Influence of Hand File Size on the Accuracy of Root ZX and iPex Electronic Apex Locators: An In Vitro Study

Objective: To investigate the effect of the hand file size on the accuracy of Root ZX (J. Morita Co., Kyoto, Japan) and iPex (NSK, Tochigi, Japan). Methods: Seventy-five single-rooted teeth were decoronated, and canals were coronally flared with Gates Glidden burs sizes 4, 3, and 2. Actual canal len...

Descripción completa

Detalles Bibliográficos
Autores principales: Siddiqui, Amna Y, Alothmani, Osama S
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2023
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10226783/
https://www.ncbi.nlm.nih.gov/pubmed/37256168
http://dx.doi.org/10.7759/cureus.39662
_version_ 1785050636515016704
author Siddiqui, Amna Y
Alothmani, Osama S
author_facet Siddiqui, Amna Y
Alothmani, Osama S
author_sort Siddiqui, Amna Y
collection PubMed
description Objective: To investigate the effect of the hand file size on the accuracy of Root ZX (J. Morita Co., Kyoto, Japan) and iPex (NSK, Tochigi, Japan). Methods: Seventy-five single-rooted teeth were decoronated, and canals were coronally flared with Gates Glidden burs sizes 4, 3, and 2. Actual canal length was determined by averaging two readings obtained by inserting K-file size 8 until its tip was apparent at the most coronal border of the apical foramen. The reference length was actual length-0.5 mm. The blinded operator utilized Root ZX and iPex following the manufacturer’s recommendations. Teeth were placed in sponge blocks soaked with Ringer’s solution. Canals were irrigated with 5% sodium hypochlorite. K-file size 8 was attached to the lip clip and introduced until the APEX/0.0 mark, then withdrawn to the 0.5 mark. A stable meter gauge for five seconds indicated an acceptable reading. Readings with sizes 10 and 15 were obtained afterward. All measurements were done twice, then averaged. Data analysis was done using ANOVA and a posthoc Bonferroni test with the significance level set at P<0.05. Results: For Root ZX, the mean length with size 8 was not significantly different from the mean reference length (P=0.205). The same was found for its mean length at size 10 (P=0.093). However, the mean Root ZX length with size 15 was significantly shorter than the mean reference length (P=0.019). Mean iPex lengths with sizes 8, 10, and 15 were all significantly shorter than the mean reference length (P=0.038, 0.006, and 0.02, respectively). Conclusion: The size of the hand file affected the precision of Root ZX and iPex.
format Online
Article
Text
id pubmed-10226783
institution National Center for Biotechnology Information
language English
publishDate 2023
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-102267832023-05-30 Influence of Hand File Size on the Accuracy of Root ZX and iPex Electronic Apex Locators: An In Vitro Study Siddiqui, Amna Y Alothmani, Osama S Cureus Dentistry Objective: To investigate the effect of the hand file size on the accuracy of Root ZX (J. Morita Co., Kyoto, Japan) and iPex (NSK, Tochigi, Japan). Methods: Seventy-five single-rooted teeth were decoronated, and canals were coronally flared with Gates Glidden burs sizes 4, 3, and 2. Actual canal length was determined by averaging two readings obtained by inserting K-file size 8 until its tip was apparent at the most coronal border of the apical foramen. The reference length was actual length-0.5 mm. The blinded operator utilized Root ZX and iPex following the manufacturer’s recommendations. Teeth were placed in sponge blocks soaked with Ringer’s solution. Canals were irrigated with 5% sodium hypochlorite. K-file size 8 was attached to the lip clip and introduced until the APEX/0.0 mark, then withdrawn to the 0.5 mark. A stable meter gauge for five seconds indicated an acceptable reading. Readings with sizes 10 and 15 were obtained afterward. All measurements were done twice, then averaged. Data analysis was done using ANOVA and a posthoc Bonferroni test with the significance level set at P<0.05. Results: For Root ZX, the mean length with size 8 was not significantly different from the mean reference length (P=0.205). The same was found for its mean length at size 10 (P=0.093). However, the mean Root ZX length with size 15 was significantly shorter than the mean reference length (P=0.019). Mean iPex lengths with sizes 8, 10, and 15 were all significantly shorter than the mean reference length (P=0.038, 0.006, and 0.02, respectively). Conclusion: The size of the hand file affected the precision of Root ZX and iPex. Cureus 2023-05-29 /pmc/articles/PMC10226783/ /pubmed/37256168 http://dx.doi.org/10.7759/cureus.39662 Text en Copyright © 2023, Siddiqui et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Dentistry
Siddiqui, Amna Y
Alothmani, Osama S
Influence of Hand File Size on the Accuracy of Root ZX and iPex Electronic Apex Locators: An In Vitro Study
title Influence of Hand File Size on the Accuracy of Root ZX and iPex Electronic Apex Locators: An In Vitro Study
title_full Influence of Hand File Size on the Accuracy of Root ZX and iPex Electronic Apex Locators: An In Vitro Study
title_fullStr Influence of Hand File Size on the Accuracy of Root ZX and iPex Electronic Apex Locators: An In Vitro Study
title_full_unstemmed Influence of Hand File Size on the Accuracy of Root ZX and iPex Electronic Apex Locators: An In Vitro Study
title_short Influence of Hand File Size on the Accuracy of Root ZX and iPex Electronic Apex Locators: An In Vitro Study
title_sort influence of hand file size on the accuracy of root zx and ipex electronic apex locators: an in vitro study
topic Dentistry
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10226783/
https://www.ncbi.nlm.nih.gov/pubmed/37256168
http://dx.doi.org/10.7759/cureus.39662
work_keys_str_mv AT siddiquiamnay influenceofhandfilesizeontheaccuracyofrootzxandipexelectronicapexlocatorsaninvitrostudy
AT alothmaniosamas influenceofhandfilesizeontheaccuracyofrootzxandipexelectronicapexlocatorsaninvitrostudy