Testing radiofrequency energy delivery and catheter entrapment in three left ventricular assist device models in mock circulation

FUNDING ACKNOWLEDGEMENTS: Type of funding sources: None. BACKGROUND: The use of Left Ventricular Assist Devices (LVAD) as a bridge to heart transplantation (HTx) or as a destination therapy is increasing. Near half of the patients had ventricular arrhythmias (VA) in the first year of LVAD support. A...

Descripción completa

Detalles Bibliográficos
Autores principales: Simsek, E, Yagmur, B, Kahraman, U, Candemir, A, Engin, C, Yagdi, T, Nalbantgil, S, Ozbaran, M
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10207026/
http://dx.doi.org/10.1093/europace/euad122.325
_version_ 1785046358124658688
author Simsek, E
Yagmur, B
Kahraman, U
Candemir, A
Engin, C
Yagdi, T
Nalbantgil, S
Ozbaran, M
author_facet Simsek, E
Yagmur, B
Kahraman, U
Candemir, A
Engin, C
Yagdi, T
Nalbantgil, S
Ozbaran, M
author_sort Simsek, E
collection PubMed
description FUNDING ACKNOWLEDGEMENTS: Type of funding sources: None. BACKGROUND: The use of Left Ventricular Assist Devices (LVAD) as a bridge to heart transplantation (HTx) or as a destination therapy is increasing. Near half of the patients had ventricular arrhythmias (VA) in the first year of LVAD support. Ablation is common method for recurrent ventricular tachycardias in LVAD patients. Even there are many case reports about successful VT ablations in LVAD patients there is no report about consequences of RF energy delivery inside LVAD or ablation catheter contact – entrapment to the rotor of the LVAD. METHOD: Mock circulation models were created for three mostly used LVAD models ( Figure 1) (HeartMate3, HeartMate2, HVAD) Sheep heart was located inside the mock circulation model and 4mm non-irrigated RF ablation catheter and RF generator used for RF energy delivery tissue ( sheep heart) , ostium, mid and rotor of the LVAD)(Figure 1). During RF delivery flow parameters of the LVAD systems were noted also external noncontact flow monitor located out flow canula of the LVAD. All LVADs out flows were tested for 24 hours after RF energy delivery for long term malfunction. Also RF ablation catheters were tested for mechanical and functional disfunctions after contact of the LVAD rotors. RESULTS: No malfunctions were observed during and after RF energy delivery inside the LVADs. Only LVAD flow parameters temporary decreased in HeartMate 3 when the catheters had a contact to the rotor of the LVAD. This phenomenon occurred irrespective of RF energy delivery status. There were no changes in HVAD and Heartmate 2. After 24 hours continuous testing of the flow of the LVADs any changes were not observed . Impedance changes were noticed during RF energy delivery. Impedance values were dropped when RF catheter moved from tissue to the ostium of the LVAD in all models. Small increases observed during contact to the rotor of the LVAD in HeartMate 3 ( Figure 2). No mechanical damage were observed on the tip of the catheters after contact to the rotors also RF lesion dimensions were compared before and after the contact and there were no differences. CONCLUSION: RF energy delivery and catheter entrapment of the RF catheters to the rotors of the LVADs caused no changes in LVAD functioning in ex-vivo conditions in a mock circulation. [Figure: see text] [Figure: see text]
format Online
Article
Text
id pubmed-10207026
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-102070262023-05-25 Testing radiofrequency energy delivery and catheter entrapment in three left ventricular assist device models in mock circulation Simsek, E Yagmur, B Kahraman, U Candemir, A Engin, C Yagdi, T Nalbantgil, S Ozbaran, M Europace 13.4.3 - Ablation of Ventricular Arrhythmias FUNDING ACKNOWLEDGEMENTS: Type of funding sources: None. BACKGROUND: The use of Left Ventricular Assist Devices (LVAD) as a bridge to heart transplantation (HTx) or as a destination therapy is increasing. Near half of the patients had ventricular arrhythmias (VA) in the first year of LVAD support. Ablation is common method for recurrent ventricular tachycardias in LVAD patients. Even there are many case reports about successful VT ablations in LVAD patients there is no report about consequences of RF energy delivery inside LVAD or ablation catheter contact – entrapment to the rotor of the LVAD. METHOD: Mock circulation models were created for three mostly used LVAD models ( Figure 1) (HeartMate3, HeartMate2, HVAD) Sheep heart was located inside the mock circulation model and 4mm non-irrigated RF ablation catheter and RF generator used for RF energy delivery tissue ( sheep heart) , ostium, mid and rotor of the LVAD)(Figure 1). During RF delivery flow parameters of the LVAD systems were noted also external noncontact flow monitor located out flow canula of the LVAD. All LVADs out flows were tested for 24 hours after RF energy delivery for long term malfunction. Also RF ablation catheters were tested for mechanical and functional disfunctions after contact of the LVAD rotors. RESULTS: No malfunctions were observed during and after RF energy delivery inside the LVADs. Only LVAD flow parameters temporary decreased in HeartMate 3 when the catheters had a contact to the rotor of the LVAD. This phenomenon occurred irrespective of RF energy delivery status. There were no changes in HVAD and Heartmate 2. After 24 hours continuous testing of the flow of the LVADs any changes were not observed . Impedance changes were noticed during RF energy delivery. Impedance values were dropped when RF catheter moved from tissue to the ostium of the LVAD in all models. Small increases observed during contact to the rotor of the LVAD in HeartMate 3 ( Figure 2). No mechanical damage were observed on the tip of the catheters after contact to the rotors also RF lesion dimensions were compared before and after the contact and there were no differences. CONCLUSION: RF energy delivery and catheter entrapment of the RF catheters to the rotors of the LVADs caused no changes in LVAD functioning in ex-vivo conditions in a mock circulation. [Figure: see text] [Figure: see text] Oxford University Press 2023-05-24 /pmc/articles/PMC10207026/ http://dx.doi.org/10.1093/europace/euad122.325 Text en © The Author(s) 2023. Published by Oxford University Press on behalf of the European Society of Cardiology. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs licence (https://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial reproduction and distribution of the work, in any medium, provided the original work is not altered or transformed in any way, and that the work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle 13.4.3 - Ablation of Ventricular Arrhythmias
Simsek, E
Yagmur, B
Kahraman, U
Candemir, A
Engin, C
Yagdi, T
Nalbantgil, S
Ozbaran, M
Testing radiofrequency energy delivery and catheter entrapment in three left ventricular assist device models in mock circulation
title Testing radiofrequency energy delivery and catheter entrapment in three left ventricular assist device models in mock circulation
title_full Testing radiofrequency energy delivery and catheter entrapment in three left ventricular assist device models in mock circulation
title_fullStr Testing radiofrequency energy delivery and catheter entrapment in three left ventricular assist device models in mock circulation
title_full_unstemmed Testing radiofrequency energy delivery and catheter entrapment in three left ventricular assist device models in mock circulation
title_short Testing radiofrequency energy delivery and catheter entrapment in three left ventricular assist device models in mock circulation
title_sort testing radiofrequency energy delivery and catheter entrapment in three left ventricular assist device models in mock circulation
topic 13.4.3 - Ablation of Ventricular Arrhythmias
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10207026/
http://dx.doi.org/10.1093/europace/euad122.325
work_keys_str_mv AT simseke testingradiofrequencyenergydeliveryandcatheterentrapmentinthreeleftventricularassistdevicemodelsinmockcirculation
AT yagmurb testingradiofrequencyenergydeliveryandcatheterentrapmentinthreeleftventricularassistdevicemodelsinmockcirculation
AT kahramanu testingradiofrequencyenergydeliveryandcatheterentrapmentinthreeleftventricularassistdevicemodelsinmockcirculation
AT candemira testingradiofrequencyenergydeliveryandcatheterentrapmentinthreeleftventricularassistdevicemodelsinmockcirculation
AT enginc testingradiofrequencyenergydeliveryandcatheterentrapmentinthreeleftventricularassistdevicemodelsinmockcirculation
AT yagdit testingradiofrequencyenergydeliveryandcatheterentrapmentinthreeleftventricularassistdevicemodelsinmockcirculation
AT nalbantgils testingradiofrequencyenergydeliveryandcatheterentrapmentinthreeleftventricularassistdevicemodelsinmockcirculation
AT ozbaranm testingradiofrequencyenergydeliveryandcatheterentrapmentinthreeleftventricularassistdevicemodelsinmockcirculation