Leadcare® II Comparison with Graphite Furnace Atomic Absorption Spectrophotometry for Blood Lead Measurement in Peruvian Highlands

Peru is one of the countries with the highest lead contamination in the world. Biological monitoring has limitations due to the shortage of laboratories with validated methodologies for the measurement of blood lead, and it is necessary to use alternative methods for its measurement in high-altitude...

Descripción completa

Detalles Bibliográficos
Autores principales: Rosales-Rimache, Jaime, Chavez-Ruiz, Manuel, Inolopú-Cucche, Jorge, Rabanal-Sanchez, Jhonatan, Rueda-Torres, Lenin, Sanchez-Holguin, Gloria
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer India 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10205923/
https://www.ncbi.nlm.nih.gov/pubmed/37234184
http://dx.doi.org/10.1007/s12291-022-01050-y
_version_ 1785046121724248064
author Rosales-Rimache, Jaime
Chavez-Ruiz, Manuel
Inolopú-Cucche, Jorge
Rabanal-Sanchez, Jhonatan
Rueda-Torres, Lenin
Sanchez-Holguin, Gloria
author_facet Rosales-Rimache, Jaime
Chavez-Ruiz, Manuel
Inolopú-Cucche, Jorge
Rabanal-Sanchez, Jhonatan
Rueda-Torres, Lenin
Sanchez-Holguin, Gloria
author_sort Rosales-Rimache, Jaime
collection PubMed
description Peru is one of the countries with the highest lead contamination in the world. Biological monitoring has limitations due to the shortage of laboratories with validated methodologies for the measurement of blood lead, and it is necessary to use alternative methods for its measurement in high-altitude cities. We aimed to compare the blood lead levels (BLL) measured by the LeadCare II (LC) method and Graphite Furnace Atomic Absorption Spectrometry (GF-AAS). We measured the BLL of 108 children from the city of La Oroya. The mean and median BLL for GF-AAS were 10.77 ± 4.18 and 10.44 µg/dL, respectively; for the LC method, the mean was 11.71 ± 4.28 and the median was 11.60 µg/dL. We found a positive linear correlation (Rho = 0.923) between both methods. Notwithstanding, the Wilcoxon test suggests a significant difference between both methods (ρ = 0.000). In addition, the Bland–Altman analysis indicates that there is a positive bias (0.94) in the LC method, and this method tends to overestimate the BLL. Likewise, we performed a generalized linear model to evaluate the influence of age and hemoglobin on BLL. We found that age and hemoglobin had a significant influence on BLL measured by the LC method. Finally, we used two non-parametric linear regression methods (Deming and Passing-Bablok regression) to compare the LC method with the GF-AAS. We found that these methods differ by at least a constant amount, and there would be a proportional difference between both. Although in general there is a positive linear correlation, the results of both methods differ significantly. Therefore, its use in cities located at high altitudes (higher than 2440 m.a.s.l.) would not be recommended.
format Online
Article
Text
id pubmed-10205923
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer India
record_format MEDLINE/PubMed
spelling pubmed-102059232023-05-25 Leadcare® II Comparison with Graphite Furnace Atomic Absorption Spectrophotometry for Blood Lead Measurement in Peruvian Highlands Rosales-Rimache, Jaime Chavez-Ruiz, Manuel Inolopú-Cucche, Jorge Rabanal-Sanchez, Jhonatan Rueda-Torres, Lenin Sanchez-Holguin, Gloria Indian J Clin Biochem Original Research Article Peru is one of the countries with the highest lead contamination in the world. Biological monitoring has limitations due to the shortage of laboratories with validated methodologies for the measurement of blood lead, and it is necessary to use alternative methods for its measurement in high-altitude cities. We aimed to compare the blood lead levels (BLL) measured by the LeadCare II (LC) method and Graphite Furnace Atomic Absorption Spectrometry (GF-AAS). We measured the BLL of 108 children from the city of La Oroya. The mean and median BLL for GF-AAS were 10.77 ± 4.18 and 10.44 µg/dL, respectively; for the LC method, the mean was 11.71 ± 4.28 and the median was 11.60 µg/dL. We found a positive linear correlation (Rho = 0.923) between both methods. Notwithstanding, the Wilcoxon test suggests a significant difference between both methods (ρ = 0.000). In addition, the Bland–Altman analysis indicates that there is a positive bias (0.94) in the LC method, and this method tends to overestimate the BLL. Likewise, we performed a generalized linear model to evaluate the influence of age and hemoglobin on BLL. We found that age and hemoglobin had a significant influence on BLL measured by the LC method. Finally, we used two non-parametric linear regression methods (Deming and Passing-Bablok regression) to compare the LC method with the GF-AAS. We found that these methods differ by at least a constant amount, and there would be a proportional difference between both. Although in general there is a positive linear correlation, the results of both methods differ significantly. Therefore, its use in cities located at high altitudes (higher than 2440 m.a.s.l.) would not be recommended. Springer India 2022-05-30 2023-07 /pmc/articles/PMC10205923/ /pubmed/37234184 http://dx.doi.org/10.1007/s12291-022-01050-y Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Research Article
Rosales-Rimache, Jaime
Chavez-Ruiz, Manuel
Inolopú-Cucche, Jorge
Rabanal-Sanchez, Jhonatan
Rueda-Torres, Lenin
Sanchez-Holguin, Gloria
Leadcare® II Comparison with Graphite Furnace Atomic Absorption Spectrophotometry for Blood Lead Measurement in Peruvian Highlands
title Leadcare® II Comparison with Graphite Furnace Atomic Absorption Spectrophotometry for Blood Lead Measurement in Peruvian Highlands
title_full Leadcare® II Comparison with Graphite Furnace Atomic Absorption Spectrophotometry for Blood Lead Measurement in Peruvian Highlands
title_fullStr Leadcare® II Comparison with Graphite Furnace Atomic Absorption Spectrophotometry for Blood Lead Measurement in Peruvian Highlands
title_full_unstemmed Leadcare® II Comparison with Graphite Furnace Atomic Absorption Spectrophotometry for Blood Lead Measurement in Peruvian Highlands
title_short Leadcare® II Comparison with Graphite Furnace Atomic Absorption Spectrophotometry for Blood Lead Measurement in Peruvian Highlands
title_sort leadcare® ii comparison with graphite furnace atomic absorption spectrophotometry for blood lead measurement in peruvian highlands
topic Original Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10205923/
https://www.ncbi.nlm.nih.gov/pubmed/37234184
http://dx.doi.org/10.1007/s12291-022-01050-y
work_keys_str_mv AT rosalesrimachejaime leadcareiicomparisonwithgraphitefurnaceatomicabsorptionspectrophotometryforbloodleadmeasurementinperuvianhighlands
AT chavezruizmanuel leadcareiicomparisonwithgraphitefurnaceatomicabsorptionspectrophotometryforbloodleadmeasurementinperuvianhighlands
AT inolopucucchejorge leadcareiicomparisonwithgraphitefurnaceatomicabsorptionspectrophotometryforbloodleadmeasurementinperuvianhighlands
AT rabanalsanchezjhonatan leadcareiicomparisonwithgraphitefurnaceatomicabsorptionspectrophotometryforbloodleadmeasurementinperuvianhighlands
AT ruedatorreslenin leadcareiicomparisonwithgraphitefurnaceatomicabsorptionspectrophotometryforbloodleadmeasurementinperuvianhighlands
AT sanchezholguingloria leadcareiicomparisonwithgraphitefurnaceatomicabsorptionspectrophotometryforbloodleadmeasurementinperuvianhighlands