Cargando…

Baseline local impedance and changes in local impedance are predictors of electrode tissue coupling in RF-ablation

FUNDING ACKNOWLEDGEMENTS: Type of funding sources: None. BACKGROUND: Radiofrequency (RF) current remains the most ablation technology in the electrophysiologic laboratory. Durable lesion formation without collateral tissue damage is key for obtaining efficient and safe procedures. RF power, duration...

Descripción completa

Detalles Bibliográficos
Autores principales: Bahlke, F, Wachter, A, Englert, F, Krafft, H, Popa, M, Erhard, N, Telishevska, M, Lengauer, S, Lennerz, C, Reents, T, Hessling, G, Deisenhofer, I, Bourier, F
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10207564/
http://dx.doi.org/10.1093/europace/euad122.711
_version_ 1785046484975091712
author Bahlke, F
Wachter, A
Englert, F
Krafft, H
Popa, M
Erhard, N
Telishevska, M
Lengauer, S
Lennerz, C
Reents, T
Hessling, G
Deisenhofer, I
Bourier, F
author_facet Bahlke, F
Wachter, A
Englert, F
Krafft, H
Popa, M
Erhard, N
Telishevska, M
Lengauer, S
Lennerz, C
Reents, T
Hessling, G
Deisenhofer, I
Bourier, F
author_sort Bahlke, F
collection PubMed
description FUNDING ACKNOWLEDGEMENTS: Type of funding sources: None. BACKGROUND: Radiofrequency (RF) current remains the most ablation technology in the electrophysiologic laboratory. Durable lesion formation without collateral tissue damage is key for obtaining efficient and safe procedures. RF power, duration and contact force are considered as main determinants of lesion size. Recently published data show that electrode tissue coupling (ETC) is another important parameter of lesion formation. This ex vivo model was created to investigate local impedance changes and lesion size when alternating ETC-levels. METHODS: RF-lesions were created using a force and local impedance sensing catheter in an ex vivo porcine cardiac model. The experimental setup consisted of a saline-filled container, a dispersive electrode, a heated thermostat and a circulation pump to imitate in vivo conditions. Global impedance was kept at 120 Ohm as well as the temperature at 37°C. RF-lesions were created using identic values of RF duration and contact force. A RF power of 20W, 30W, 40W, and 50W was used. The ETC levels were systematically varied between minor and full coupling. In minor ETC-level, only the distal end of the catheter is in contact with the tissue. In full ETC-level, the whole catheter tip is in contact with the tissue. All parameters (power, temperature, global and local impedance, contact force, ETC, lesion size) were measured constantly during application of RF-current, enabling real-time correlation of RF parameters and lesion size. In case of an audible steam pop, RF application was stopped. RESULTS: 8654 measurements of lesion depth and diameter were analyzed after creating 72 lesions. Lesion depth and diameter in full ETC-level were significantly higher compared to minor ETC-level (s. Table 1). Interestingly, lesions in lower power levels (20-30 Watt), but full ETC-level became wider and deeper than lesions in high power levels (40-50 Watts), but minor ETC-level (s. Figure 1). These observations were made despite shorter RF-duration (54.00 ± 13.07 s vs. 47.11 ± 17.57 s, p=0.032) in lesions with full ETC-level due to higher incidence of steam pops (14 vs. 6 steam pops, p = 0.032). Baseline LI and LI-drop were significantly higher in full ETC-level (Figure 2), but no relevant difference was observed in baseline GI. CONCLUSION: Baseline LI and LI changes during ablation provide information about ETC-levels and interaction of the RF ablation catheter with cardiac tissue. ETC-levels, as predicted by LI, had a significantly higher impact on RF lesion size than RF power or RF-duration. Therefore, LI should be monitored during RF ablation for estimating ETC-levels and lesion formation. [Figure: see text] [Figure: see text]
format Online
Article
Text
id pubmed-10207564
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-102075642023-05-25 Baseline local impedance and changes in local impedance are predictors of electrode tissue coupling in RF-ablation Bahlke, F Wachter, A Englert, F Krafft, H Popa, M Erhard, N Telishevska, M Lengauer, S Lennerz, C Reents, T Hessling, G Deisenhofer, I Bourier, F Europace 9.4.4 - Catheter Ablation of Arrhythmias FUNDING ACKNOWLEDGEMENTS: Type of funding sources: None. BACKGROUND: Radiofrequency (RF) current remains the most ablation technology in the electrophysiologic laboratory. Durable lesion formation without collateral tissue damage is key for obtaining efficient and safe procedures. RF power, duration and contact force are considered as main determinants of lesion size. Recently published data show that electrode tissue coupling (ETC) is another important parameter of lesion formation. This ex vivo model was created to investigate local impedance changes and lesion size when alternating ETC-levels. METHODS: RF-lesions were created using a force and local impedance sensing catheter in an ex vivo porcine cardiac model. The experimental setup consisted of a saline-filled container, a dispersive electrode, a heated thermostat and a circulation pump to imitate in vivo conditions. Global impedance was kept at 120 Ohm as well as the temperature at 37°C. RF-lesions were created using identic values of RF duration and contact force. A RF power of 20W, 30W, 40W, and 50W was used. The ETC levels were systematically varied between minor and full coupling. In minor ETC-level, only the distal end of the catheter is in contact with the tissue. In full ETC-level, the whole catheter tip is in contact with the tissue. All parameters (power, temperature, global and local impedance, contact force, ETC, lesion size) were measured constantly during application of RF-current, enabling real-time correlation of RF parameters and lesion size. In case of an audible steam pop, RF application was stopped. RESULTS: 8654 measurements of lesion depth and diameter were analyzed after creating 72 lesions. Lesion depth and diameter in full ETC-level were significantly higher compared to minor ETC-level (s. Table 1). Interestingly, lesions in lower power levels (20-30 Watt), but full ETC-level became wider and deeper than lesions in high power levels (40-50 Watts), but minor ETC-level (s. Figure 1). These observations were made despite shorter RF-duration (54.00 ± 13.07 s vs. 47.11 ± 17.57 s, p=0.032) in lesions with full ETC-level due to higher incidence of steam pops (14 vs. 6 steam pops, p = 0.032). Baseline LI and LI-drop were significantly higher in full ETC-level (Figure 2), but no relevant difference was observed in baseline GI. CONCLUSION: Baseline LI and LI changes during ablation provide information about ETC-levels and interaction of the RF ablation catheter with cardiac tissue. ETC-levels, as predicted by LI, had a significantly higher impact on RF lesion size than RF power or RF-duration. Therefore, LI should be monitored during RF ablation for estimating ETC-levels and lesion formation. [Figure: see text] [Figure: see text] Oxford University Press 2023-05-24 /pmc/articles/PMC10207564/ http://dx.doi.org/10.1093/europace/euad122.711 Text en © The Author(s) 2023. Published by Oxford University Press on behalf of the European Society of Cardiology. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs licence (https://creativecommons.org/licenses/by-nc-nd/4.0/), which permits non-commercial reproduction and distribution of the work, in any medium, provided the original work is not altered or transformed in any way, and that the work is properly cited. For commercial re-use, please contact journals.permissions@oup.com
spellingShingle 9.4.4 - Catheter Ablation of Arrhythmias
Bahlke, F
Wachter, A
Englert, F
Krafft, H
Popa, M
Erhard, N
Telishevska, M
Lengauer, S
Lennerz, C
Reents, T
Hessling, G
Deisenhofer, I
Bourier, F
Baseline local impedance and changes in local impedance are predictors of electrode tissue coupling in RF-ablation
title Baseline local impedance and changes in local impedance are predictors of electrode tissue coupling in RF-ablation
title_full Baseline local impedance and changes in local impedance are predictors of electrode tissue coupling in RF-ablation
title_fullStr Baseline local impedance and changes in local impedance are predictors of electrode tissue coupling in RF-ablation
title_full_unstemmed Baseline local impedance and changes in local impedance are predictors of electrode tissue coupling in RF-ablation
title_short Baseline local impedance and changes in local impedance are predictors of electrode tissue coupling in RF-ablation
title_sort baseline local impedance and changes in local impedance are predictors of electrode tissue coupling in rf-ablation
topic 9.4.4 - Catheter Ablation of Arrhythmias
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10207564/
http://dx.doi.org/10.1093/europace/euad122.711
work_keys_str_mv AT bahlkef baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT wachtera baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT englertf baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT kraffth baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT popam baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT erhardn baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT telishevskam baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT lengauers baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT lennerzc baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT reentst baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT hesslingg baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT deisenhoferi baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation
AT bourierf baselinelocalimpedanceandchangesinlocalimpedancearepredictorsofelectrodetissuecouplinginrfablation