Ventilation Difficulty Caused by Obstructed Heated Breathing Circuit

For perioperative hypothermia prevention, a heated, humidified breathing circuit equipped with a fluid-warming unit inside the inspiratory limb has been developed. We report a ventilation difficulty caused by an obstructed heated breathing circuit. Cotton surrounding the hot wire, temperature sensor...

Descripción completa

Detalles Bibliográficos
Autores principales: Kwon, Jiyeon, Kang, Eunsu, Shin, Sunghyun, Lee, Byeongcheol, Ko, Myoungjin, Kim, Sehun, Lee, Soojee
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10222461/
https://www.ncbi.nlm.nih.gov/pubmed/37241197
http://dx.doi.org/10.3390/medicina59050965
_version_ 1785049704464121856
author Kwon, Jiyeon
Kang, Eunsu
Shin, Sunghyun
Lee, Byeongcheol
Ko, Myoungjin
Kim, Sehun
Lee, Soojee
author_facet Kwon, Jiyeon
Kang, Eunsu
Shin, Sunghyun
Lee, Byeongcheol
Ko, Myoungjin
Kim, Sehun
Lee, Soojee
author_sort Kwon, Jiyeon
collection PubMed
description For perioperative hypothermia prevention, a heated, humidified breathing circuit equipped with a fluid-warming unit inside the inspiratory limb has been developed. We report a ventilation difficulty caused by an obstructed heated breathing circuit. Cotton surrounding the hot wire, temperature sensor, and fluid tubing in the distal inspiratory limb was irregularly thicker than that of a normal circuit and nearly blocked the lumen. Despite carrying out routine checks on the anesthesia workstation preoperatively, we failed to make a prediagnosis by omitting the flow test after changing the circuit. This case puts emphasis on a routine flow test with a meticulous examination of the heated breathing circuit before every procedure.
format Online
Article
Text
id pubmed-10222461
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-102224612023-05-28 Ventilation Difficulty Caused by Obstructed Heated Breathing Circuit Kwon, Jiyeon Kang, Eunsu Shin, Sunghyun Lee, Byeongcheol Ko, Myoungjin Kim, Sehun Lee, Soojee Medicina (Kaunas) Case Report For perioperative hypothermia prevention, a heated, humidified breathing circuit equipped with a fluid-warming unit inside the inspiratory limb has been developed. We report a ventilation difficulty caused by an obstructed heated breathing circuit. Cotton surrounding the hot wire, temperature sensor, and fluid tubing in the distal inspiratory limb was irregularly thicker than that of a normal circuit and nearly blocked the lumen. Despite carrying out routine checks on the anesthesia workstation preoperatively, we failed to make a prediagnosis by omitting the flow test after changing the circuit. This case puts emphasis on a routine flow test with a meticulous examination of the heated breathing circuit before every procedure. MDPI 2023-05-17 /pmc/articles/PMC10222461/ /pubmed/37241197 http://dx.doi.org/10.3390/medicina59050965 Text en © 2023 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Case Report
Kwon, Jiyeon
Kang, Eunsu
Shin, Sunghyun
Lee, Byeongcheol
Ko, Myoungjin
Kim, Sehun
Lee, Soojee
Ventilation Difficulty Caused by Obstructed Heated Breathing Circuit
title Ventilation Difficulty Caused by Obstructed Heated Breathing Circuit
title_full Ventilation Difficulty Caused by Obstructed Heated Breathing Circuit
title_fullStr Ventilation Difficulty Caused by Obstructed Heated Breathing Circuit
title_full_unstemmed Ventilation Difficulty Caused by Obstructed Heated Breathing Circuit
title_short Ventilation Difficulty Caused by Obstructed Heated Breathing Circuit
title_sort ventilation difficulty caused by obstructed heated breathing circuit
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10222461/
https://www.ncbi.nlm.nih.gov/pubmed/37241197
http://dx.doi.org/10.3390/medicina59050965
work_keys_str_mv AT kwonjiyeon ventilationdifficultycausedbyobstructedheatedbreathingcircuit
AT kangeunsu ventilationdifficultycausedbyobstructedheatedbreathingcircuit
AT shinsunghyun ventilationdifficultycausedbyobstructedheatedbreathingcircuit
AT leebyeongcheol ventilationdifficultycausedbyobstructedheatedbreathingcircuit
AT komyoungjin ventilationdifficultycausedbyobstructedheatedbreathingcircuit
AT kimsehun ventilationdifficultycausedbyobstructedheatedbreathingcircuit
AT leesoojee ventilationdifficultycausedbyobstructedheatedbreathingcircuit