonoftwogenerationsofthoracicaorticstentgraftsandtheirimpactonaorticstiffnessinanexvivoporcinemodel
AT contimichele comparisonoftwogenerationsofthoracicaorticstentgraftsandtheirimpactonaorticstiffnessinanexvivoporcinemodel
AT allievisara comparisonoftwogenerationsofthoracicaorticstentgraftsandtheirimpactonaorticstiffnessinanexvivoporcinemodel
AT dedolafrancesca comparisonoftwogenerationsofthoracicaorticstentgraftsandtheirimpactonaorticstiffnessinanexvivoporcinemodel
AT bissaccodaniele comparisonoftwogenerationsofthoracicaorticstentgraftsandtheirimpactonaorticstiffnessinanexvivoporcinemodel
AT bianchidaniele comparisonoftwogenerationsofthoracicaorticstentgraftsandtheirimpactonaorticstiffnessinanexvivoporcinemodel
AT marconistefania comparisonoftwogenerationsofthoracicaorticstentgraftsandtheirimpactonaorticstiffnessinanexvivoporcinemodel
AT domaninmaurizio comparisonoftwogenerationsofthoracicaorticstentgraftsandtheirimpactonaorticstiffnessinanexvivoporcinemodel
AT vanherwaardenjoosta comparisonoftwogenerationsofthoracicaorticstentgraftsandtheirimpactonaorticstiffnessinanexvivoporcinemodel
AT auricchioferdinando comparisonoftwogenerationsofthoracicaorticstentgraftsandtheirimpactonaorticstiffnessinanexvivoporcinemodel
AT trimarchisanti comparisonoftwogenerationsofthoracicaorticstentgraftsandtheirimpactonaorticstiffnessinanexvivoporcinemodel