Characterization of US Food and Drug Administration Class I Recalls from 2018 to 2022 for Moderate- and High-Risk Medical Devices: A Cross-Sectional Study

BACKGROUND: Medical device recalls are initiated in response to safety concerns. Class I (highest severity) recalls imply a reasonable likelihood of serious adverse events or death associated with device use. Recalled devices must be identified, assessed, and corrected or removed, upon which a recal...

Descripción completa

Detalles Bibliográficos
Autores principales: Mooghali, Maryam, Ross, Joseph S, Kadakia, Kushal T, Dhruva, Sanket S
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10204764/
https://www.ncbi.nlm.nih.gov/pubmed/37229515
http://dx.doi.org/10.2147/MDER.S412802
Descripción
Sumario:BACKGROUND: Medical device recalls are initiated in response to safety concerns. Class I (highest severity) recalls imply a reasonable likelihood of serious adverse events or death associated with device use. Recalled devices must be identified, assessed, and corrected or removed, upon which a recall can be terminated. OBJECTIVE: To characterize Class I medical device recalls and corresponding recalled devices. METHODS: This was a cross-sectional study of Class I recalls posted on the Food and Drug Administration’s annual log from January 1, 2018 to June 30, 2022 for moderate-risk and high-risk medical devices. Devices were categorized by therapeutic use, need for implantation, and life-sustaining designation; recalls were categorized by reason, status, and time elapsed. RESULTS: There were 189 unique Class I medical device recalls, including 151 (79.9%) for moderate-risk and 34 (18.0%) for high-risk devices. Sixty-five (34.4%) recalls were for cardiovascular devices, 36 (19.0%) for implanted devices, and 37 (19.6%) for life-sustaining devices. The median number of device units recalled in the US per recall notice was 4620 (interquartile range [IQR], 578–42,591), with 11 (5.8%) recalls associated with more than 1 million device units. Overall, 125 (66.1%) devices had multiple recalls, with a median of 4 (IQR, 3–11) recalls issued per recalled device. As of September 15, 2022, 50 (26.5%) recalls were terminated, with a median of 24 (IQR, 17.3–30.8) months elapsed between recall initiation and termination. Recalls were terminated more commonly among devices recalled once compared to those recalled multiple times (36.2% vs 19.2%; p=0.02) and for recalls that recommended discontinuing further use of affected devices compared to those that recommended device assessment and/or education of affected population (31.8% vs 18.2%; p=0.04). CONCLUSION: High-severity medical device recalls are common and affect millions of device units annually in the US. Recall termination takes a significant amount of time, putting patients at risk for serious safety concerns.